Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data

22. ledna 2019 aktualizováno: MC10 Inc.

A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings

Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.

Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.

Přehled studie

Typ studie

Intervenční

Zápis (Aktuální)

30

Fáze

  • Nelze použít

Kontakty a umístění

Tato část poskytuje kontaktní údaje pro ty, kteří studii provádějí, a informace o tom, kde se tato studie provádí.

Studijní místa

    • Ohio
      • Cincinnati, Ohio, Spojené státy, 45212
        • CTI Clinical Research Center

Kritéria účasti

Výzkumníci hledají lidi, kteří odpovídají určitému popisu, kterému se říká kritéria způsobilosti. Některé příklady těchto kritérií jsou celkový zdravotní stav osoby nebo předchozí léčba.

Kritéria způsobilosti

Věk způsobilý ke studiu

18 let až 99 let (Dospělý, Starší dospělý)

Přijímá zdravé dobrovolníky

Ano

Pohlaví způsobilá ke studiu

Všechno

Popis

Inclusion Criteria:

  1. Male or female volunteers, at least 18 years of age at the time of screening visit;
  2. Fluent in English;
  3. The subject is willing to comply with the protocol specified evaluations;
  4. Subject is willing and cognitively able to sign informed consent

Exclusion Criteria:

  1. Pregnancy;
  2. Positive urine drug screen
  3. History of active (clinically significant) skin disorders;
  4. History of allergic response to silicones or adhesives;
  5. Subjects with electronic implants of any kind (e.g. pacemaker)
  6. History of sleep disorders or self-reported insomnia or other sleep conditions;
  7. Broken, damaged or irritated skin or rashes near the sensor application sites;
  8. Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
  9. Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.

Studijní plán

Tato část poskytuje podrobnosti o studijním plánu, včetně toho, jak je studie navržena a co studie měří.

Jak je studie koncipována?

Detaily designu

  • Primární účel: Jiný
  • Přidělení: N/A
  • Intervenční model: Přiřazení jedné skupiny
  • Maskování: Žádné (otevřený štítek)

Zbraně a zásahy

Skupina účastníků / Arm
Intervence / Léčba
Jiný: Single Arm

This single arm consists of all subjects which will interact with the device under investigation as well as comparator devices.

This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention.

Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.
FDA cleared reference device that monitors heart rate and heart rate variability in subjects
FDA cleared reference device that monitors respiration rate in subjects
Visual annotation of subject posture and other activities used for reference

Co je měření studie?

Primární výstupní opatření

Měření výsledku
Popis opatření
Časové okno
Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device
Časové okno: 2 days after informed consent
Heart rate as measured by the subject device in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device
Časové okno: 2 days after informed consent
Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device
Časové okno: 2 days after informed consent
HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device
Časové okno: 2 days after informed consent
Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.
2 days after informed consent
Accuracy of Activity Classification as Compared to Visual Annotation
Časové okno: 2 days after informed consent
The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer. The device's Activity Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.
2 days after informed consent
Accuracy of Step Count Compared to an Observer's Manual Count
Časové okno: 2 days after informed consent
The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.
2 days after informed consent
Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation
Časové okno: 2 days after informed consent
Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times. The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.
2 days after informed consent
Accuracy of Posture Classification as Compared to Visual Annotation
Časové okno: 2 days after informed consent
The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer. The device's Posture Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.
2 days after informed consent
Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale
Časové okno: 2 days after informed consent
The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject). Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion. The percentage of sensors which had acceptable adhesion at sensor removal is presented.
2 days after informed consent

Spolupracovníci a vyšetřovatelé

Zde najdete lidi a organizace zapojené do této studie.

Sponzor

Termíny studijních záznamů

Tato data sledují průběh záznamů studie a předkládání souhrnných výsledků na ClinicalTrials.gov. Záznamy ze studií a hlášené výsledky jsou před zveřejněním na veřejné webové stránce přezkoumány Národní lékařskou knihovnou (NLM), aby se ujistily, že splňují specifické standardy kontroly kvality.

Hlavní termíny studia

Začátek studia (Aktuální)

31. července 2017

Primární dokončení (Aktuální)

25. srpna 2017

Dokončení studie (Aktuální)

30. srpna 2017

Termíny zápisu do studia

První předloženo

16. srpna 2017

První předloženo, které splnilo kritéria kontroly kvality

17. srpna 2017

První zveřejněno (Aktuální)

22. srpna 2017

Aktualizace studijních záznamů

Poslední zveřejněná aktualizace (Aktuální)

25. ledna 2019

Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality

22. ledna 2019

Naposledy ověřeno

1. ledna 2019

Více informací

Termíny související s touto studií

Další identifikační čísla studie

  • MC10-PTL-103

Plán pro data jednotlivých účastníků (IPD)

Plánujete sdílet data jednotlivých účastníků (IPD)?

NE

Informace o lécích a zařízeních, studijní dokumenty

Studuje lékový produkt regulovaný americkým FDA

Ne

Studuje produkt zařízení regulovaný americkým úřadem FDA

Ano

produkt vyrobený a vyvážený z USA

Ne

Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .