- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT04080362
Safety and Performance Study of the MitralStitch Repair System.
A Prospective, Multicenter, Single Group Assignment Study for Evaluating the Safety and Effectiveness of MitralStitch Mitral Valve Repair System in Patients With Moderate to Severe and Severe Mitral Regurgitation.
Study Overview
Detailed Description
The clinical trial was designed as a prospective, multicenter, single group assignment study. The participants with moderate to severe and severe mitral valve regurgitation will receive a transapical beating-heart mitral valve repair surgery using MitralStitch mitral valve repair system after signed the informed consent. The follow-up will be conducted after 30 days, 3 months, 6 months and 12 months of the operation.
The following conditions evaluate the performance of the system and the safety and efficiency of using MitralStitch mitral valve repair system in treating patients with severe mitral regurgitation.
The none occurrence of these conditions will be approved as the main validity index: death, mitral valve related surgery without mitral valve failure and moderate or severe mitral regurgitation (MR > 2+). The secondary validity index was technical success rate, instrument success rate, surgical success rate, cardiac function improvement and quality of life improvement. And the safety evaluation indicators were assessed by the incidence of major adverse events, adverse events, serious adverse events, and device defects.
Study Type
Enrollment (Anticipated)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Beijing
-
Beijing, Beijing, China
- Structural Heart Disease Center, Fuwai Hospital
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Age ≥18 yrs
- Moderate to severe (3+) or severe (4+) mitral valve regurgitation confirmed by echocardiography (defined as MR >2+);
- NYHA functional class II-V (Appendix 3);
- Anatomically suitable for mitral valve repair;
- 40mm ≤LVESD≤55 mm; 25% ≤LVEF ≤60%; tolerable for small incision thoracic surgery;
- Mitral valve diameter ≤45mm;
- The independent expert committee of this study determined that surgical operation was contraindicated or high-risk, and the recommended reference standard was: surgical valve replacement with STS score ≥ 8; or surgical valve repair with STS score ≥ 6; or other risk factors (e.g., There are at least two moderate to severe indicators of weakness; potential operational disorders; at least major organ dysfunction that cannot be improved after surgery, etc.)
- The subject or the subject's legal representative fully understand and agree to join to the clinic trial.
Exclusion Criteria:
- Echocardiographic evidence of intracardiac mass, thrombus or vegetation;
- Severe mitral valve calcification;
- Concomitant with greater than moderate aortic stenosis or regurgitation;
- Mitral regurgitation caused by special pathological mechanisms, such as leaflet perforation and leaflet fissure;
- Severe untreated coronary artery stenosis requiring revascularization; or with other cardiovascular disease requiring surgical treatment;
- Hypertrophic cardiomyopathy with or without obstruction, restrictive cardiomyopathy, constrictive pericarditis, or other structural heart diseases that cause heart failure in patients expect dilated cardiomyopathy;
- Myocardial infarction was performed within 4 weeks before the intervention;
- Any vascular intervention, cardiac surgery, cardiac resynchronization therapy (CRT, CRT-D), and an implantation of implantable cardioverter defibrillator (ICD) were performed within 30 days before the intervention;
- Any heart transplantation, mitral valve surgery or percutaneous mitral valve operation were performed before the intervention;
- Life expectancy is less than 12 months;
- History of, or active, rheumatic heart disease;
- Active phase of bacterial endocarditis;
- The active infections require concurrent antibiotic treatment;
- History of acute peptic ulcer or upper gastrointestinal bleeding within 3 months;
- Dysplasia of the blood system, including granulocytopenia (WBC < 3×109/L), acute anemia (HB < 90g/L), thrombocytopenia (PLT < 50×109/L), hemorrhagic constitution, and coagulopathy;
- Modified Rankin scale ≥4;
- Pregnant or lactating women;
- The subject is suffered from a disease which may cause difficulty in evaluating the treatment (e.g., cancer, infection, severe metabolic disease, psychosis, etc.); or special cases were evaluated by the local experimental center heart team as not suitable for the surgical application of this clinical trial instrument;
- Participants who have participated in clinical trials of any drug and/or medical device within 1 month before this study;
- In the judgment of the Investigator, patients were not compliant enough to complete the study as required.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: MitralStitch repair system
Experimental group is allocated to use novel mitral vavle repair system manufactured by Hangzhou Valgen Medtech Co., Ltd
|
With transesophageal echocardiographic guidance,implanting ePTFE sutures as artificial neochordae or ege-to-ege repair using MitralStitch System
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The success rate
Time Frame: 1 year
|
Combined incidence of freedom from: death, surgery for Valve Dysfunction, and MR > 2+ (moderate to severe (3+) or severe (4+) mitral regurgitation) at 12 months
|
1 year
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
The instant procedural success
Time Frame: Immediately after repair
|
Technical success is defined as the absence of death, the successful implantation of the device and complete retraction of the delivery system without any need for emergency surgery or secondary intervention (including device related and surgical approach related
|
Immediately after repair
|
|
The success of the repair system
Time Frame: 12 months
|
The success rate of the device is defined as the successful delivery and implantation of the device, and the successful withdrawal of the implanted parts after the completion of the release, without any technical failure or complications related to the device.
|
12 months
|
|
The success of the surgery
Time Frame: 1 month
|
Surgical success was defined as the success of the instrument, MR 2+, and no major adverse events related to the instrument or operation occurred during the follow-up period after the implantation of the instrument.
|
1 month
|
|
New York Heart Association (NYHA) class
Time Frame: 12 months
|
NYHA grading was used to evaluate the cardiac function status of subjects
|
12 months
|
|
Living quality
Time Frame: 12 months
|
Postoperative SF-12 was used to evaluate the quality of life of the subjects
|
12 months
|
|
Incidence of major adverse events(MAE)
Time Frame: 12 months
|
MAE definition: Any complications related to devices or surgeries, including but not limited to: death, emergency surgery, severe cardiac tamponade requiring pericardiocentesis or operation, hemorrhage, stroke related to operation.
|
12 months
|
Collaborators and Investigators
Sponsor
Investigators
- Principal Investigator: Shengshou Hu, Chinese Academy of Medical Science, Fuwai hospital
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Anticipated)
Study Completion (Anticipated)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- MIRACLE_II
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Mitral Regurgitation
-
Ancora Heart, Inc.CompletedMitral Regurgitation | Mitral Valve Regurgitation | Functional Mitral RegurgitationAustria, Lithuania, Germany
-
Abbott Medical DevicesCompletedFunctional Mitral Regurgitation | Degenerrative Mitral RegurgitationRussian Federation
-
Xiao-dong ZhuangNot yet recruitingMitral Regurgitation | Functional Mitral Regurgitation | Degenerative Mitral Valve RegurgitationChina
-
Heinrich-Heine University, DuesseldorfCompletedMitral Regurgitation | Functional Mitral Regurgitation | Degenerative Mitral Valve RegurgitationGermany
-
Abbott Medical DevicesCompletedMitral Regurgitation | Mitral Valve RegurgitationUnited States
-
Mitre Medical Corp.BSWRI Cardiac Imaging Core Lab (CICL)Not yet recruitingIschemic Mitral Regurgitation | Functional Mitral RegurgitationUnited States
-
Abbott Medical DevicesCompletedMitral Valve Regurgitation | Mitral Valve Incompetence | Mitral Insufficiency | Functional Mitral Regurgitation | Mitral Stenosis With InsufficiencyJapan
-
Sun Yat-sen UniversityRecruitingMitral Insufficiency | Mitral Regurgitation (MR)China
-
Hospital Clinic of BarcelonaAbbottUnknownMITRAL REGURGITATIONSpain
-
Edwards LifesciencesActive, not recruitingMitral Regurgitation | Mitral Insufficiency | Functional Mitral Regurgitation | Degenerative Mitral Valve DiseaseUnited States, Canada, Switzerland, Germany
Clinical Trials on MitralStitch
-
Xijing HospitalUnknown