Feasibility Performance Study of ABL90 FLEX PLUS HEM (ABL90FLEXHEM)
January 7, 2026 updated by: Radiometer Medical ApS
Feasibility Performance Study Protocol for ABL90 FLEX PLUS HEM
This feasibility performance study evaluates the ABL90 FLEX PLUS HEM device for its analytical performance in a clinical setting.
The study aims to evaluate the performance of the hemolysis detection feature.
Study Overview
Status
Recruiting
Conditions
Detailed Description
The study is designed to assess the feasibility and performance of the ABL90 FLEX PLUS HEM device in a clinical environment.
It will involve testing whole blood samples using the device and comparing results against established reference method.
The primary focus is on validating the device's performance of the hemolysis detection function.
The study will be conducted at one selected clinical site with trained personnel and will follow a predefined protocol to ensure consistency and reliability of data collection.
The outcomes will inform future clinical studies aimed at supporting regulatory submissions and clinical adoption.
Study Type
Observational
Enrollment (Estimated)
95
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Contact
- Name: Vadim Fedulov
- Phone Number: +4526311594
- Email: vadim.fedulov@radiometer.dk
Study Locations
-
-
NV
-
Copenhagen, NV, Denmark, 2400
- Recruiting
- Bispebjerg Hospital
-
Contact:
- Theis Itenov, MD, PhD
- Phone Number: +45 51 34 19 45
- Email: theis.skovsgaard.itenov@regionh.dk
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
- Adult
- Older Adult
Accepts Healthy Volunteers
No
Sampling Method
Non-Probability Sample
Study Population
Adults, 18 years of age or older, that have been admitted to the hospital and undergo blood collection as part of their standard care.
Description
Inclusion Criteria:
- Subject must be 18 years of age or older
Informed consent is obtained:
- Directly from the subject, if they are competent and able to understand the provided information and voluntarily agree to participate, or
- If the subject is temporarily incapacitated and unable to provide informed consent at the time of sample collection, written informed consent will be obtained from the subject's legally designated representative (LDR). when the subject regains capacity after enrollment, they will be informed of study inclusion and given the opportunity to withdraw consent.
- Subject evaluated as suitable according to the protocol and for the study by the principal investigator or designee.
Exclusion Criteria:
- Subject where sample collection is evaluated by PI or designee to impose unnecessary risk.
- Subject with an invalid written informed consent or who has withdrawn consent.
- Subject with known pregnancy or who is breastfeeding.
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Adults, 18 years of age or older, that have been admitted to hospital and undergo blood collection
Adults, 18 years of age or older, that have been admitted to the hospital and undergo blood collection as part of their standard care will be eligible for participation.
Many of these patients are admitted to the Intensive Care Unit (ICU) with acute injury or critical illness and are temporarily unconscious or sedated at the time of admission or during recovery in the intermediate care unit and post-anesthesia care unit due to their medical condition and required treatment (e.g., mechanical ventilation or sedation) and have an A- and or a V-line established for routine sample collection without additional burden as part of their standard care.
|
The hemolysis detection function is a newly developed sample quality feature on the ABL90 FLEX PLUS HEM intended to detect hemolyzed samples and to improve the reliability of the results provided by blood gas analyzer.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Percent of samples correctly classified into hemolysis bins (0-4) using ABL90 FLEX PLUS HEM hemolysis flag
Time Frame: At time of sample analysis (single session per sample; Day 1)
|
Whole blood samples (native and contrived) are classified into five hemolysis bins based on reference ccfHb (mg/dL): bin 0 (0-100), bin 1 (101-200), bin 2 (201-400), bin 3 (401-1000), bin 4 (>1000).
"Correct classification" is the proportion of samples where the ABL90 FLEX PLUS HEM hemolysis flag matches the reference bin assignment.
|
At time of sample analysis (single session per sample; Day 1)
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Slope of change in ccfHb concentration over 0-45 minutes (mg/dL per minute)
Time Frame: 0-45 minutes after t=0 measurement (within one sample session; Day 1)
|
For each sample, ccfHb is measured on ABL90 FLEX PLUS HEM at t=0, 15, 30, and 45 minutes (±5 minutes).
The outcome is the slope (mg/dL/min) from linear regression of ccfHb versus time over 0-45 minutes.
|
0-45 minutes after t=0 measurement (within one sample session; Day 1)
|
|
Pearson correlation coefficient (r) between ABL90 ccfHb (mg/dL) and Cobas 8000 Hemolysis Index (HI)
Time Frame: At time of paired analysis per sample (Day 1)
|
Paired measurements are obtained from the same sample: ABL90 FLEX PLUS HEM measures ccfHb (mg/dL) in S65 mode, and Roche cobas 8000 reports Hemolysis Index (HI) using the hemolysis index assay.
The outcome is the Pearson correlation coefficient (r) across paired ABL90 ccfHb and Cobas HI values (unitless r).
|
At time of paired analysis per sample (Day 1)
|
|
Pearson correlation coefficient (r) between ccfHb and potassium concentration (cK+, mmol/L) after RMED aspiration manipulation
Time Frame: During RMED manipulation session (baseline to after 2 aspiration rounds; Day 1)
|
For each sample, ABL90 FLEX PLUS HEM measures ccfHb and potassium (cK+) at baseline (no manipulation), after 1 round of RMED aspiration manipulation, and after 2 rounds.
The outcome is Pearson correlation coefficient (r) between ccfHb and potassium values pooled across the three measurements (unitless r).
|
During RMED manipulation session (baseline to after 2 aspiration rounds; Day 1)
|
|
Percentage of native whole blood samples with hemolysis (ccfHb > 20 mg/dL)
Time Frame: At time of native sample analysis (Day 1)
|
Among native (non-contrived) leftover arterial and venous whole blood samples, the outcome is the percentage of samples with ccfHb > 20 mg/dL as measured on ABL90 FLEX PLUS HEM.
|
At time of native sample analysis (Day 1)
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Collaborators
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
August 4, 2025
Primary Completion (Estimated)
May 10, 2026
Study Completion (Estimated)
June 10, 2026
Study Registration Dates
First Submitted
September 11, 2025
First Submitted That Met QC Criteria
January 7, 2026
First Posted (Actual)
January 15, 2026
Study Record Updates
Last Update Posted (Actual)
January 15, 2026
Last Update Submitted That Met QC Criteria
January 7, 2026
Last Verified
January 1, 2026
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- DC-089793-ABL90FLEX-HEM-Rev3
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
Yes
product manufactured in and exported from the U.S.
No
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hemolysis
-
Guy's and St Thomas' NHS Foundation TrustUniversity of OxfordNot yet recruitingHemolysis Neonatal
-
Hadassah Medical OrganizationUnknown
-
National Institutes of Health Clinical Center (CC)CompletedIsoagglutinin-Mediated Hemolysis | Probiotic ToxicityUnited States
-
Hospices Civils de LyonCompletedChronic HemolysisFrance
-
Cedars-Sinai Medical CenterWithdrawnAcute Decompensated Heart Failure | Hemolysis IntravascularUnited States
-
Aswan UniversityRecruitingObstetric Hemostasis | Intravascular Hemolysis | Major Obstetric Hemorrhage | Manual Syringing Transfusion | Pneumatic Pressure Bag TransfusionEgypt
-
Sanquin-LUMC J.J van Rood Center for Clinical Transfusion...Leiden University Medical CenterCompleted
-
Medical University of GrazJoanneum Research Forschungsgesellschaft mbHCompleted
-
National Heart, Lung, and Blood Institute (NHLBI)Completed
-
CARE Pharma Shanghai Ltd.Recruiting