Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data
A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings
Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.
Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.
Studieoversigt
Status
Betingelser
Undersøgelsestype
Tilmelding (Faktiske)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
Ohio
-
Cincinnati, Ohio, Forenede Stater, 45212
- CTI Clinical Research Center
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Beskrivelse
Inclusion Criteria:
- Male or female volunteers, at least 18 years of age at the time of screening visit;
- Fluent in English;
- The subject is willing to comply with the protocol specified evaluations;
- Subject is willing and cognitively able to sign informed consent
Exclusion Criteria:
- Pregnancy;
- Positive urine drug screen
- History of active (clinically significant) skin disorders;
- History of allergic response to silicones or adhesives;
- Subjects with electronic implants of any kind (e.g. pacemaker)
- History of sleep disorders or self-reported insomnia or other sleep conditions;
- Broken, damaged or irritated skin or rashes near the sensor application sites;
- Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
- Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Andet
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Andet: Single Arm
This single arm consists of all subjects which will interact with the device under investigation as well as comparator devices. This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention. |
Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.
FDA cleared reference device that monitors heart rate and heart rate variability in subjects
FDA cleared reference device that monitors respiration rate in subjects
Visual annotation of subject posture and other activities used for reference
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device
Tidsramme: 2 days after informed consent
|
Heart rate as measured by the subject device in beats per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device
Tidsramme: 2 days after informed consent
|
Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device
Tidsramme: 2 days after informed consent
|
HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device
Tidsramme: 2 days after informed consent
|
Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Activity Classification as Compared to Visual Annotation
Tidsramme: 2 days after informed consent
|
The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer.
The device's Activity Classification will be compared to the observer's annotation.
The percentage of correct classifications by the device against indicated visual observation is presented.
The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.
|
2 days after informed consent
|
|
Accuracy of Step Count Compared to an Observer's Manual Count
Tidsramme: 2 days after informed consent
|
The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.
|
2 days after informed consent
|
|
Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation
Tidsramme: 2 days after informed consent
|
Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times.
The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.
|
2 days after informed consent
|
|
Accuracy of Posture Classification as Compared to Visual Annotation
Tidsramme: 2 days after informed consent
|
The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer.
The device's Posture Classification will be compared to the observer's annotation.
The percentage of correct classifications by the device against indicated visual observation is presented.
The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.
|
2 days after informed consent
|
|
Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale
Tidsramme: 2 days after informed consent
|
The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject).
Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion.
The percentage of sensors which had acceptable adhesion at sensor removal is presented.
|
2 days after informed consent
|
Samarbejdspartnere og efterforskere
Sponsor
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Faktiske)
Studieafslutning (Faktiske)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Andre undersøgelses-id-numre
- MC10-PTL-103
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
produkt fremstillet i og eksporteret fra U.S.A.
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Sunde normaler
-
Assiut UniversityIkke rekrutterer endnuNormal arbejdskraft
-
Massachusetts General HospitalRekrutteringNormal fysiologiForenede Stater
-
National Institute of Environmental Health Sciences...Rekruttering
-
Assistance Publique - Hôpitaux de ParisURC Necker Cochin, FranceRekrutteringNormal graviditetFrankrig
-
National Institute of Neurological Disorders and...RekrutteringNormal fysiologiForenede Stater
-
National Institute of Neurological Disorders and...AfsluttetNormal fysiologiForenede Stater
-
National Institute of Neurological Disorders and...AfsluttetNormal fysiologiForenede Stater
Kliniske forsøg med Physiological signal monitor
-
Dana-Farber Cancer InstituteNational Institute of Nursing Research (NINR)Trukket tilbageLivmoderhalskræft | Gynækologisk kræft | Livmoderhalskræft | Æggelederkræft | Primært peritonealt karcinom | LivmoderkræftForenede Stater
-
Promedius IncRekrutteringOsteoporose, OsteopeniDet Forenede Kongerige, Holland
-
Luca ValciAfsluttetCerebral aneurismeSchweiz
-
UMC UtrechtAfsluttet
-
Western University, CanadaIkke rekrutterer endnuParkinsons sygdomCanada
-
University Hospital, CaenRekrutteringHjertesygdom, lungesygdomFrankrig
-
Signal ReliefProxima CRO; ITC Imaging, LLCAfsluttet
-
Tidal Medical TechnologiesRekruttering
-
Hebrew University of JerusalemTilmelding efter invitationFødevarerespons | Madstop | Food Response-hæmningIsrael
-
Bio-Signal Group Corp.National Institute of Neurological Disorders and Stroke (NINDS); State... og andre samarbejdspartnereAfsluttet