Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data

tiistai 22. tammikuuta 2019 päivittänyt: MC10 Inc.

A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings

Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.

Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.

Tutkimuksen yleiskatsaus

Opintotyyppi

Interventio

Ilmoittautuminen (Todellinen)

30

Vaihe

  • Ei sovellettavissa

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskelupaikat

    • Ohio
      • Cincinnati, Ohio, Yhdysvallat, 45212
        • CTI Clinical Research Center

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

18 vuotta - 99 vuotta (Aikuinen, Vanhempi Aikuinen)

Hyväksyy terveitä vapaaehtoisia

Joo

Sukupuolet, jotka voivat opiskella

Kaikki

Kuvaus

Inclusion Criteria:

  1. Male or female volunteers, at least 18 years of age at the time of screening visit;
  2. Fluent in English;
  3. The subject is willing to comply with the protocol specified evaluations;
  4. Subject is willing and cognitively able to sign informed consent

Exclusion Criteria:

  1. Pregnancy;
  2. Positive urine drug screen
  3. History of active (clinically significant) skin disorders;
  4. History of allergic response to silicones or adhesives;
  5. Subjects with electronic implants of any kind (e.g. pacemaker)
  6. History of sleep disorders or self-reported insomnia or other sleep conditions;
  7. Broken, damaged or irritated skin or rashes near the sensor application sites;
  8. Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
  9. Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Muut
  • Jako: Ei käytössä
  • Inventiomalli: Yksittäinen ryhmätehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Muut: Single Arm

This single arm consists of all subjects which will interact with the device under investigation as well as comparator devices.

This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention.

Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.
FDA cleared reference device that monitors heart rate and heart rate variability in subjects
FDA cleared reference device that monitors respiration rate in subjects
Visual annotation of subject posture and other activities used for reference

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device
Aikaikkuna: 2 days after informed consent
Heart rate as measured by the subject device in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device
Aikaikkuna: 2 days after informed consent
Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device
Aikaikkuna: 2 days after informed consent
HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device
Aikaikkuna: 2 days after informed consent
Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Activity Classification as Compared to Visual Annotation
Aikaikkuna: 2 days after informed consent
The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer. The device's Activity Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.
2 days after informed consent
Accuracy of Step Count Compared to an Observer's Manual Count
Aikaikkuna: 2 days after informed consent
The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.
2 days after informed consent
Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation
Aikaikkuna: 2 days after informed consent
Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times. The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.
2 days after informed consent
Accuracy of Posture Classification as Compared to Visual Annotation
Aikaikkuna: 2 days after informed consent
The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer. The device's Posture Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.
2 days after informed consent
Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale
Aikaikkuna: 2 days after informed consent
The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject). Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion. The percentage of sensors which had acceptable adhesion at sensor removal is presented.
2 days after informed consent

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Sponsori

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Maanantai 31. heinäkuuta 2017

Ensisijainen valmistuminen (Todellinen)

Perjantai 25. elokuuta 2017

Opintojen valmistuminen (Todellinen)

Keskiviikko 30. elokuuta 2017

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Keskiviikko 16. elokuuta 2017

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Torstai 17. elokuuta 2017

Ensimmäinen Lähetetty (Todellinen)

Tiistai 22. elokuuta 2017

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Perjantai 25. tammikuuta 2019

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Tiistai 22. tammikuuta 2019

Viimeksi vahvistettu

Tiistai 1. tammikuuta 2019

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muut tutkimustunnusnumerot

  • MC10-PTL-103

Yksittäisten osallistujien tietojen suunnitelma (IPD)

Aiotko jakaa yksittäisten osallistujien tietoja (IPD)?

EI

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Joo

Yhdysvalloissa valmistettu ja sieltä viety tuote

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .

Kliiniset tutkimukset Terveet normaalit

Kliiniset tutkimukset Physiological signal monitor

Tilaa