Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data
A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings
Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.
Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.
Visão geral do estudo
Status
Condições
Tipo de estudo
Inscrição (Real)
Estágio
- Não aplicável
Contactos e Locais
Locais de estudo
-
-
Ohio
-
Cincinnati, Ohio, Estados Unidos, 45212
- CTI Clinical Research Center
-
-
Critérios de participação
Critérios de elegibilidade
Idades elegíveis para estudo
Aceita Voluntários Saudáveis
Gêneros Elegíveis para o Estudo
Descrição
Inclusion Criteria:
- Male or female volunteers, at least 18 years of age at the time of screening visit;
- Fluent in English;
- The subject is willing to comply with the protocol specified evaluations;
- Subject is willing and cognitively able to sign informed consent
Exclusion Criteria:
- Pregnancy;
- Positive urine drug screen
- History of active (clinically significant) skin disorders;
- History of allergic response to silicones or adhesives;
- Subjects with electronic implants of any kind (e.g. pacemaker)
- History of sleep disorders or self-reported insomnia or other sleep conditions;
- Broken, damaged or irritated skin or rashes near the sensor application sites;
- Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
- Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.
Plano de estudo
Como o estudo é projetado?
Detalhes do projeto
- Finalidade Principal: Outro
- Alocação: N / D
- Modelo Intervencional: Atribuição de grupo único
- Mascaramento: Nenhum (rótulo aberto)
Armas e Intervenções
Grupo de Participantes / Braço |
Intervenção / Tratamento |
|---|---|
|
Outro: Single Arm
This single arm consists of all subjects which will interact with the device under investigation as well as comparator devices. This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention. |
Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.
FDA cleared reference device that monitors heart rate and heart rate variability in subjects
FDA cleared reference device that monitors respiration rate in subjects
Visual annotation of subject posture and other activities used for reference
|
O que o estudo está medindo?
Medidas de resultados primários
Medida de resultado |
Descrição da medida |
Prazo |
|---|---|---|
|
Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device
Prazo: 2 days after informed consent
|
Heart rate as measured by the subject device in beats per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device
Prazo: 2 days after informed consent
|
Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device
Prazo: 2 days after informed consent
|
HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device
Prazo: 2 days after informed consent
|
Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device.
The Mean Absolute Error (MAE) between measurements from each device is presented.
|
2 days after informed consent
|
|
Accuracy of Activity Classification as Compared to Visual Annotation
Prazo: 2 days after informed consent
|
The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer.
The device's Activity Classification will be compared to the observer's annotation.
The percentage of correct classifications by the device against indicated visual observation is presented.
The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.
|
2 days after informed consent
|
|
Accuracy of Step Count Compared to an Observer's Manual Count
Prazo: 2 days after informed consent
|
The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.
|
2 days after informed consent
|
|
Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation
Prazo: 2 days after informed consent
|
Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times.
The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.
|
2 days after informed consent
|
|
Accuracy of Posture Classification as Compared to Visual Annotation
Prazo: 2 days after informed consent
|
The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer.
The device's Posture Classification will be compared to the observer's annotation.
The percentage of correct classifications by the device against indicated visual observation is presented.
The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.
|
2 days after informed consent
|
|
Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale
Prazo: 2 days after informed consent
|
The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject).
Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion.
The percentage of sensors which had acceptable adhesion at sensor removal is presented.
|
2 days after informed consent
|
Colaboradores e Investigadores
Patrocinador
Datas de registro do estudo
Datas Principais do Estudo
Início do estudo (Real)
Conclusão Primária (Real)
Conclusão do estudo (Real)
Datas de inscrição no estudo
Enviado pela primeira vez
Enviado pela primeira vez que atendeu aos critérios de CQ
Primeira postagem (Real)
Atualizações de registro de estudo
Última Atualização Postada (Real)
Última atualização enviada que atendeu aos critérios de controle de qualidade
Última verificação
Mais Informações
Termos relacionados a este estudo
Outros números de identificação do estudo
- MC10-PTL-103
Plano para dados de participantes individuais (IPD)
Planeja compartilhar dados de participantes individuais (IPD)?
Informações sobre medicamentos e dispositivos, documentos de estudo
Estuda um medicamento regulamentado pela FDA dos EUA
Estuda um produto de dispositivo regulamentado pela FDA dos EUA
produto fabricado e exportado dos EUA
Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .
Ensaios clínicos em Normais Saudáveis
-
Mahidol UniversityConcluído
-
Umeå UniversityConcluídoHidrocefalia de pressão normal idiopática
-
University Hospital, Basel, SwitzerlandRecrutamentoHidrocefalia de pressão normal idiopática (INPH)Suíça
-
DeepsonbioRecrutamentoHidrocefalia de pressão normal idiopática (iNPH)Republica da Coréia
-
Uppsala University HospitalUppsala University; Swedish Society for Medical ResearchConcluídoHidrocefalia de pressão normal idiopática (INPH)Suécia
-
Johns Hopkins UniversityUniversity of Utah; Integra LifeSciences CorporationConcluídoHidrocefalia de pressão normal idiopática (INPH)Estados Unidos, Canadá, Suécia
-
Assiut UniversityAinda não está recrutando
-
Massachusetts General HospitalRecrutamentoFisiologia NormalEstados Unidos
Ensaios clínicos em Physiological signal monitor
-
Bio-Signal Group Corp.National Institute of Neurological Disorders and Stroke (NINDS); State University... e outros colaboradoresConcluídoEstado mental alteradoEstados Unidos
-
National Institute of Geriatrics, Rheumatology...RecrutamentoFibromialgia | Síndrome da Fadiga Crônica (SFC)Polônia
-
Mayo ClinicSnap40 Ltd.ConcluídoMonitoramento remotoEstados Unidos
-
Marmara UniversityConcluídoMétodos de imagem da placa dentáriaPeru
-
Stryker InstrumentsConcluídoAnestesia, GeralEstados Unidos
-
Weill Medical College of Cornell UniversityWeill Cornell Medical College in QatarRetirado
-
William Marsh Rice UniversityKamuzu University of Health SciencesRecrutamento
-
Exempla Saint Joseph HospitalRescindidoHipertensão | Diabetes MellitusEstados Unidos
-
Rush University Medical CenterNational Institute of Nursing Research (NINR); University of Illinois at Chicago e outros colaboradoresConcluídoAtividade físicaEstados Unidos
-
Kuopio University HospitalConcluídoCirurgia cardíacaFinlândia