Evaluation of the Sevoflurane Consumption During General Anesthesia When Using the MemsorbTM Membrane (MEMSEV)

January 19, 2023 updated by: Philippe Richebe, Ciusss de L'Est de l'Île de Montréal

Evaluation of the Sevoflurane Consumption During General Anesthesia When Using the MemsorbTM Membrane and Low Gas Flow Versus the Classical Chemical Absorbents With 2L.Min-1 Gas Flow for CO2 Removal From the Anesthesia Circuit.

The study aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.

Study Overview

Detailed Description

Assessing the impact of anesthesia practice on global warming and carbon footprint becomes part of the standard of care and is a growing concern within the anesthesia community. Global Warming Potential (GWP) is a measure of how much a given mass of greenhouse gas contributes to global warming over a specified time period. The time period used, 20 versus 100 years, might drastically change the way we see impact of anesthesia on climate changes and GWP20 of CO2 is, by definition set at "1". Inhaled anesthetics have various GWP20: 349 for sevoflurane and 3714 for desflurane. These numbers might slightly change from one to another report/study. However, GWP20 and CDE20 alone are not sufficient to evaluate the environmental impact of anesthetic gases.

Other parameters must be included in the analysis: fresh gas flow (FGF), carrier gas (air, O2, N2O) and potency of the anesthetic gas. Unfortunately, the majority of trials did not fully consider the FGF reduction and the fact that desflurane can be administered with new closed or very low-flow anesthesia circuits as opposed as the recommended 2L.min-1 that must be used for sevoflurane according to its monography when classical chemical absorbents are used by the anesthesia team. Most of the calculations were made on a purely theoretical approach that could be different from actual measurements based on a strictly monitored anesthesia practice.

For sevoflurane, the standard FGF must be set at 2L.min-1 as there is still controversy concerning impact on renal function at lower flows if classical CO2 absorbents are used. It is still not recommended to administer sevoflurane during anesthesia with a FGF lower than 2L.min-1 in Canada when classical chemical absorbents are used according to Baxter monography when classical absorbents are used on the anesthesia circuit (http://www.baxter.ca/fr_CA/assets/downloads/monographs/Sevoflurane_FR.pdf). When continuous and accurate gas monitoring and analysis is used as recommended nowadays by all GCP guidelines (see Canadian guidelines for anesthesia practice), the use of closed or semi-closed-circuit anesthesia with very low FGF might allow for a reduction of more than 80% of the anesthetic gas administration and its consequent pollution.

Moreover, there are few clinical trials looking at the sparing effect on the consumption of anesthetic gases when the depth of anesthesia is properly monitored, with the bispectral index for instance. Indeed, and because the lack of appropriate and precise technology at the time of completion of these trials, most of the studies used only end-tidal concentration in % of the anesthetic gases to argue that they had decreased the consumption of gas during surgery. This remains a very indirect assessing method based on extrapolations as opposed to direct measurements. Studying the effect of the combination of BIS/NOL indices (depth of hypnosis / depth of analgesia) monitoring and the use of our Drager ventilators with low-FGF on precise consumption in mL of each gas in a clinical environment allows to get high quality data never reported in the past. The fact this study also uses the NOL index to ensure that level of analgesia is controlled and equivalent in all groups will also reinforce the idea that what this study measures in terms of anesthetic gas consumption is based on the real need for hypnosis for each participant, and not an overconsumption of gas because of poor control of nociception and analgesia.

Recently a new device was developed to extract CO2 from the ventilation-anesthesia circuit: the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada). All details on the technology of this new membrane are given in the Appendix 1 attached to this proposal (see at the bottom of the present text).

For the last 5 years, clinical trials have been conducted in humans on the use of this new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) and compared this membrane to the classical Drägersorb 800+ CO2 absorbent (Dräger, Lübeck, Germany). After REB committee approval and investigational testing authorization (ITA) by Health Canada, Dr O. Hung (from Halifax, Canada) reported out of 200 patients (100 with MemsorbTM, 100 with DraegersorbTM; ClinicalTrials.gov Identifier: NCT03014336) comparable data regarding the end-tidal CO2 with a median of 5.1% for memsorb™ and 5.0% for control (DraegersorbTM), stable over 2h of anesthesia. Vapor consumption data did not differ significantly between the 2 groups. Anesthesiologist used fresh gas flows at their discretion (from 0.3L.min-1 to 2.7L.min-1). As a conclusion to their study, Memsorb™ was shown to remove CO2 out of an anesthesia circuit as safely and as efficiently than the classical CO2 absorbent (Draegersorb) BUT without the known limitations of chemical absorbents at very low gas flow (0.3L.min-1).

The study here aims at determining whether replacing the classical chemical absorber Dräegersorb 800+ on Dräeger Perseus A500 machines (Dräeger, Lübeck, Germany) by the new membrane technology-based product (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) with the help of high-quality monitoring (BIS and NOL) and high-end ventilators (Dräeger Perseus A500 machines; Dräeger, Lübeck, Germany) that allow minimal fresh gas flow, will significantly decrease the use of sevoflurane and its related atmospheric pollution.

Indeed, MemsorbTM membrane allows to safely reduce the gas flow of the Dräger A500 ventilator as low as 0.2L.min-1 for the administration of sevoflurane (as there is no chemical reaction between the memsorb membrane and the sevoflurane, see ref 16) into the breathing circuit whereas the classical DräegersorbTM has to use a minimal gas flow of 2L.min-1 as recommended by the above cited monography (as below 2L.min-1, the Draegersorb might interact with sevoflurane and produce toxic compounds).

The impact of the present study will be that it will demonstrate sevoflurane administration at 0.2L.min-1 when using the MemsorbTM membrane (Memsorb™, DMF Medical Inc., Halifax, NS, Canada) reduces anesthesia related pollution to its minimum compared to the mandatory FGF at 2L.min-1 that MUST be used with the classical Draegersorb CO2 absorbent. As a consequence of this study, it is expected that anesthesiologists will drastically change towards the use of MemsorbTM in their clinical practice to significantly lessen the major impact they have on the environment.

The primary endpoint of this study is to show a significant decrease of at least 25% of the sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).

Study Type

Observational

Enrollment (Actual)

50

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Quebec
      • Montréal-Est, Quebec, Canada, H1T2M4
        • Hopital Maisonneuve Rosemont, CIUSSS de l'est de l'Ile de Montréal

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Enrollment of 50 randomized adult patients undergoing general anesthesia for laparoscopic general, gynecological or urologic surgery.

Type of surgery: laparoscopic general, gynecological and urologic surgery with general anesthesia that does not require additional regional anesthesia such as an epidural analgesia.

Description

Inclusion Criteria:

  • ASA 1-3,
  • Laparoscopic general, gynecological or urologic surgery requiring general anesthesia without additional regional anesthesia technique,
  • Fully consented,
  • BMI < 40,
  • Age > 18.

Exclusion Criteria:

  • Allergy or contra-indication to any drug used in the study protocol,
  • History of unstable coronary artery disease,
  • Serious cardiac arrhythmia (including atrial fibrillation),
  • History of substance abuse in the last 2 months prior to surgery,
  • Chronic use of psychotropic and/or opioid drugs (still existing within the last month prior to surgery,
  • History of psychiatric diseases,
  • History of refractory PONV in previous surgery,
  • Allergy to any drug used in the study protocol,
  • Non-scheduled surgery,
  • Refusal of the patient for participation in the study.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Case-Control
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Memsorb
M for memsorbTM group with the minimal gas flow possible (sevoflurane administration at 0.2L.min-1) with the ventilator Draeger A-500 Perseus,
Sevoflurane administration at 0.2L.min-1 (minimal gas flow)
Dräegersorb
D for DraegersorbTM group with gas flow at 2L.min-1 (classical sevoflurane administration) with the same ventilator Draeger A-500 Perseus.
Sevoflurane administration at 2L.min-1 (classical gas flow)

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
To show a significant decrease of at least 25% of the sevoflurane consumption
Time Frame: Intraoperative
To compare the total sevoflurane consumption when using MemsorbTM versus DrägersorbTM. This will be expressed in mL.kg-1.h-1 of surgery and the primary objective will focus on H1 of surgery, H1 starting at the time of incision (T0).
Intraoperative

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
To evaluate sevoflurane consumption for each hour of surgery
Time Frame: Intraoperative
To evaluate sevoflurane consumption in mL.kg-1.h-1 for each hour of surgery (H2, 3, 4 etc…) in laparoscopic abdominal surgeries of more than 1hour.
Intraoperative
To evaluate the pollution induced by anesthesia at H1 of surgery and consequent hours
Time Frame: Intraoperative
To evaluate the pollution induced by anesthesia in grams of CO2 when using MemsorbTM versus DrägersorbTM at H1 of surgery
Intraoperative
To evaluate the levels of expired CO2
Time Frame: Intraoperative
To evaluate the levels of expired CO2 analyzed by the Dräger Perseus A500 ventilator (electronic records per seconds) throughout the surgery when using MemsorbTM versus DrägersorbTM. Mean expired CO2 for each hour of surgery will be used in mmHg
Intraoperative
To evaluate the global costs of sevoflurane consumption
Time Frame: Intraoperative
To evaluate the global costs of sevoflurane consumption in both the groups and also the costs related to the use of the MemsorbTM versus DrägersorbTM. Will be given in CAN$
Intraoperative

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Philippe PR Richebe, MD, PhD, CIUSSS Est de l'île de Montréal

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 4, 2021

Primary Completion (Actual)

February 18, 2022

Study Completion (Actual)

February 18, 2022

Study Registration Dates

First Submitted

August 31, 2020

First Submitted That Met QC Criteria

August 31, 2020

First Posted (Actual)

September 7, 2020

Study Record Updates

Last Update Posted (Actual)

January 23, 2023

Last Update Submitted That Met QC Criteria

January 19, 2023

Last Verified

January 1, 2023

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.