Left Ventricular Septum Pacing in Patients by Transvenous Approach Through the Inter-ventricular Septum

April 7, 2016 updated by: Maastricht University Medical Center

Left Ventricular Septum Pacing in Patients by Transvenous Approach Through the Inter-ventricular Septum - Feasibility, Long-term Lead Stability and Safety

Cardiac pacing is the only effective treatment for symptomatic bradycardia. The right ventricular apex (RVA) has become the most frequently used ventricular pacing site. However, RVA pacing has been shown to cause left ventricular (LV) dyssynchrony wich can lead to LV dysfunction and development of heart failure. Recent studies in animals have demonstrated that pacing at the LV septum induces significantly less ventricular dyssynchrony than RVA pacing and is able to improve LV function to a similar degree as biventricular (BiV) pacing. In addition it was shown that a LV septum lead can be placed permanently by driving a lead with extended helix from the RV side through the inter-ventricular septum into the LV endocardial layer. This was shown to be a feasible and safe procedure and lead stability was shown during four months of follow-up in otherwise healthy and active canines. LV septum pacing may therefore be a good treatment alternative in patients with symptomatic bradycardia, as well as patients with an indication for cardiac resynchronization therapy (CRT). The purpose of this study is to translate the findings from preclinical studies to the clinical situation by investigating the feasibility, long-term lead stability and safety of LV septum pacing by transvenous approach through the inter-ventricular septum in patients.

Study Overview

Detailed Description

Cardiac pacing is the only effective treatment for symptomatic bradycardia. After the introduction of the implantable pacemaker (PM), the right ventricular apex (RVA) has become the most frequently used ventricular pacing site. However, clinical studies have shown that RVA pacing leads to left ventricular (LV) dyssynchrony, and on the long run to adverse structural changes (remodeling), a higher risk of developing atrial fibrillation and heart failure, and higher mortality.

Recognition of the adverse effects of RVA pacing has generated interest in alternative ventricular pacing sites. Recent studies in animals have shown that pacing at the LV septum induces significantly less ventricular dyssynchrony than RVA pacing. In addition it was shown that a LV septum lead can be placed permanently by introducing a custom lead with extended helix transvenously and, after positioning it against the RV septum, driving it through the inter-ventricular septum into the LV endocardial layer. This was shown to be a feasible and safe procedure and lead stability was shown during four months of follow-up in otherwise healthy and active canines.

Animal studies have also shown that LV septum pacing is able to improve LV function to a similar degree as simultaneous biventricular (BiV) pacing used in cardiac resynchronization therapy (CRT). CRT was introduced to restore the abnormal ventricular activation and contraction in patients with heart failure and intra-ventricular conduction delay. Large clinical trials have shown that CRT improves LV systolic pump function, reverses structural remodelling, improves quality of life and exercise tolerance, and decreases mortality. Unfortunately, problems encountered during positioning and fixation of the LV pacing lead in the coronary vein result in suboptimal or loss of CRT in at least a quarter of CRT candidates and require re-operation in 7% during follow up.

The adverse effects of RVA pacing, the limitations of BiV pacing and the promising effects of LV septum pacing in preclinical studies have led to the idea that LV septum pacing may be a good treatment alternative in patients with symptomatic bradycardia, as well as patients with an indication for CRT. In the latter category, CRT could then be performed using a single ventricular pacing lead, thus limiting the number of lead implantations, and thereby reducing complication rate and implantation costs as well as avoiding the difficult access route through the coronary vein.

It is the aim of this study to translate the findings from preclinical studies to the clinical situation by investigating the feasibility, long-term lead stability and safety of LV septum pacing by transvenous approach through the inter-ventricular septum in patients with a PM or CRT indication. The results may have a large impact on future pacing therapy. The LV septum may become the universal pacing site, being preferred for anti-bradycardia therapy, and being an equal alternative for BiV pacing, but easier to apply, less invasive and more cost-effective.

Study Type

Observational

Enrollment (Actual)

12

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Limburg
      • Maastricht, Limburg, Netherlands, 6202 AZ
        • Maastricht University Medical Centre

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 75 years (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Study subjects are recruited from patients referred to the out-patient pacemaker/ICD clinic of our hospital.

