- ICH GCP
- Registro de ensaios clínicos dos EUA
- Ensaio Clínico NCT02438228
Cardiac Output by Nine Different Pulse Contour Algorithms
Estimation of Cardiac Output by Nine Different Pulse Contour Algorithms Compared With Transpulmonary Thermodilution
Visão geral do estudo
Status
Condições
Intervenção / Tratamento
Descrição detalhada
After induction of anesthesia, a central venous catheter and a transpulmonary thermodilution catheter (Pulsion Medical Systems, Munich, Germany) are introduced in the right internal jugular vein and in the femoral artery, respectively. The thermodilution catheter is connected to a PiCCO2 monitor (Software version 1.3.0.8). Thereafter, placement of an esophageal doppler probe is performed and the probe is connected to a unique Monitoring System (Deltex Medical, Chichester, UK) with nine different pulse contour algorithms. Cardiac Output by esophageal doppler is used to calibrate the nine different pulse contour algorithms.
After induction of anesthesia and establishment of all monitoring devices, a passive leg raising maneouvre (PLR) is performed and hemodynamic variables including CO by transpulmonary thermodilution (COTPTD) and CO by nine different pulse contour algorithms (COX1-9) are recorded before, during and after PLR. Subsequently, measurements of COTPTD and COX1-9 are carried out every 10 minutes until the beginning of CPB. Fifteen minutes after weaning from cardiopulmonary bypass calibration of nine different pulse contour algorithms by esophageal doppler are performed again and measurements of CITPTD and CIPFX are restarted up to the end of the surgical Intervention.
Tipo de estudo
Inscrição (Real)
Estágio
- Não aplicável
Contactos e Locais
Locais de estudo
-
-
-
Hamburg, Alemanha, 20099
- Berthold Bein
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
- Patients > 18 years of age with a left ventricular ejection fraction ≥ 0.5.
- Patients undergoing coronary artery bypass grafting.
Exclusion Criteria:
- Patients <18 years of age and with a left ventricular ejection fraction ≤0.5 are not considered for the study.
- Emergency procedures and patients with hemodynamic instability requiring continuous pharmacologic support, intracardiac shunts, severe aortic-, tricuspid- or mitral stenosis or insufficiency and mechanical circulatory support were also excluded.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Outro: Cardiac output measurement
|
Measurement of cardiac output by transpulmonary thermodilution, nine different pulse contour algorithms and esophageal doppler before, during and after a PLR-maneouvre and every 10 minutes before and after cardiopulmonary Bypass.
|
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Prazo |
|---|---|
|
Accuracy of cardiac output in l/min by nine different pulse contour algorithms compared with transpulmonary thermodilution
Prazo: After induction of anesthesia until the end of surgery
|
After induction of anesthesia until the end of surgery
|
Colaboradores e Investigadores
Patrocinador
Investigadores
- Investigador principal: Berthold Bein, MD, PhD, Asklepios Klinik St. Georg
Publicações e links úteis
Publicações Gerais
- Critchley LA, Lee A, Ho AM. A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010 Nov;111(5):1180-92. doi: 10.1213/ANE.0b013e3181f08a5b. Epub 2010 Aug 24.
- Suehiro K, Tanaka K, Funao T, Matsuura T, Mori T, Nishikawa K. Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth. 2013 Aug;111(2):170-7. doi: 10.1093/bja/aet022. Epub 2013 Mar 10.
- Schloglhofer T, Gilly H, Schima H. Semi-invasive measurement of cardiac output based on pulse contour: a review and analysis. Can J Anaesth. 2014 May;61(5):452-79. doi: 10.1007/s12630-014-0135-8. Epub 2014 Mar 19.
- Smetkin AA, Hussain A, Kuzkov VV, Bjertnaes LJ, Kirov MY. Validation of cardiac output monitoring based on uncalibrated pulse contour analysis vs transpulmonary thermodilution during off-pump coronary artery bypass grafting. Br J Anaesth. 2014 Jun;112(6):1024-31. doi: 10.1093/bja/aet489. Epub 2014 Feb 13.
- Huang L, Critchley LA. An assessment of two Doppler-based monitors to track cardiac output changes in anaesthetised patients undergoing major surgery. Anaesth Intensive Care. 2014 Sep;42(5):631-9. doi: 10.1177/0310057X1404200514.
- Singer M. Oesophageal Doppler. Curr Opin Crit Care. 2009 Jun;15(3):244-8. doi: 10.1097/MCC.0b013e32832b7083.
- Broch O, Bein B, Gruenewald M, Masing S, Huenges K, Haneya A, Steinfath M, Renner J. Accuracy of Cardiac Output by Nine Different Pulse Contour Algorithms in Cardiac Surgery Patients: A Comparison with Transpulmonary Thermodilution. Biomed Res Int. 2016;2016:3468015. doi: 10.1155/2016/3468015. Epub 2016 Dec 28.
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo
Conclusão Primária (Real)
Conclusão do estudo (Real)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Estimativa)
Atualizações de registro de estudo
Última Atualização Postada (Estimativa)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Palavras-chave
Outros números de identificação do estudo
- AZ-138
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .
Ensaios clínicos em Débito cardíaco
-
Assiut UniversityAinda não está recrutandoCardiac CT TOF
Ensaios clínicos em Cardiac output measurement
-
Santa Barbara Cottage HospitalConcluídoHipertensão pulmonarEstados Unidos
-
Bodyport Inc.RecrutamentoDoenca renal | Doença renal terminal (ESRD) | Diálise Peritoneal (DP)Estados Unidos, Canadá
-
University of California, San FranciscoConcluído
-
St. Luke's-Roosevelt Hospital CenterConcluído
-
Bodyport Inc.Duke Clinical Research InstituteConcluídoInsuficiência cardíacaEstados Unidos
-
Medtronic Cardiac Rhythm and Heart FailureMedtronic International Trading SarlConcluídoFibrilação Atrial ParoxísticaTcheca
-
University Health Network, TorontoCanadian Institutes of Health Research (CIHR)Ainda não está recrutandoDoenças cardíacas | Doenças cardiovasculares | Conhecimentos, Atitudes e Práticas em Saúde | Comportamento, Saúde
-
Seattle Institute for Cardiac ResearchCardiac Science Corporation.ConcluídoFibrilação atrial | Síncope | Taquicardia supraventricular | Pré-síncopeEstados Unidos, Nova Zelândia
-
Medtronic Cardiac Rhythm and Heart FailureRescindidoInsuficiência cardíacaHolanda
-
Respicardia, Inc.ConcluídoApnéia do sono | Distúrbios Respiratórios do Sono | Respiração Periódica | Apneia central do sono | Respiração de Cheyne StokesEstados Unidos, Polônia, Alemanha, Itália