Effect of Leadless Pacing on Heart Function

October 26, 2023 updated by: Universitaire Ziekenhuizen KU Leuven

Ventricular Mechanical Impact, Contraction Efficiency and Heart Function With Right Ventricular Pacing by MICRATMTPS. Comparison With Conventional RV Pacing

This prospective, un-blinded, randomized, noninferiority, single centre study with 12 month follow-up included patients who were scheduled for a single-chamber PM implant. Patients were 1:1 randomized to undergo the implantation of a conventional VVI pacing system (Medtronic, Advisa ADSR03) with a ventricular pacing lead (Medtronic, 5076-58) in the right ventricle (called 'conventional group') or to be implanted with a leadless Micra TPS (called 'Micra group').

The primary objective was to assess and compare the mechanical impact of right ventricular pacing using the Micra TPS versus a conventional PM on the left ventricular function.

Study Overview

Detailed Description

The primary endpoint was to assess the mechanical impact of the Micra TPS on the left ventricular function. The investigators measured the change of Global Longitudinal Strain (GLS) over one year after pacemaker implantation.

As secondary endpoint, the investigators studied the effect of the Micra TPS on the right ventricle using speckle tracking for 2-D myocardial motion estimation. The investigators also assessed conventionally the function of the right ventricle using tricuspid annular plane systolic excursion (TAPSE) in accordance with the guidelines of the European Association of Cardiovascular Imaging and American Society of Echocardiography. Evolution of clinical and biological parameters as NYHA (New York Heart Association) classification, '6 minutes 'walk distance and measurement of brain natriuretic peptide (NT-proBNP) test was also be performed at baseline and after one year follow-up.

Design

It was an interventional prospective not blinded randomized single centre study. The investigators randomized fifty patients into two groups (1:1 ratio) between May 2018 and December 2021:

Group 1 : 25 patients underwent the implantation of a conventional ventricular pacemaker (Medtronic) with placement of a ventricular lead Medtronic 5076-58 in the right ventricle.

Group 2 : 25 patients underwent the implantation of a leadless pacing system Micra TPS.

All implantation procedures were performed under local anaesthesia. To minimize the impact of the pacing site, the investigators targeted as first position the mid septum of the right ventricle. In the presence of inacceptable electrical measurements, an apical positioning was reached. All pacing positions were reviewed according to the Lieberman criteria.

Patients follow-up Patients were followed for a minimal period of 1 year (at 1 month after implantation, 6 months and 12 months as in routine care). Two-dimensional echocardiography was performed 1 day before implantation and at each visit. A 6 minutes' walk distance was performed at day 10 and at 1 year follow-up. As biological marker of the left ventricular function, brain natriuretic peptide (NT-proBNP) was also measured at baseline and at 1 year follow-up.

Device data collections:

The implant procedure duration as the procedure fluoroscopy times were also recorded. Follow-up of the electrical measurements was performed at the same time than the echocardiographic recording.

All complications (implantation failure, device or lead dislodgements, implantation related infection, death) were also prospectively documented.

Conventional 2D-Echocardiography:

Transthoracic echocardiography was performed using a commercially available ultrasound machine (VividTM E95, GE Vingmed Ultrasound AS; Epiq, Philips Medical Systems). Baseline and follow-up echocardiographic examinations were performed using the same machine. Conventional echocardiographic data were analysed according to the recommendations of the European Association of Cardiovascular Imaging. Standard parasternal long- and short-axis and apical four and two-chamber and long axis view were acquired. Left ventricular end-diastolic volume (LVEDV), end-systolic volume (LVESV) and left ventricular ejection fraction were calculated using Simpson's rule in a four-chamber view.

The ystolic function of the right ventricle was evaluated by tricuspid annular plane systolic excursion (TAPSE) in a standard apical four-chamber view. Finally, valve function and particularly tricuspid function was assessed following the European recommendations.

All echocardiographic values will be obtained from the average of measurements from six continuous normal cardiac beats.

Ethical considerations:

The investigators followed the declaration of Helsinki regarding human experimentation. Moreover, the investigators respected national legislation regarding clinical research and confidentiality of the data obtained in this study. All the participants were informed about the aims and procedures of the study and signed an informed consent form before the inclusion in the study. The local ethical committees reviewed and approved this protocol.

Study Type

Interventional

Enrollment (Actual)

51

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Leuven, Belgium, 3000
        • University Hospitals of Leuven

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Patients with a Class I or II indication for single chamber pacemaker following international guidelines

Exclusion Criteria:

  • patients with previously implanted cardiac devices / mechanical valves that would interfere with the echocardiographic measurements
  • patients with inadequate image quality at baseline that prohibited the assessment of myocardial mechanics using echocardiography
  • patients with a left ventricular ejection fraction (LVEF) ≤40% at baseline
  • patients considered to have a pre-existing condition challenging/precluding the implant of a conventional pacemaker
  • patients who refused or were not able to provide written informed consent.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Active Comparator: Conventional pacing
24 patients underwent the implantation of a conventional VVI pacing system (Medtronic) with placement of a ventricular pacing lead Medtronic 5076-58 in the right ventricle.
Implantation of a conventional ventricular pacemaker (VVI and VV(R) pacing mode) and right ventricular lead (Medtronic 5076-58)/ Echocardiographic and clinical follow-up during 12 months
Experimental: leadless pacing
27 patients underwent the implantation of a leadless pacing system Micra TPS
Implantation of a leadless pacemaker type Micra TPS. Echocardiographic and clinical follow-up during 12 months

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Assessment and comparison of the mechanical impact of right ventricular pacing using the Micra TPS versus a conventional PM on the left ventricular function.
Time Frame: 1 year
The primary endpoints were the change in left ventricular ejection fraction (LVEF) and global longitudinal strain (GLS)
1 year

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Evolution of the right ventricular, tricuspid valve and mitral valve function
Time Frame: 1 year
Echocardiographic measurements at baseline and 1-year follow-up
1 year
Evolution of NT-pro-BNP
Time Frame: 1 year
Measurement of NT-pro BNP at baseline and 1-year follow-up
1 year
Evolution of pacemaker performance
Time Frame: 1 year
Measurement of electrical parameters: pacing threshold, impedance and R-wave sensing. Measurement of percentage of ventricular pacing in both groups.
1 year

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Christophe Garweg, MD, PhD, University Hospitals of Leuven

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 1, 2018

Primary Completion (Actual)

November 30, 2021

Study Completion (Actual)

December 31, 2021

Study Registration Dates

First Submitted

October 20, 2023

First Submitted That Met QC Criteria

October 20, 2023

First Posted (Actual)

October 25, 2023

Study Record Updates

Last Update Posted (Actual)

October 31, 2023

Last Update Submitted That Met QC Criteria

October 26, 2023

Last Verified

October 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.