X-ray Dose Reduction Study for Endovascular Interventional Radiology

Clinical Validation and Evaluation of X-ray Image Processing for Endovascular Digital Subtraction Angiography and Intervention

ClarityIQ is a novel X-ray imaging technology, that combines advanced real-time image noise reduction algorithms, with state-of-the-art hardware to reduce patient entrance dose significantly. This is realized by anatomy-specific optimization of the full acquisition chain (grid switch, beam filtering, pulse width, spot size, detector and image processing engine) for every clinical task individually. Furthermore, smaller focal spot sizes, shorter pulses and the introduction of automatic real-time motion compensation in subtraction imaging are used, which are known to positively influence image quality . The final effect on the clinical image quality in peripheral angiography is investigated in this study.

Study Overview

Detailed Description

X-ray dose and image quality are related by laws of physics. Low dose and high image quality cannot be achieved at the same time. However, image processing algorithms can help an x-ray system to acquire images with lower dose without influencing image quality or achieving higher image quality with equal dose. The primary aim of this study is to verify if the diagnostic image quality is equal or better when using a significant reduction in X-ray dose for endovascular digital subtraction angiography in combination with a novel imaging technology compared to using normal dose with conventional image technology.

Study Type

Interventional

Enrollment (Actual)

51

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Nieuwegein, Netherlands, 3430 EM
        • St. Antonius ziekenhuis

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Patients older than 18 years of age undergoing digital subtraction angiography for diagnosis or treatment of iliac artery disease.

Exclusion Criteria:

  • Patients not willing or unable to give consent to participate
  • Patients already involved in a clinical trial
  • Patients under the age of 18
  • Pregnant or breastfeeding women
  • Patients with kidney disease (eGFR < 60)

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Non-Randomized
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: ClarityIQ
Low-dose DSA (83% reducation compared to normal dose) with novel X-ray imaging technology.
Digital Subtraction Angiography (DSA) of iliac artery with reduced dose settings (83% dose reduction expected) in combination with novel X-ray imaging technology.
Other Names:
  • ClarityIQ
Active Comparator: AlluraXper
Normal dose DSA with conventional X-ray imaging technology.
Digital Subtraction Angiography (DSA) of iliac artery with normal dose settings in combination with conventional X-ray imaging technology.
Other Names:
  • AlluraXper

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Image Quality
Time Frame: 1 Day
Overall proportion where diagnostic image quality of ClarityIQ is scored equal or better compared to AlluraXper by the blinded reviewers. Reading is performed by simultaneous visual comparison of image quality of AlluraXper and ClarityIQ by multiple blinded readers. All blinded readers have rated all side-by-side presented images. The images are presented in a random order and blinded to the reader. The hypothesis is that the overall proportion where diagnostic image quality of ClarityIQ is scored equal or better is ≥ than 0.80. Combined for all raters, the lower bound of the one-sided 95% CI (lower bound of the two-sided 90% CI) is used.
1 Day

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Radiation Dose Measurements: Dose Area Product (DAP)
Time Frame: Participants were followed for the duration of the procedure

Percentage dose change of ClarityIQ vs. AlluraXper in DAP calculated by DAP/frame for DSA. Negative change means a reduction in dose for ClarityIQ vs. AlluraXper.

DAP was measured during the ClarityIQ and AlluraXper runs during the procedure.

Participants were followed for the duration of the procedure
Radiation Dose Measurements: Air Kerma (AK)
Time Frame: Participants were followed for the duration of the procedure

Percentage dose change of ClarityIQ vs. AlluraXper in AK calculated by AK/frame for DSA. Negative change means a reduction in dose for ClarityIQ vs. AlluraXper.

AK was measured during the ClarityIQ and AlluraXper runs during the procedure.

Participants were followed for the duration of the procedure

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Marco van Strijen, MD, St. Antonius ziekenhuis

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

June 1, 2012

Primary Completion (Actual)

May 1, 2013

Study Completion (Actual)

May 1, 2013

Study Registration Dates

First Submitted

May 8, 2012

First Submitted That Met QC Criteria

May 15, 2012

First Posted (Estimate)

May 16, 2012

Study Record Updates

Last Update Posted (Actual)

March 8, 2022

Last Update Submitted That Met QC Criteria

March 7, 2022

Last Verified

March 1, 2022

More Information

Terms related to this study

Additional Relevant MeSH Terms

Other Study ID Numbers

  • XCY607-100093

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Cardiovascular Disease

Clinical Trials on Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology.

Subscribe