Inspiratory Flow Rates Achieved by the COPD Patients Through Breezhaler®, Ellipta® and Handihaler® Inhaler Devices

June 25, 2018 updated by: Novartis Pharmaceuticals

Multicenter Open-label Cross-over Study to Compare Inspiratory Flow Rates Achieved by the COPD Patients Through Breezhaler®, Ellipta® and Handihaler® Dry Powder Inhaler Devices

The purpose of this study was to compare dynamic inspiratory flow rates achieved by a population of Chronic Obstructive Pulmonary Disease (COPD) patients through the Breezhaler®, Ellipta® and Handihaler® dry powder inhaler (DPI) devices. No active drug or placebo were administered to patients in this study.

Study Overview

Detailed Description

The inspiratory flow rates achieved by the COPD patients, via the breath actuated devices, are one of the important factors that can influence the efficiency of the drug delivery by the inhalation route. The peak inspiratory flow (PIF) rates achieved by COPD patients through different breath actuated dry power inhalers may differ because of the differences in internal resistance of the different type of devices. The results of inspiratory airflow rate and pressure drop over time, from this study, provided an assessment of the range and variability of inhalation profile characteristics generated by COPD patients within the population selected for this study (e.g. demographics, gender, disease control) through the Breezhaler and the other marketed comparator DPIs.

The inhalation profiles were obtained through an Inhalation Profile Recorder (IPR). The inhalation profile recorder was a data acquisition device which consists of a computer, an interface unit and a pressure transducer. This allowed it to measure the real time dynamic pressure drop at the mouthpiece of the inhaler during an inhalation maneuver. On the graphic user interface, plots of inspiratory pressure drop versus time and flow rate versus time are displayed in real time. The investigator or designated study personnel at site was responsible for typing in the Test Location, selecting the Inhaler Type (Breezhaler, Ellipta, Handihaler), typing in the Patient ID and Patient Age, selecting Male or Female, and recording Patient Comments made during the test session, as necessary.

Each patient were required to generate three (03) inhalational profiles through each of the three (03) study devices: Breezhaler®, Ellipta® and Handihaler®.

Study Type

Interventional

Enrollment (Actual)

97

Phase

  • Phase 4

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Ciudad Autonoma de Bs As, Argentina, C1425FVH
        • Novartis Investigative Site
    • Buenos Aires
      • Caba, Buenos Aires, Argentina, C1426ABP
        • Novartis Investigative Site
      • Florida, Buenos Aires, Argentina, B1602DQD
        • Novartis Investigative Site
      • La Plata, Buenos Aires, Argentina, 1900
        • Novartis Investigative Site
      • Mar del Plata, Buenos Aires, Argentina, 7600
        • Novartis Investigative Site

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

40 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

- Current or ex-smokers who have a smoking history of at least 10 pack years (e.g., 10 pack years = 1 pack/day × 10 years, ½ pack/day × 20 years, etc.).

Note: A pack of cigarettes is equal to 20 cigarettes. Occasional smoking of cigars is not relevant to smoking history.

An ex-smoker is defined as a patient who has not smoked for ≥6 months at screening.

  • COPD patients with moderate to very severe airflow limitation (spirometric classification: GOLD 2, 3 or 4) at time of screening,

    • Post-bronchodilator FEV1 < 80% of the predicted normal, and
    • Post-bronchodilator FEV1/forced vital capacity (FVC) <0.70. (Post-bronchodilator refers to 1 hour after sequential inhalation of 84 µg ipratropium bromide and 400 µg salbutamol)
  • Willing patients assessed as suitable by investigator to reproducibly perform inhalational manoeuvers through study devices as required by the standard protocol
  • Willing patients assessed as suitable by investigator to comprehend and follow the instructions for use of the inhalational devices to be tested in the study

Exclusion Criteria:

