Evaluación de la eficacia y seguridad de RuiXin-CoronaryAI para el diagnóstico de estenosis de las arterias coronarias
Evaluación de la eficacia y seguridad del software de análisis angiográfico por TC coronaria basado en IA (RuiXin-CoronaryAI) para el diagnóstico de estenosis de las arterias coronarias
With the emergence of advanced technology to date in the artificial intelligence (AI), computer aided diagnosis has gradually gained its popularity in the field of healthcare. Particularly, in the clinical practice of coronary artery disease diagnosis, the application of AI could be of great implication in alleviating the shortage of medical sources. To evaluate the effectiveness and safety of the AI-based coronary CT angiographic analysis software (RuiXin-CoronaryAI) for diagnosis of coronary artery stenosis, a retrospective, multi-center, cross-over designed, blinded, sensitivity superiority and specificity non-inferiority clinical trial will be conducted.
Descripción general del estudio
Estado
Estado
Condiciones
Condiciones
Intervención / Tratamiento
Intervención / Tratamiento
Descripción detallada
Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA will be included. CCTA images of subjects should be of good quality up to the DICOM 3.0 standards, obtained by CT scan with ≥64-slices. The subjects with unqualified CTA will be excluded. CCTA images will be analyzed in three methods (3 groups). Control group: CCTA images will be visually evaluated by physicians. Experiment group: CCTA images will be evaluated by physicians using RuiXin-CoronaryAI. Reference group: CCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard. Primary outcomes are diagnostic sensitivity and specificity of RuiXin-CoronaryAI and coronary CTA for diagnosis of ischemia on a per-vessel basis. The effectiveness of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis will be conducted by testing superiority of diagnostic sensitivity and non-inferiority of specificity.
Tipo de estudio
Tipo de estudio
Inscripción (Actual)
Inscripción
Contactos y Ubicaciones
Ubicaciones de estudio
-
-
Beijing Municipality
-
Beijing, Beijing Municipality, Porcelana, 100730
- Beijing Hospital
-
-
Guangdong
-
Guangzhou, Guangdong, Porcelana, 510120
- Sun Yat-sen Memorial Hospital
-
-
Hubei
-
Wuhan, Hubei, Porcelana, 430030
- Tongji Hospital
-
-
Criterios de participación
Criterio de elegibilidad
Criterio de elegibilidad
Edades elegibles para estudiar
Acepta Voluntarios Saludables
Géneros elegibles para el estudio
Método de muestreo
Población de estudio
Patients ≥18 years old with suspected or known coronary artery disease who underwent CCTA examination.
Descripción
Inclusion Criteria:
- layer thickness of CCTA images should be less than 1mm, image quality should be up to DICOM 3.0 standards;
- vessels should be clearly developed, contrast medium ought to be well filled, the average of CT value of aortic root cavity should be between 325-600HU in CCTA image;
- remodeling of vessels should be intact, including coronary artery and branches, without missed or inaccurate slices;
- CCTA image should be obtained from single- or dual-source computed tomography (CT) scanners with a minimum of 64 detector rows.
Exclusion Criteria:
- CCTA image is of poor quality due to motion artifact, severe calcification, metal coverage, noise, poor contrast medium injection and other variables influencing the diagnosis of stenosis;
- previous percutaneous coronary intervention (PCI) or coronary artery bypass grafting (CABG);
- anomalous origin of coronary artery;
- other non-atherosclerosis-related coronary diseases like coronary artery fistula, aneurysm, coronary artery ectasia, arteritis coronaria, etc.;
- repeated enrollment;
- other conditions not suitable for enrollment.
Plan de estudios
¿Cómo está diseñado el estudio?
Detalles de diseño
- Modelos observacionales: COHORT
- Perspectivas temporales: RETROSPECTIVE
Número de grupos/cohortes
Cohortes e Intervenciones
Grupo / CohorteGrupo / Cohorte |
Intervención / TratamientoIntervención / Tratamiento |
|---|---|
|
Control group
CCTA images will be visually evaluated by physicians. |
|
|
Experiment group
CCTA images will be evaluated by physicians using RuiXin-CoronaryAI. |
RuiXin-CoronaryAI, based on Computed Tomography Angiography (CTA) and was independently designed by RaysightMed Inc., which has been already authorized by National Medical Products Administration (NMPA). |
|
Reference group
CCTA images will be visually evaluated by cardiologists with at least 10 years experiences, and the conclusions they offer will be used as golden standard. |
¿Qué mide el estudio?
Medidas de resultado primarias
Medidas de resultado primarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Per-vessel diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis |
1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
|
Per-vessel diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-vessel basis |
1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Medidas de resultado secundarias
Medidas de resultado secundarias
Medida de resultado |
Medida Descripción |
Periodo de tiempo |
|---|---|---|
|
Per-patient diagnostic sensitivity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis |
1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
|
Per-patient diagnostic specificity of RuiXin-CoronaryAI for diagnosis of coronary artery stenosis
Periodo de tiempo: 1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Outcome measures were comparing RuiXin-CoronaryAI to CTA on a per-patient basis |
1 day; Incident time for CTA examination was dependent on the length of time on the CT scaner. RuiXin-CoronaryAI examination was done remotely at Raysight's processing center in Shenzhen with a turnaround time of 24 hours from CT scan.
|
Colaboradores e Investigadores
Patrocinador
Patrocinador
Colaboradores
Colaboradores
Investigadores
Investigadores
- Director de estudio: Liming Xia, Tongji Hospital
Fechas de registro del estudio
Fechas importantes del estudio
Inicio del estudio (Actual)
Inicio del estudio
Finalización primaria (Actual)
Finalización primaria
Finalización del estudio (Actual)
Finalización del estudio
Fechas de registro del estudio
Enviado por primera vez
Enviado por primera vez
Primero enviado que cumplió con los criterios de control de calidad
Primero enviado que cumplió con los criterios de control de calidad
Publicado por primera vez (Actual)
Publicado por primera vez
Actualizaciones de registros de estudio
Última actualización publicada (Actual)
Última actualización publicada
Última actualización enviada que cumplió con los criterios de control de calidad
Última actualización enviada que cumplió con los criterios de control de calidad
Última verificación
Última verificación
Más información
Términos relacionados con este estudio
Términos MeSH relevantes adicionales
Otros números de identificación del estudio
Otros números de identificación del estudio
- RuiXin-CoronaryAI
Información sobre medicamentos y dispositivos, documentos del estudio
Estudia un producto farmacéutico regulado por la FDA de EE. UU.
Estudia un producto de dispositivo regulado por la FDA de EE. UU.
Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .