Validation of a Connected Patch, an Alternative to Conventional Monitoring

maanantai 3. helmikuuta 2020 päivittänyt: Hopital Foch
The hypothesis is that this connected patch could be used in surgical departments to detect early a cardiovascular or respiratory complication and therefore to treat it without delay. However, it has not been validated in a hospital setting; that is the purpose of this project. This step of validating the connected monitoring object is required in different places: post-interventional surveillance room that approaches a conventional hospital room, and operating room where artifacts are numerous. Validation will be done by comparing data from General Electric's conventional monitors, with data from patches.

Tutkimuksen yleiskatsaus

Tila

Peruutettu

Ehdot

Interventio / Hoito

Yksityiskohtainen kuvaus

When patients operated on outside intensive care or intensive care units are hospitalized, vital signs are collected intermittently by various conventional devices. During the follow-up period of these patients, it is possible that abnormal vital signs that may identify clinical deterioration may appear a few minutes to a few hours before a serious adverse event (infection etc.) occurs. These signals often go unnoticed due to infrequent monitoring.

The establishment of teams to ensure a rapid response would improve the safety of hospitalized patients for whom the detection of abnormal vital signs predictive of clinical deterioration would generate an alarm. This system is based on the identification of patients at risk and the rapid intervention of the team.

One of the crucial issues is therefore the speed of the alarm and its reliability.

The evolution of electronic technology offers the possibility of using inexpensive portable devices (connected patches) that can provide information on vital signs in real time. Vital signs collected by these devices could be analyzed to predict clinical deterioration prior to hospital staff visits, especially at night and on weekends when staffing levels are generally lower.

The connected patch used in this research is the Scaléo Médical company's device "myAngel VitalSigns". The long-term objective would be to be able to use this connected patch in surgical departments to detect early a complication, particularly of a cardiovascular or respiratory nature, and therefore to treat it without wasting time. However, it has not been validated in a hospital setting; that is the purpose of the project.

The objective of this research is therefore to validate Scaléo Médical's connected patch in different locations: post-interventional surveillance room that approaches a conventional hospital room, operating room where artifacts are numerous.

Opintotyyppi

Interventio

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskelupaikat

      • Suresnes, Ranska, 92150
        • Hôpital FOCH

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

18 vuotta - 85 vuotta (Aikuinen, Vanhempi Aikuinen)

Hyväksyy terveitä vapaaehtoisia

Ei

Sukupuolet, jotka voivat opiskella

Kaikki

Kuvaus

Inclusion Criteria:

  • patient over 18 and under 85 years of age
  • patient requiring general anesthesia for extra-thoracic surgery
  • the patient's position during the operation is in supine position only
  • patient affiliated to or benefiting from social security, excluding State Medical Assistance
  • patient with written consent

Exclusion Criteria:

  • pregnant or breastfeeding patients
  • patients known to have severe skin reactions to adhésives
  • patient whose surgical procedure prevents the patch from being placed (intervention on the thorax)
  • patients deprived of their liberty or under guardianship

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Muut
  • Jako: Ei käytössä
  • Inventiomalli: Yksittäinen ryhmätehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: Yhdistetyn korjaustiedoston vahvistus
Yhdistetyn laastarin käyttö verenpaineen, sykkeen, hengitystiheyden, valtimoiden happisaturaation ja lämpötilan keräämiseen

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Concordance of blood pressure value measured by 2 different devices after surgery
Aikaikkuna: 36 hours maximum after the surgery
Comparison between blood pressure measured by conventional medical monitoring device and by connected patch
36 hours maximum after the surgery
Concordance of heart rate value measured by 2 different devices after surgery
Aikaikkuna: 36 hours maximum after the surgery
Comparison between heart rate measured by conventional medical monitoring device and by connected patch
36 hours maximum after the surgery
Concordance of respiratory rate value measured by 2 different devices after surgery
Aikaikkuna: 36 hours maximum after the surgery
Comparison between respiratory rate measured by conventional medical monitoring device and by connected patch
36 hours maximum after the surgery
Concordance of arterial oxygen saturation value measured by 2 different devices after surgery
Aikaikkuna: 36 hours maximum after the surgery
Comparison between arterial oxygen saturation measured by conventional medical monitoring device and by connected patch
36 hours maximum after the surgery
Concordance of temperature value measured by 2 different devices after surgery
Aikaikkuna: 36 hours maximum after the surgery
Comparison between temperature measured by conventional medical monitoring device and by connected patch
36 hours maximum after the surgery

Toissijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Concordance of blood pressure value measured by 2 different devices during surgery
Aikaikkuna: During surgery
Comparison between blood pressure measured by conventional medical monitoring device and by connected patch
During surgery
Concordance of heart rate value measured by 2 different devices during surgery
Aikaikkuna: During surgery
Comparison between heart rate measured by conventional medical monitoring device and by connected patch
During surgery
Concordance of respiratory rate value measured by 2 different devices during surgery
Aikaikkuna: During surgery
Comparison between respiratory rate measured by conventional medical monitoring device and by connected patch
During surgery
Concordance of arterial oxygen saturation value measured by 2 different devices during surgery
Aikaikkuna: During surgery
Comparison between arterial oxygen saturation measured by conventional medical monitoring device and by connected patch
During surgery
Concordance of temperature value measured by 2 different devices during surgery
Aikaikkuna: During surgery
Comparison between temperature value measured by conventional medical monitoring device and by connected patch
During surgery
Blood pressure artifact
Aikaikkuna: From surgery to 36 hours maximum after the surgery
Value of blood pressure will be considered as artifact if systolic blood pressure value < 80 or > 200 and diastolic blood pressure > 110
From surgery to 36 hours maximum after the surgery
Heart rate artifact
Aikaikkuna: From surgery to 36 hours maximum after the surgery
Value of heart rate will be considered as artifact if value < 40 or > 140
From surgery to 36 hours maximum after the surgery
Respiratory rate artifact
Aikaikkuna: From surgery to 36 hours maximum after the surgery
Value of respiratory rate will be considered as artifact if value < 8 or > 36
From surgery to 36 hours maximum after the surgery
Arterial oxygen saturation artifact
Aikaikkuna: From surgery to 36 hours maximum after the surgery
Value of arterial oxygen saturation will be considered as artifact if value < 85 %
From surgery to 36 hours maximum after the surgery
Acceptability of connected patch: numerical scale
Aikaikkuna: During the surgery and through study completion
Evaluation of connected patch with a simple numerical scale between 0 and 10: 0 representing a very poor tolerance of the device and leading to premature interruption of the collection and 10 a perfect tolerance allowing the patch to be maintained for several consecutive days
During the surgery and through study completion

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Sponsori

Tutkijat

  • Päätutkija: Morgan Le Guen, Hôpital FOCH

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Odotettu)

Keskiviikko 1. tammikuuta 2020

Ensisijainen valmistuminen (Odotettu)

Perjantai 4. syyskuuta 2020

Opintojen valmistuminen (Odotettu)

Perjantai 4. syyskuuta 2020

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Tiistai 6. elokuuta 2019

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Perjantai 9. elokuuta 2019

Ensimmäinen Lähetetty (Todellinen)

Maanantai 12. elokuuta 2019

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Keskiviikko 5. helmikuuta 2020

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Maanantai 3. helmikuuta 2020

Viimeksi vahvistettu

Torstai 1. elokuuta 2019

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muut tutkimustunnusnumerot

  • 2017055F

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

Ei

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Yhdysvalloissa valmistettu ja sieltä viety tuote

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .

Kliiniset tutkimukset Patch validointi

3
Tilaa