  • Patients with a history of asthma or onset of respiratory symptoms prior to the age of 40 years
  • Use of short acting bronchodilating agent (SABA) as rescue (reliever) medication within 6 hours prior to or during inhalational profile assessments for the study [Note: Use of rescue medication should not be restricted if patient feels the need of the rescue/ reliever medication because of the diseased state. Safety and disease management should be priority and the suitability of such a patient for the study or relevance of the (study) assessments done for the study should be reviewed, as appropriate.]
  • Patients with any history of premature birth less than 33 weeks gestation or significant level of respiratory care including mechanical ventilation required as neonate affecting the respiratory tract or chronic lung diseases, which in the opinion of the investigator or designated study personnel at site may interfere with the study evaluation or optimal participation in the study.
  • Any major chronic illness including but not limited to a diagnosis of non-skin cancer, cystic fibrosis, bronchiectasis, α-1 anti-trypsin deficiency, myelomeningocele, sickle cell anemia, endocrine disease, congenital heart disease, unstable arrhythmia, congestive heart failure, stroke, severe hypertension, insulin-dependent diabetes mellitus, renal failure, liver disorders, immunodeficiency states, significant neurodevelopmental delay or behavioral disorders (excluding mild attention deficit hyperactivity disorder).
  • Patients who have had a COPD exacerbation that required treatment with antibiotics or oral corticosteroids or hospitalization in the 6 weeks prior to screening
  • Patients who, within 7 days prior to the screening visit (Visit 1) OR prior to Visit 2, increased use of rescue bronchodilators amounting to more than double the average number of puffs used in the preceding week or more than 8 puffs of SABA on any 3 consecutive days or more than 12 puffs of SABA on any 2 consecutive days
  • Respiratory tract infections (sinus, middle ear, oropharyngeal, upper or lower respiratory tract infection) within the 4weeks before the visit

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Breezhaler®
Each patient was required to inhale via Breezhaler® in a randomized cross-over sequence. The randomization numbers were generated using 6 sequences: BEH, EHB, HBE, BHE, EBH, HEB (B: Breezhaler, E: Ellipta, H: Handihaler).
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.
Other: Ellipta®
Each patient were required to inhale via Ellipta® in a randomized cross-over sequence. The randomization numbers were generated using 6 sequences: BEH, EHB, HBE, BHE, EBH, HEB (B: Breezhaler, E: Ellipta, H: Handihaler).
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.
Other: Handihaler®
Each patient were required to inhale via Handihaler® in a randomized cross-over sequence. The randomization numbers were generated using 6 sequences: BEH, EHB, HBE, BHE, EBH, HEB (B: Breezhaler, E: Ellipta, H: Handihaler).
No active drug or placebo were administered to patients during this study. Each patient were required to inhale via all three (03) study devices (Breezhaler®, Ellipta® and Handihaler®). Each patient was required to produce at least three (03) inhalational profiles via each inhalational device, following the training and demonstration for correct inhalational procedure, for each device.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Peak Inspiratory Flows Rates Summary by Inhalation Devices - PPS
Time Frame: Visit 2 (Day 1)
The peak inspiratory flows (PIF) rates obtained from the inhalation flow profiles generated by the COPD patients through the three dry powder inhalation (DPI) devices(Breezhaler®, Ellipta® and Handihaler®) were measured and compared (without drug or placebo administration). Each patient were required to generate inhalation flow profiles through all three DPI devices in a randomized cross-over sequence. The inspiratory measurements were taken in each of these devices in the same visit.
Visit 2 (Day 1)
Peak Inspiratory Flows Rates Summary by Inhalation Devices - FAS
Time Frame: Visit 2 (Day 1)

The peak inspiratory flows (PIF) rates obtained from the inhalation flow profiles generated by the COPD patients through the three dry powder inhalation (DPI) devices(Breezhaler®, Ellipta® and Handihaler®) were measured and compared (without drug or placebo administration). Each patient were required to generate inhalation flow profiles through all three DPI devices in a randomized cross-over sequence. The inspiratory measurements were taken in each of these devices in the same visit.

This FAS dataset includes PIF data from additional patients with corrected inhaler internal resistance values.

Visit 2 (Day 1)

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Study Director: Novartis Pharma, Novartis Pharmaceuticals

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

December 16, 2015

Primary Completion (Actual)

April 29, 2016

Study Completion (Actual)

April 29, 2016

Study Registration Dates

First Submitted

October 14, 2015

First Submitted That Met QC Criteria

November 2, 2015

First Posted (Estimate)

November 4, 2015

Study Record Updates

Last Update Posted (Actual)

June 26, 2018

Last Update Submitted That Met QC Criteria

June 25, 2018

Last Verified

June 1, 2018

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Pulmonary Disease, Chronic Obstructive (COPD)

Clinical Trials on Breezhaler®

Subscribe