Description

Inclusion Criteria:

  1. Cardiac pacing indication for sinus node dysfunction (dual chamber pacemaker) or indication for CRT (BiV pacemaker).

    • Patient with sinus node dysfunction:
    • Normal left ventricular ejection fraction (LVEF > 50%)
    • Left ventricular end-diastolic diameter (LVEDD) < 55mm
    • QRS duration < 100ms
    • Patient with CRT indication:
    • Non-ischemic cardiomyopathy
    • LVEF is < 35%
    • Typical left bundle-branch block on surface ECG
    • QRS duration >150ms
    • NYHA functional class I-III
  2. Greater than 18 years of age
  3. Stable sinus rhythm at the time of inclusion
  4. Willing and capable of giving informed consent

Exclusion Criteria:

  1. Ventricular pacing dependent
  2. High degree AV block
  3. Previous septal myocardial infarction
  4. Previously implanted pacing device
  5. Abnormal venous anatomy
  6. Left ventricular septum wall thickness >10mm determined by echocardiography
  7. Presence of severe valvular disease
  8. Presence of an ongoing progressive terminal disease associated with a reduced likelihood of survival for the duration of the study

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Sinus node dysfunction
Patients with sinus node dysfunction and structurally normal hearts
A left ventricular septum pacing lead is permanently placed by introducing a custom lead with extended helix transvenously and, after positioning it against the RV septum, driving it through the inter-ventricular septum into the LV endocardial layer.
Other Names:
  • Adapted Medtronic Select Secure 3830 pacing lead
CRT indication
Patients with an indication for CRT (heart failure with an LVEF <35% and LBBB)
A left ventricular septum pacing lead is permanently placed by introducing a custom lead with extended helix transvenously and, after positioning it against the RV septum, driving it through the inter-ventricular septum into the LV endocardial layer.
Other Names:
  • Adapted Medtronic Select Secure 3830 pacing lead

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Number of successful LV septum lead implantations
Time Frame: Baseline and 6 months
LV septum lead implantation is considered successful if the lead can be placed in the endocardial LV septum and adequate and stable sensing and pacing tresholds are achieved.
Baseline and 6 months
Change in sensing values, pacing threshold and lead impedance
Time Frame: Baseline and 6 months
Maintenance of stable lead function is determined by assessment of change in the electrical parameters: sensing value, pacing threshold and lead impedance at baseline versus 6 months follow-up.
Baseline and 6 months
Number of participants with procedure or lead related adverse events
Time Frame: Baseline and 6 months
The number of participant with and the kind of procedure- or lead related complications are assessed at the time of the procedure and up to 6 months after lead implantation. Special attention is given to the presence of perforation or rupture of the inter-ventricular septum or free wall, pericardial effusion, thrombosis, heart rhythm disturbances, conduction disturbances, and lead-/screw dislocation/-fracture.
Baseline and 6 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Acute change in LVdP/dtmax
Time Frame: Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours

Acute change in LV systolic function using the hemodynamic parameter LVdP/dtmax during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours
Acute change in LV stroke volume
Time Frame: Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours

Acute change in LV systolic function using the hemodynamic parameter LV stroke volume measured by Nexfin, during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours
Acute change in QRS duration
Time Frame: Baseline and 4 weeks

Acute change in sequence of LV electrical activation using the electrocardiographic parameter QRS duration, during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Baseline and 4 weeks
Acute change in 3-dimensional QRS vector direction
Time Frame: Baseline and 4 weeks

Acute change in sequence of LV electrical activation using the electrocardiographic parameter 3-dimensional QRS vector direction, during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Baseline and 4 weeks
Acute change in SPECKLE-tracking strain
Time Frame: Baseline and 4 weeks

Acute change in regional myocardial deformation patterns using the echocardiographic parameter SPECKLE-tracking strain, during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Baseline and 4 weeks
Difference between relative changes in LV stroke volume assessed by Nexfin and relative changes in invasively measured LVdp/dtmax.
Time Frame: Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours

Correlation between the relative change in non-invasively measured LV stroke volume using Nexfin and the relatieve change in invasively determined LV dP/dtmax, during LV septum pacing compared to:

  • RVA pacing and baseline (atrial pacing / AAI mode) in patients with SND.
  • RVA pacing, single LV epicardial lateral wall pacing, Bi-LV pacing (LV septum and LV lateral wall pacing), BiV pacing (RV septum and LV lateral wall pacing) and baseline (atrial pacing / AAI mode) in heart failure patients who are candidates for CRT.
Acute measurements are performed for the duration of the lead implantation procedure, an expected average of two hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Frits Prinzen, PhD, Maastricht University
  • Study Director: Kevin Vernooy, MD, PhD, Maastricht University Medical Centre

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

November 1, 2012

Primary Completion (ACTUAL)

February 1, 2015

Study Completion (ACTUAL)

February 1, 2015

Study Registration Dates

First Submitted

April 5, 2012

First Submitted That Met QC Criteria

May 31, 2012

First Posted (ESTIMATE)

June 1, 2012

Study Record Updates

Last Update Posted (ESTIMATE)

April 8, 2016

Last Update Submitted That Met QC Criteria

April 7, 2016

Last Verified

April 1, 2016

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Heart Failure

Clinical Trials on Left ventricular septum pacing lead