Closing the Loop in Adults With Type 1 Diabetes in the Home Setting

January 28, 2014 updated by: Hood Thabit, University of Cambridge

An Open-label, Three-centre, Randomised, Two-period, Crossover Study to Assess the Efficacy, Safety and Utility of Real-time Continuous Subcutaneous Glucose Monitoring Combined With Overnight Closed-loop Glucose Control in the Home Setting in Comparison With Real-time Continuous Subcutaneous Glucose Monitoring Alone in Adults With Type 1 Diabetes on Subcutaneous Insulin Infusion Pump Therapy

The main study objective is to compare real-time continuous subcutaneous glucose monitoring (CGM) combined with overnight automated closed-loop glucose control, and real-time CGM alone in the home setting.

Study Overview

Detailed Description

Achievement of tight glycaemic control in type 1 diabetes mellitus (T1D) using intensive insulin regimens, which has been shown to be important for the prevention of long term diabetes-related complications, is limited by a significantly increased risk of hypoglycaemia. The average patient with T1D suffers two symptomatic episodes of hypoglycaemia per week, and one episode of severe hypoglycaemia, defined as an event requiring assistance of another person to administer rescue treatment in the form of carbohydrate and/or glucagon, per year.Despite the rapid advancements in insulin pump technology and the ongoing development of more physiological insulin preparations, the currently available therapeutic regimens are still unable to achieve optimal glycaemic control.The emergence of continuous glucose monitoring (CGM) over the last decade, which enables users to view in real-time estimates of plasma glucose and receive alarms for impending hypo- or hyperglycaemia, thus facilitating appropriate changes in insulin therapy, is a major step towards improved diabetes monitoring.

The desirable goal is the development of an insulin delivery that is glucose responsive and the development of effective real time glucose monitoring should allow this. Glucose responsive insulin delivery should allow achievement of ideal glucose targets with less risk of hypoglycaemia. Closed-loop systems may provide a realistic treatment option for people with T1D. The research we are conducting at the University of Cambridge has been focused on developing a closed-loop system for overnight glucose control in patients with T1D. The studies that have been performed so far employ model predictive control (MPC) - this algorithm estimates patient-specific parameters from CGM measurements taken every 1 to 15 minutes and makes predictions of glucose excursions, which are then used to calculate basal insulin infusion rates. We hypothesize that overnight automated closed-loop glucose control in the home setting will be efficacious and safe compared to CGM alone, in T1D subjects on insulin pump treatment.

Study Type

Interventional

Enrollment (Actual)

24

Phase

  • Phase 2

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Cambridge, United Kingdom, CB2 0QQ
        • Addenbrooke's Hospital
      • London, United Kingdom, SE5 9RJ
        • King's College London
      • Sheffield, United Kingdom, S5 7AU
        • Northern General Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 65 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Type 1 diabetes, as defined by WHO for at least 6 months or confirmed C-peptide negative.
  • On insulin pump therapy for at least 3 months

Exclusion Criteria:

  • Non-type 1 diabetes mellitus
  • Any physical/psychological disease likely to interfere with the study
  • Taking medication likely to interfere with interpretation of the results
  • Known/suspected allergy against insulin
  • Patients with clinically significant nephropathy, neuropathy or retinopathy as judged by the investigator
  • Ongoing severe recurrent hypoglycaemia as judged by the investigator.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Closed loop (algorithm)
Subcutaneous delivery of Novorapid insulin, dose calculated by control algorithm, based on continuous glucose sensor readings.
Other Names:
  • Automated CL
Placebo Comparator: Open loop
Subcutaneous delivery of Novorapid insulin according to usual pump regime

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of CGM (Continuous glucose monitoring) values in target (3.9 - 8.0 mmol/l).
Time Frame: 4 weeks
Participants will be assessed overnight (00:00 to 07:00) for 4 weeks in each arm.
4 weeks

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of CGM values below 3.9 mmol/l.
Time Frame: 4 weeks
Participants will be assessed overnight (00:00 to 07:00) for 4 weeks in each arm.
4 weeks
Time spent with glucose levels below 3.9 mmol/l and above 8.0 mmol/l, as recorded by CGM and other CGM-based metrics
Time Frame: 4 weeks
Participants will be assessed overnight (00:00 to 07:00) for 4 weeks in each arm.
4 weeks
Glycaemic control assessed by fructosamine and HbA1c
Time Frame: 4 weeks
Participants will be assessed for 4 weeks in each arm.
4 weeks

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Roman Hovorka, PhD, MSc, BSc, University of Cambridge

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

December 1, 2012

Primary Completion (Actual)

December 1, 2013

Study Completion (Actual)

January 1, 2014

Study Registration Dates

First Submitted

September 20, 2011

First Submitted That Met QC Criteria

September 22, 2011

First Posted (Estimate)

September 26, 2011

Study Record Updates

Last Update Posted (Estimate)

January 29, 2014

Last Update Submitted That Met QC Criteria

January 28, 2014

Last Verified

January 1, 2014

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type 1 Diabetes Mellitus

Clinical Trials on Closed-loop

3
Subscribe