Descending Aorta Blood Flow to Guide Fluid Therapy During Surgery

April 24, 2024 updated by: Torkjell Nostdahl, Sykehuset Telemark
The investigators will perform a clinical observational study of patients undergoing abdominal surgery. The study will be performed at Sykehuset Telemark, Skien. The aims of the study are to compare suprasternal and transoesophageal Doppler (reference method) to measure blood flow in the proximal descending aorta and to explore the variability of hemodynamic variables during surgery.

Study Overview

Detailed Description

Background and aim: The best validated method to estimate cardiac output with ultrasound requires measuring the Doppler velocity time integral (VTI) in the left ventricular outflow tract measured from the apical (lateral chest) window/ projection. However, this projection is often out of reach for anaesthesiologists during many surgical procedures (e.g. abdominal surgery) due to draping, and is often difficult to perform if the patient cannot be placed in the left lateral decubitus position. Alternatively, VTI in the proximal descending aorta may be measured via the suprasternal view, which is one of the standard projections in transthoracic echocardiography. This view is often available for anaesthesiologists during surgery on a patient in supine position. Assuming that changes in cardiac output are associated with changes in VTI in proximal descending aorta, this may provide the anaesthesiologist with a non-invasive measure of the response to a fluid challenge. The first aim of this study is to evaluate the ability of suprasternal Doppler to measure blood flow in the proximal descending aorta, and its agreement with the reference method oesophageal Doppler, during ongoing abdominal surgery.

Fluid challenges are typically given when stroke volume is reduced by 10-15%, which is considered a significant reduction. This is however dependent on the precision of the measurements, which is estimated during hemodynamic stability. During surgery, on the other hand, the variability may be larger, leading to erroneously concluding that stroke volume is reduced due to volume loss, when in fact this may be due to other stimuli related to surgery (mechanical, e.g. compression of large veins or the thoracic cavity, or stimuli affecting the autonomic nervous system, e.g. pain). The proportion of potentially falsely measured reduction may be calculated by performing repeated measurements during ongoing surgery. Thus, an other aim with this study is to explore the variability of measurements of stroke volume during surgery.

Study design: Clinical observational study.

Data collection: 30 patients scheduled to undergo laparoscopic or open abdominal surgical procedures (gastrointestinal or gynecological) of at least 1.5 hrs estimated duration. The patients should be of American Society of Anesthesiologists physical status 1-3 and have no contraindications to the use of oesophageal ultrasound. Hemodynamic data will be downloaded to a PC continuously from the clinical monitoring equipment (Philips Intellivue) using the VSCapture software (https://github.com/xeonfusion/VSCaptureMP) in VisualStudio (Microsoft). "Landmarks" in the procedure (e.g. surgical incision, manipulation in the surgical field and hemorrhage) will be recorded manually to relate the hemodynamic data to the different steps in the procedure. Blood flow velocity in the descending aorta will be measured using ultrasound machines in ordinary clinical use (GE Venue R 2.5; GE Healthcare). The Doppler recordings will be analyzed using commercially available software (EchoPAC; General Electric) after blinding.

Data analyzes:

  • Considerations regarding sample size for agreement analyzes: The investigators will focus on within-subject variability, and the main outcome will be the corresponding within-subject limits of agreement. If both methods have a precision of 2.9%, limits of agreement will be 8.3%. Given 18 measurements per subject (one measurement every 5 min over 1.5 hrs), 25 subjects will give an 80% power for an upper confidence interval of the limits of agreement of 18%. By including 30 patients, the investigators allow for some unsuccessful measurements due to e.g. poor echogenicity.
  • Considerations regarding sample size for variability analyzes: Given that the SD of oesophageal Doppler is 2.9%, the precision is 2.9%×1.96=5.7%. The probability of each evaluation of showing a false reduction is 0.8%. If surgery lasts 1.5 hrs with one evaluation every 5 min, the probability of at least one false reduction is 12%. If the precision is reduced with a doubled SD, the probability of at least one false reduction is 88%. Assuming a true SD of 2.9%; studying 25 subjects gives a power of 0.8 to detect a SD <3.0%.

Study Type

Observational

Enrollment (Actual)

30

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Vestfold Og Telemark
      • Skien, Vestfold Og Telemark, Norway, 3710
        • Telemark Hospital Trust

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Sampling Method

Probability Sample

Study Population

Patients that plan to undergo abdominal surgery (gastrointestinal or gynecological) at Telemark hospital, Skien, Norway.

Description

Inclusion Criteria:

  • Patients scheduled to undergo laparoscopic or open abdominal surgical procedures (gastrointestinal or gynecological) of at least 1.5 h estimated duration.
  • The patients should be of American Society of Anesthesiologists physical status 1-3

Exclusion Criteria:

  • Pathology in the larynx, oesophagus or stomach
  • Former surgery in larynx, oesophagus or stomach
  • Thoracic aortic aneurysm.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Patients undergoing abdominal surgery
Inclusion criteria: > 18 years old, American Society of Anesthesiologists physical status 1-3, abdominal surgery of > 1,5 hrs estimated duration, no contraindications to the use of oesophageal ultrasound.
While study subjects are in general anesthesia the investigator will perform repeated measurements every 5 min.
While study subjects are in general anesthesia the investigator will perform repeated measurements every 5 min.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in blood flow velocity in descending aorta
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Suprasternal Doppler
During surgery, every 5 minutes. At least over a time period of 1,5 hours
Change in blood flow velocity in descending aorta
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Transoesophageal Doppler
During surgery, every 5 minutes. At least over a time period of 1,5 hours
Change in stroke volume
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Calculated from the measurements by transoesophageal Doppler
During surgery, every 5 minutes. At least over a time period of 1,5 hours

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Change in heart rate
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Measured in beats per minute, collected from standard monitoring device during surgery
During surgery, every 5 minutes. At least over a time period of 1,5 hours
Change in systemic vascular resistance
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Measured in dynes/second/cm^5. Calculated from the measurements by transoesophageal Doppler and standard monitoring device (mean arterial pressure)
During surgery, every 5 minutes. At least over a time period of 1,5 hours
Change in blood pressure
Time Frame: During surgery, every 5 minutes. At least over a time period of 1,5 hours
Measured in mmHg, collected from standard monitoring device during surgery
During surgery, every 5 minutes. At least over a time period of 1,5 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Torkjell Nøstdahl, MD, PhD, Telemark Hospital Trust

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 18, 2022

Primary Completion (Actual)

January 29, 2024

Study Completion (Actual)

January 29, 2024

Study Registration Dates

First Submitted

December 20, 2021

First Submitted That Met QC Criteria

January 26, 2022

First Posted (Actual)

February 7, 2022

Study Record Updates

Last Update Posted (Actual)

April 25, 2024

Last Update Submitted That Met QC Criteria

April 24, 2024

Last Verified

April 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • 264472

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Hemodynamic Instability

Clinical Trials on Measurement of blood flow in descending aorta by suprasternal ultrasound Doppler technique

Subscribe