Functional Outcomes in CondUction System Pacing and Right Ventricular Synchrony (FOCUS-Right) (FOCUS-Right)

April 15, 2026 updated by: Virginia Commonwealth University

Functional Outcomes in CondUction System Pacing and Right Ventricular Synchrony The FOCUS-Right Pilot Study (FOCUS-Right)

The goal of this prospective single center clinical trial is to evaluate the safety and feasibility of performing simultaneous exercise stress cardiac magnetic resonance (CPET-CMR) and cardiopulmonary exercise testing in patients with pre-existing left bundle branch area pacemakers (LBBAP) programmed to an atrial sensing mode. Measurements of right ventricular, left ventricular function, and exercise capacity will be obtained at various LBBAP programming parameters at rest and during low intensity exercise. The main aims of the study are:

  • Demonstrate the safety and feasibility of performing CPET-CMR in patients with pre-existing LBBAP programmed to P-synchronous ventricular pacing mode.
  • Generate preliminary data evaluating differences in RV function, LV function, and exercise capacity during various pacemaker programming settings.

Study Overview

Detailed Description

Conduction system pacing, namely left bundle branch area pacing (LBBAP) represents an important pacing modality introduced to avoid the deleterious effects of non-physiologic ventricular activation during chronic right ventricular pacing and /or left bundle branch block (LBBBP). Previous studies have demonstrated that LBBAP exhibits similar or superior left ventricular resynchronization compared to conventional treatment modalities, the impact of this novel pacing strategy on the right ventricle has not yet been addressed. This study aims to demonstrate the safety and feasibility of assessing LBBAP performance with a focus on right ventricular function in a novel way utilizing Cardiopulmonary Exercise Testing and Exercise Stress Cardiac Magnetic Resonance.

The investigators plan a 1-day prospective non-randomized controlled pilot study in which up to 20 subjects with left bundle branch area pacemakers with evidence of output dependent anodal capture will undergo low intensity exercise CPET-CMR at three different LBBAP settings (underlying rhythm, non-selective left bundle pacing, non-selective left bundle pacing with anodal stimulation).

Study Type

Interventional

Enrollment (Estimated)

20

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • presence of dual chamber LBBAP implantation
  • presence of anodal capture with bipolar pacing configuration as determined at pacing implant.

Exclusion Criteria:

  • Known history of chronotropic incompetence
  • high degree or complete heart block
  • inability to tolerate NVP during initial device interrogation
  • pregnancy
  • inability to undergo MRI or exercise utilizing supine bicycle at target work-load
  • presence of MRI unsafe pacemaker components.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Non-Randomized
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: No Ventricular Pacing (NVP)
Assessment 1 (No Ventricular Pacing) will be performed with no ventricular pacing to assess baseline right and left ventricular function at rest and during exercise.
Prior to the CMR study, subjects will undergo baseline interrogation of their dual chamber LBBAP to assess for appropriate device function. Lead capture characteristics will then be assessed with continuous intracardiac electrogram characteristics recorded from the atrial and ventricular lead with simultaneous 12-lead ECG rhythm strip recordings to assess for settings at which the investigators can differentiate aLBBAP, sLBBAP and nsLBBAP. Pacing output will be increased until left bundle branch area pacing with anodal capture is achieved, and then decreased until anodal capture is lost. These programming parameters will be recorded and utilized for Phase 2-3 of the CPET-CMR scan. 12-lead ECG rhythm strips of QRS morphology with each parameter set will be obtained and saved for later analysis.
Following ECG and pacemaker analysis as above for safety, subjects will be fitted with 2 sets of MRI-compatible ECG monitoring systems a BP monitoring system , and a finger-tip pulse gating system. Imaging will be performed on a Magnetom Vida 3 Tesla scanner by trained CRM radiology technologists.
CPT will be established and supervised by a cardiac electrophysiologist to monitor for evidence of noise or artifact which may lead to inability to appropriately sense native atrial depolarization in DDD mode. Subjects with PM dependence are excluded. If noise is noted during scanning with NVP, then scanning will NOT be performed during P-synchronous pacing and will instead be performed during overdrive pacing at a rate above the patients maximal achieved heart rate with exertion during Phase 1. Thus, there is minimal risk to the subject from PM programming during CPET-CMR.
Active Comparator: Non-selective or Selective Left Bundle Branch Area Pacing (nsLBBAP)
Phase 2 (non-selective or selective Left Bundle Branch Area Pacing or ns/sLBBAP) will assess right and left ventricular function at rest and during exercise with non-selective or selective left bundle branch area pacing
Prior to the CMR study, subjects will undergo baseline interrogation of their dual chamber LBBAP to assess for appropriate device function. Lead capture characteristics will then be assessed with continuous intracardiac electrogram characteristics recorded from the atrial and ventricular lead with simultaneous 12-lead ECG rhythm strip recordings to assess for settings at which the investigators can differentiate aLBBAP, sLBBAP and nsLBBAP. Pacing output will be increased until left bundle branch area pacing with anodal capture is achieved, and then decreased until anodal capture is lost. These programming parameters will be recorded and utilized for Phase 2-3 of the CPET-CMR scan. 12-lead ECG rhythm strips of QRS morphology with each parameter set will be obtained and saved for later analysis.
Following ECG and pacemaker analysis as above for safety, subjects will be fitted with 2 sets of MRI-compatible ECG monitoring systems a BP monitoring system , and a finger-tip pulse gating system. Imaging will be performed on a Magnetom Vida 3 Tesla scanner by trained CRM radiology technologists.
CPT will be established and supervised by a cardiac electrophysiologist to monitor for evidence of noise or artifact which may lead to inability to appropriately sense native atrial depolarization in DDD mode. Subjects with PM dependence are excluded. If noise is noted during scanning with NVP, then scanning will NOT be performed during P-synchronous pacing and will instead be performed during overdrive pacing at a rate above the patients maximal achieved heart rate with exertion during Phase 1. Thus, there is minimal risk to the subject from PM programming during CPET-CMR.
Experimental: Left Bundle Branch Area Pacing with Anodal Capture (aLBBAP)
Phase 3 (Left Bundle Branch Area Pacing with Anodal Capture or aLBBAP) will assess right and left ventricular function at rest and during exercise with left bundle branch area pacing with anodal capture.
Prior to the CMR study, subjects will undergo baseline interrogation of their dual chamber LBBAP to assess for appropriate device function. Lead capture characteristics will then be assessed with continuous intracardiac electrogram characteristics recorded from the atrial and ventricular lead with simultaneous 12-lead ECG rhythm strip recordings to assess for settings at which the investigators can differentiate aLBBAP, sLBBAP and nsLBBAP. Pacing output will be increased until left bundle branch area pacing with anodal capture is achieved, and then decreased until anodal capture is lost. These programming parameters will be recorded and utilized for Phase 2-3 of the CPET-CMR scan. 12-lead ECG rhythm strips of QRS morphology with each parameter set will be obtained and saved for later analysis.
Following ECG and pacemaker analysis as above for safety, subjects will be fitted with 2 sets of MRI-compatible ECG monitoring systems a BP monitoring system , and a finger-tip pulse gating system. Imaging will be performed on a Magnetom Vida 3 Tesla scanner by trained CRM radiology technologists.
CPT will be established and supervised by a cardiac electrophysiologist to monitor for evidence of noise or artifact which may lead to inability to appropriately sense native atrial depolarization in DDD mode. Subjects with PM dependence are excluded. If noise is noted during scanning with NVP, then scanning will NOT be performed during P-synchronous pacing and will instead be performed during overdrive pacing at a rate above the patients maximal achieved heart rate with exertion during Phase 1. Thus, there is minimal risk to the subject from PM programming during CPET-CMR.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Incidence of Pacemaker Atrial (P Wave) Over-Sensed Events
Time Frame: Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Incidence of pacemaker atrial (P Wave) over sensing as derived from continuous pacemaker telemetry at optimal programmed atrial sensitivity.
Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Incidence of Pacemaker Atrial (P Wave) Under-Sensed Events
Time Frame: Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Incidence of pacemaker atrial (P Wave) under sensing as derived from continuous pacemaker telemetry at optimal programmed atrial sensitivity.
Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Left Ventricular Ejection Fraction (%)
Time Frame: Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Left ventricular ejection fraction (%) as measured by cardiac magnetic resonance imaging.
Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Right Ventricular Ejection Fraction (%)
Time Frame: Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Right ventricular ejection fraction (%) as measured by cardiac magnetic resonance imaging.
Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.

Other Outcome Measures

Outcome Measure
Measure Description
Time Frame
Slope of VO2 as a Function of Work Rate
Time Frame: Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.
Slope of VO2 as a function of work rate as quantified by cardiopulmonary exercise testing.
Day 1: From beginning of CPET-CMR imaging procedure to end of CPET-CMR imaging procedure on study day.

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Ajay Pillai, Virginia Commonwealth University

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

August 23, 2024

Primary Completion (Estimated)

May 1, 2026

Study Completion (Estimated)

May 1, 2026

Study Registration Dates

First Submitted

July 9, 2024

First Submitted That Met QC Criteria

September 13, 2024

First Posted (Actual)

September 19, 2024

Study Record Updates

Last Update Posted (Actual)

April 16, 2026

Last Update Submitted That Met QC Criteria

April 15, 2026

Last Verified

April 1, 2026

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.