Evaluación de la eficacia y seguridad de RuiXin-CoronaryAI para el diagnóstico de estenosis de las arterias coronarias

22 de noviembre de 2026 actualizado por: Shenzhen Raysight Intelligent Medical Technology Co., Ltd.

Evaluación de la eficacia y seguridad del software de análisis angiográfico por TC coronaria basado en IA (RuiXin-CoronaryAI) para el diagnóstico de estenosis de las arterias coronarias

With the emergence of advanced technology to date in the artificial intelligence (AI), computer aided diagnosis has gradually gained its popularity in the field of healthcare. Particularly, in the clinical practice of coronary artery disease diagnosis, the application of AI could be of great implication in alleviating the shortage of medical sources. To evaluate the effectiveness and safety of the AI-based coronary CT angiographic analysis software (RuiXin-CoronaryAI) for diagnosis of coronary artery stenosis, a retrospective, multi-center, cross-over designed, blinded, sensitivity superiority and specificity non-inferiority clinical trial will be conducted.

Descripción general del estudio

Estado

Terminado

Condiciones

Intervención / Tratamiento

Descripción detallada

Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA will be included. CCTA images of subjects should be of good quality up to the DICOM 3.0 standards, obtained by CT scan with ≥64-slices. The subjects with unqualified CTA will be excluded. CCTA images will be analyzed in three methods (3 groups). Control group: CCTA images will be visually evaluated by physicians. Experiment group: CCTA images will be evaluated by physicians using RuiXin-CoronaryAI. Reference group: CCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard. Primary outcomes are diagnostic sensitivity and specificity of RuiXin-CoronaryAI and coronary CTA for diagnosis of ischemia on a per-vessel basis. The effectiveness of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis will be conducted by testing superiority of diagnostic sensitivity and non-inferiority of specificity.

Tipo de estudio

De observación

Inscripción (Actual)

739

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Beijing Municipality
      • Beijing, Beijing Municipality, Porcelana, 100730
        • Beijing Hospital
    • Guangdong
      • Guangzhou, Guangdong, Porcelana, 510120
        • Sun Yat-sen Memorial Hospital
    • Hubei
      • Wuhan, Hubei, Porcelana, 430030
        • Tongji Hospital

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años y mayores (Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Todos

Método de muestreo

Muestra de probabilidad

Población de estudio

Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA examination.

Descripción

Inclusion Criteria:

  1. layer thickness of CCTA images should be less than 1mm, image quality should be up to DICOM 3.0 standards;
  2. vessels should be clearly developed, contrast medium ought to be well filled, the average of CT value of aortic root cavity should be between 325-600HU in CCTA image;
  3. remodeling of vessels should be intact, including coronary artery and branches, without missed or inaccurate slices;
  4. CCTA image should be obtained from single- or dual-source computed tomography (CT) scanners with a minimum of 64 detector rows.

Exclusion Criteria:

  1. CCTA image is of poor quality due to motion artifact, severe calcification, metal coverage, noise, poor contrast medium injection and other variables influencing the diagnosis of stenosis;
  2. previous percutaneous coronary intervention (PCI) or coronary artery bypass grafting (CABG);
  3. anomalous origin of coronary artery;
  4. other non-atherosclerosis-related coronary diseases like coronary artery fistula, aneurysm, coronary artery ectasia, arteritis coronaria, etc.;
  5. repeated enrollment;
  6. other conditions not suitable for enrollment.

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Modelos observacionales: COHORT
  • Perspectivas temporales: RETROSPECTIVE

Cohortes e Intervenciones

Grupo / Cohorte
Intervención / Tratamiento
Control group

CCTA images will be visually evaluated by physicians.

Experiment group

CCTA images will be evaluated by physicians using RuiXin-CoronaryAI.

RuiXin-CoronaryAI, based on Computed Tomography Angiography (CTA) and was independently designed by RaysightMed Inc., which has been already authorized by National Medical Products Administration (NMPA).

Reference group

CCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Per-vessel diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis

1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
Per-vessel diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis

1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Per-patient diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis

1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
Per-patient diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis

1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Patrocinador

Colaboradores

Investigadores

  • Director de estudio: Liming Xia, Tongji Hospital

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

26 de mayo de 2021

Finalización primaria (Actual)

6 de marzo de 2022

Finalización del estudio (Actual)

19 de noviembre de 2023

Fechas de registro del estudio

Enviado por primera vez

16 de marzo de 2022

Primero enviado que cumplió con los criterios de control de calidad

1 de abril de 2022

Publicado por primera vez (Actual)

11 de abril de 2022

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

24 de noviembre de 2026

Última actualización enviada que cumplió con los criterios de control de calidad

22 de noviembre de 2026

Última verificación

1 de noviembre de 2026

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • RuiXin-CoronaryAI

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .