- ICH GCP
- Реестр клинических исследований США
- Клиническое испытание NCT04862689
Исследование перфузионного баллонного катетера Рингера (Ringer PTCA)
Проспективное многоцентровое исследование перфузионного баллонного катетера Рингера в одной группе (исследование Рингера PTCA)
The objective of this study is to demonstrate safe and effective use of the Ringer perfusion balloon catheter for dilatation/pre-dilatation of coronary stenoses during percutaneous coronary intervention (PCI).
Обзор исследования
Статус
Условия
Вмешательство/лечение
Подробное описание
A prospective, multicenter, single-arm clinical study. The study will be conducted in 3 to 5 investigational sites in the US and Canada and will enroll up to 60 participants. The population for this study is adult participants undergoing non-emergent percutaneous coronary intervention (PCI) as a stand-alone procedure or following non-emergent diagnostic angiography performed during the same procedure.
Тип исследования
Регистрация (Действительный)
Фаза
- Непригодный
Контакты и местонахождение
Места учебы
-
-
Ontario
-
Scarborough Village, Ontario, Канада, M1B 4Z8
- Heart Health Institute
-
Toronto, Ontario, Канада, M5B 1W8
- St. Michael's Hospital
-
Toronto, Ontario, Канада, M4N3M5
- Sunnybrook Heath Sciences Center
-
-
-
-
California
-
Torrance, California, Соединенные Штаты, 90505
- Torrance Memorial Medical Center
-
-
Massachusetts
-
Boston, Massachusetts, Соединенные Штаты, 02115
- Brigham & Women's Hospital
-
-
Missouri
-
Kansas City, Missouri, Соединенные Штаты, 64111
- St. Luke's Hospital
-
-
Washington
-
Seattle, Washington, Соединенные Штаты, 98196
- University of Washington Medical Center
-
-
Критерии участия
Критерии приемлемости
Возраст, подходящий для обучения
Принимает здоровых добровольцев
Полы, имеющие право на обучение
Описание
Inclusion Criteria:
- Male and Female subjects >18 years old.
- Patients undergoing clinically indicated non-emergent PCI as a stand-alone procedure or in combination with diagnostic angiography during the same procedure.
- Study target lesion characteristics: a) native de-novo coronary stenosis of at least 70% visually estimated stenosis, or at least 50% visually estimated stenosis and shown to be physiologically significant by flow-mediated criteria (FFR <0.80, IFR <0.91); b) estimated reference vessel diameter 2.0-4.5mm; c) lesion length <25mm; d) calcification not greater than mild.
- Per investigator assessment, the patient may benefit from inflation distal perfusion during balloon inflation. Examples include, but are not limited to: (i) LV systolic dysfunction with LVEF<50% by any modality; (ii) a territory subtended by the target vessel downstream from the target lesion that is estimated to be in excess of 25% of viable LV mass; (iii) prior hemodynamic or electrical instablility during intervention on the target lesion; (iv) the potential to avoid placement of a coronary stent if a stent-like result can be achieved through prolonged balloon inflation.
- Female subjects of childbearing potential must have a negative pregnancy test per standard of care for PCI and be practicing contraception.
- Informed consent provided.
Exclusion Criteria:
- Hemodynamic or arrhythmic instability within 48 hours or a myocardial infarction within 72 hours of PCI.
- More than 2 lesions planned during index PCI.
- Procedural complication developing prior to PCI of study target.
- Ejection fraction (EF) <25%.
- Creatinine clearance (Cr-Cl) <25 mg/dL.
- Baseline flow at study target <TIMI-2.
Учебный план
Как устроено исследование?
Детали дизайна
- Основная цель: Уход
- Распределение: Н/Д
- Интервенционная модель: Одногрупповое задание
- Маскировка: Нет (открытая этикетка)
Оружие и интервенции
Группа участников / Армия |
Вмешательство/лечение |
|---|---|
|
Экспериментальный: Experimental
Adult subjects clinically indicated for non-emergent percutaneous coronary intervention (PCI) as a stand-alone procedure or following non-emergent diagnostic angiography performed during the same procedure that, in the physician's estimation, requires prolonged balloon inflation with distal perfusion. |
Non-emergent Percutaneous Coronary Intervention (PCI) for treatment of a coronary artery lesion with Ringer Perfusion Balloon Catheter. |
Что измеряет исследование?
Первичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Rate of Device Success in Dilatation/Pre-dilatation of Coronary Stenosis During PCI.
Временное ограничение: During Procedure
|
Procedure success defined as successful delivery of the Ringer PBC to and across the target lesion; Successful inflation and deflation of the Ringer (and subsequent removal); No vessel perforation, no flow-limiting vessel dissection, no reduction in TIMI flow from baseline, and no clinically significant arrhythmias requiring medical treatment or device intervention following dilatation with the Ringer; Achievement of a final TIMI flow grade of 3 at the conclusion of the PCI procedure for the lesion. |
During Procedure
|
|
Rate of Ringer Related Complications Arising During Device Delivery
Временное ограничение: Discharge or 30 days, whichever comes first
|
Rate of Ringer related complications arising during delivery, inflation or removal of Ringer; freedom from major cardiac adverse events (MACE) |
Discharge or 30 days, whichever comes first
|
Вторичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Number of Successful PCI
Временное ограничение: During Procedure
|
Final residual stenosis ≤20% diameter stenosis (in-stent) or ≤50% diameter stenosis (stand-alone balloon) by visual angiographic estimation and freedom from MACE. |
During Procedure
|
|
TIMI Flow During Ringer Balloon Inflation
Временное ограничение: During Procedure
|
Maintenance of TIMI-2 or -3 flow into distal coronary bed during Ringer inflation |
During Procedure
|
|
Tolerance of Ringer Inflation of Greater Than One Minutes
Временное ограничение: During Procedure
|
Tolerance of at least one Ringer inflation greater than one minute (prolonged inflation) at target site (measured with symptoms, ECG, hemodynamics). |
During Procedure
|
Соавторы и исследователи
Спонсор
Следователи
- Директор по исследованиям: Sherry Lane, Teleflex
- Главный следователь: Kathleen Kearney, MD, University of Washington
Даты записи исследования
Изучение основных дат
Начало исследования (Действительный)
Первичное завершение (Действительный)
Завершение исследования (Действительный)
Даты регистрации исследования
Первый отправленный
Впервые представлено, что соответствует критериям контроля качества
Первый опубликованный (Действительный)
Обновления учебных записей
Последнее опубликованное обновление (Действительный)
Последнее отправленное обновление, отвечающее критериям контроля качества
Последняя проверка
Дополнительная информация
Термины, связанные с этим исследованием
Дополнительные соответствующие термины MeSH
Другие идентификационные номера исследования
- ST3170
Планирование данных отдельных участников (IPD)
Планируете делиться данными об отдельных участниках (IPD)?
Информация о лекарствах и устройствах, исследовательские документы
Изучает лекарственный продукт, регулируемый FDA США.
Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.
Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .