Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data

22 gennaio 2019 aggiornato da: MC10 Inc.

A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings

Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.

Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.

Panoramica dello studio

Tipo di studio

Interventistico

Iscrizione (Effettivo)

30

Fase

  • Non applicabile

Contatti e Sedi

Questa sezione fornisce i recapiti di coloro che conducono lo studio e informazioni su dove viene condotto lo studio.

Luoghi di studio

    • Ohio
      • Cincinnati, Ohio, Stati Uniti, 45212
        • CTI Clinical Research Center

Criteri di partecipazione

I ricercatori cercano persone che corrispondano a una certa descrizione, chiamata criteri di ammissibilità. Alcuni esempi di questi criteri sono le condizioni generali di salute di una persona o trattamenti precedenti.

Criteri di ammissibilità

Età idonea allo studio

Da 18 anni a 99 anni (Adulto, Adulto più anziano)

Accetta volontari sani

Sessi ammissibili allo studio

Tutto

Descrizione

Inclusion Criteria:

  1. Male or female volunteers, at least 18 years of age at the time of screening visit;
  2. Fluent in English;
  3. The subject is willing to comply with the protocol specified evaluations;
  4. Subject is willing and cognitively able to sign informed consent

Exclusion Criteria:

  1. Pregnancy;
  2. Positive urine drug screen
  3. History of active (clinically significant) skin disorders;
  4. History of allergic response to silicones or adhesives;
  5. Subjects with electronic implants of any kind (e.g. pacemaker)
  6. History of sleep disorders or self-reported insomnia or other sleep conditions;
  7. Broken, damaged or irritated skin or rashes near the sensor application sites;
  8. Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
  9. Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.

Piano di studio

Questa sezione fornisce i dettagli del piano di studio, compreso il modo in cui lo studio è progettato e ciò che lo studio sta misurando.

Come è strutturato lo studio?

Dettagli di progettazione

  • Scopo principale: Altro
  • Assegnazione: N / A
  • Modello interventistico: Assegnazione di gruppo singolo
  • Mascheramento: Nessuno (etichetta aperta)

Armi e interventi

Gruppo di partecipanti / Arm
Intervento / Trattamento
Altro: Single Arm

This single arm consists of all subjects which will interact with the device under investigation as well as comparator devices.

This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention.

Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.
FDA cleared reference device that monitors heart rate and heart rate variability in subjects
FDA cleared reference device that monitors respiration rate in subjects
Visual annotation of subject posture and other activities used for reference

Cosa sta misurando lo studio?

Misure di risultato primarie

Misura del risultato
Misura Descrizione
Lasso di tempo
Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device
Lasso di tempo: 2 days after informed consent
Heart rate as measured by the subject device in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device
Lasso di tempo: 2 days after informed consent
Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device
Lasso di tempo: 2 days after informed consent
HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device
Lasso di tempo: 2 days after informed consent
Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Activity Classification as Compared to Visual Annotation
Lasso di tempo: 2 days after informed consent
The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer. The device's Activity Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.
2 days after informed consent
Accuracy of Step Count Compared to an Observer's Manual Count
Lasso di tempo: 2 days after informed consent
The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.
2 days after informed consent
Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation
Lasso di tempo: 2 days after informed consent
Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times. The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.
2 days after informed consent
Accuracy of Posture Classification as Compared to Visual Annotation
Lasso di tempo: 2 days after informed consent
The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer. The device's Posture Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.
2 days after informed consent
Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale
Lasso di tempo: 2 days after informed consent
The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject). Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion. The percentage of sensors which had acceptable adhesion at sensor removal is presented.
2 days after informed consent

Collaboratori e investigatori

Qui è dove troverai le persone e le organizzazioni coinvolte in questo studio.

Sponsor

Studiare le date dei record

Queste date tengono traccia dell'avanzamento della registrazione dello studio e dell'invio dei risultati di sintesi a ClinicalTrials.gov. I record degli studi e i risultati riportati vengono esaminati dalla National Library of Medicine (NLM) per assicurarsi che soddisfino specifici standard di controllo della qualità prima di essere pubblicati sul sito Web pubblico.

Studia le date principali

Inizio studio (Effettivo)

31 luglio 2017

Completamento primario (Effettivo)

25 agosto 2017

Completamento dello studio (Effettivo)

30 agosto 2017

Date di iscrizione allo studio

Primo inviato

16 agosto 2017

Primo inviato che soddisfa i criteri di controllo qualità

17 agosto 2017

Primo Inserito (Effettivo)

22 agosto 2017

Aggiornamenti dei record di studio

Ultimo aggiornamento pubblicato (Effettivo)

25 gennaio 2019

Ultimo aggiornamento inviato che soddisfa i criteri QC

22 gennaio 2019

Ultimo verificato

1 gennaio 2019

Maggiori informazioni

Termini relativi a questo studio

Altri numeri di identificazione dello studio

  • MC10-PTL-103

Piano per i dati dei singoli partecipanti (IPD)

Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?

NO

Informazioni su farmaci e dispositivi, documenti di studio

Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti

No

Studia un dispositivo regolamentato dalla FDA degli Stati Uniti

prodotto fabbricato ed esportato dagli Stati Uniti

No

Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .

Prove cliniche su Physiological signal monitor

Sottoscrivi