- ICH GCP
- USA klinikai vizsgálatok nyilvántartása
- Klinikai vizsgálat NCT01896115
CUSTOM-DBS: Current Steering to Optimize Deep Brain Stimulation (CUSTOM-DBS)
Current Steering to Optimize Deep Brain Stimulation
A tanulmány áttekintése
Részletes leírás
The study is a prospective, multi-center, double-blind, randomized controlled trial.
This study will compare various program settings for the bilateral stimulation of the STN using the BSC implantable Vercise™ DBS System for the treatment of levodopa-responsive, moderate to severe idiopathic PD.
Tanulmány típusa
Beiratkozás (Tényleges)
Fázis
- Nem alkalmazható
Kapcsolatok és helyek
Tanulmányi helyek
-
-
-
Innsbruck, Ausztria
- A.o. LKH Univ.-Kliniken Innsbruck
-
-
-
-
-
Bristol, Egyesült Királyság
- Southmead Hospital Bristol
-
London, Egyesült Királyság
- Charing Cross Hospital
-
-
-
-
California
-
Valencia, California, Egyesült Államok, 91355
- Boston Scientific Clinical Research Information Toll Free Number
-
-
-
-
-
Berlin, Németország
- University Berlin, Charite Virchow
-
Düsseldorf, Németország
- Universitaetsklinikum Dusseldorf
-
Kiel, Németország
- Universitätsklinikum Campus Kiel
-
Köln, Németország
- Uniklinik Köln
-
Wurzburg, Németország
- Universitaetsklinikum Wuerzburg
-
-
-
-
-
Udine, Olaszország
- Osp. S. Maria Della Misericordia
-
-
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Leírás
Inclusion Criteria:
- Implanted bilaterally in STN with a Vercise DBS system for Parkinson's disease for at least three months with programming optimized according to standard of care.
- UPDRS subset III score of ≥30 in the in the pre-operative meds off state.
- DBS must improve PD symptoms by ≥30% in the meds off state, as measured by UPDRS subset III score.
- Medical and mental fitness to comply with programming visit and study related procedures.
- Able to understand the study requirements and the treatment procedures and provides written informed consent before any study-specific tests or procedures are performed.
- Off symptom rigidity score of ≥2 in the evaluated arm as determined by the UPDRS-III.
Exclusion Criteria:
- Have any significant medical condition that is likely to interfere with study procedures or likely to confound evaluation of study endpoints, including any terminal illness with survival <12 months.
- Resting and/or action tremor score of ≥3 on the evaluated side as determined by the UPDRS-III in pre-operative meds off condition.
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
- Elsődleges cél: Alapvető tudomány
- Kiosztás: Véletlenszerűsített
- Beavatkozó modell: Crossover kiosztás
- Maszkolás: Kettős
Fegyverek és beavatkozások
Résztvevő csoport / kar |
Beavatkozás / kezelés |
---|---|
Kísérleti: Short PW
Patients with a Vercise DBS system programmed to 30 microseconds pulse width
|
Deep brain stimulation set at short pulse width and monopolar stimulation using current steering
|
Kísérleti: Conventional PW
Patients with a Vercise DBS system programmed to 60 microseconds pulse width
|
Deep brain stimulation set at short pulse width and monopolar stimulation using current steering
|
Kísérleti: Ventral current steering
Patients with a Vercise DBS system programmed to steer current ventrally
|
Deep brain stimulation set at short pulse width and monopolar stimulation using current steering
|
Kísérleti: Dorsal current steering
Patients with a Vercise DBS system programmed to steer current dorsally.
|
Deep brain stimulation set at short pulse width and monopolar stimulation using current steering
|
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
---|---|---|
Therapeutic Window
Időkeret: Day 1 programming visit
|
The therapeutic window refers to the range of stimulus amplitudes that provide a therapeutic effect without side effects.
In other words, it is the amplitude difference between the first stimulation-induced side effect threshold (e.g., eye deviation, muscle contraction, and speech) and full rigidity control threshold at 60 µs and 30 µs pulse width DBS settings.
|
Day 1 programming visit
|
Unified Parkinson's Disease Rating Scale III
Időkeret: Day 1 programming visit
|
The Unified Parkinson's Disease Rating Scale (UPDRS) has four sections (I-IV) that ask patients to rate aspects of their mental state including mood (I), aspects of daily activities (II), aspects of motor function (III), and complications of treatment (IV). Here, we ask subjects to rate their motor function (UPDRS III) following interventions with 30 µs and 60 µs pulse width DBS settings. The UPDRS III scale has 14 categories including speech, facial expression, tremor at rest, action tremor, rigidity, finger tapping ability, ability to open and close hands, ability to rapidly alternate hand movements, leg agility, ability to rise from a chair, posture, gait, response to postural displacement (e.g., push), and bradykinesia. Patients rate each of these categories from 0 to 4, with 0 being normal function and 4 being the worst. Categories assessing appendages are rated for both left and right sides, allowing a maximum score (worst outcome) of 108. |
Day 1 programming visit
|
Másodlagos eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
---|---|---|
Side Effect Thresholds - Single Contact vs. Steering
Időkeret: Day 1 programming visit
|
This endpoint determined how much current (mA) could be applied before side effects appeared when using 60 microsecond pulse widths.
Values were obtained for when current was delivered through a single contact or divided between two contacts (steering).
|
Day 1 programming visit
|
Resting Tremor Severity - Single Contact vs. Steering
Időkeret: Day 1 programming visit
|
Resting tremor was measured by a motion sensor system (Kinesia System) while either using a single contact or steering current between two contacts.
Stimulation was at amplitudes defined as the therapeutic threshold for rigidity.
The Kinesia System was worn by patients to measure motion parameters including linear acceleration and angular velocity during different tasks, then provided an output score on a scale of 0 (no symptoms) to 4 (severe symptoms).
|
Day 1 programming visit
|
Finger Tapping Amplitude - Single Contact vs. Steering
Időkeret: Day 1 programming visit
|
Severity of finger-tapping bradykinesia was measured by a motion sensor system (Kinesia System) when either using a single contact or steering current between two contacts.
Stimulation was at amplitudes defined as the therapeutic threshold for rigidity.
The Kinesia System was worn by patients to measure motion parameters including linear acceleration and angular velocity during different tasks, then provided an output score on a scale of 0 (no symptoms) to 4 (severe symptoms).
|
Day 1 programming visit
|
Egyéb eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
---|---|---|
Dorsal-Ventral Current Steering Therapeutic Window
Időkeret: Day 1 programming visit
|
The therapeutic window refers to the range of stimulus amplitudes that provide a therapeutic effect without side effects.
In other words, this measure reports the stimulus amplitude difference between the full rigidity control threshold and the first stimulation induced side effect threshold at current steering settings (current divided 50% between adjacent electrodes).
|
Day 1 programming visit
|
Resting Tremor Severity - Pulse Width and Dorsal-Ventral Steering
Időkeret: Day 1 programming Visit
|
Resting tremor was measured by a motion sensor system (Kinesia System) at 60 µs, 30 µs, and current steering settings, at amplitudes defined as the therapeutic threshold for rigidity.
The Kinesia System was worn by patients to measure motion parameters including linear acceleration and angular velocity during different tasks, then provided an output score on a scale of 0 (no symptoms) to 4 (severe symptoms).
|
Day 1 programming Visit
|
Finger Tapping Amplitude - Pulse Width and Dorsal-Ventral Steering
Időkeret: Day 1 programming visit
|
Severity of finger-tapping bradykinesia was measured by a motion sensor system (Kinesia System) at 60 µs, 30 µs, and dorsal and ventral current steering settings.
Stimulation was at amplitudes defined as the therapeutic threshold for rigidity.
The Kinesia System was worn by patients to measure motion parameters including linear acceleration and angular velocity during different tasks, then provided an output score on a scale of 0 (no symptoms) to 4 (severe symptoms).
|
Day 1 programming visit
|
Együttműködők és nyomozók
Szponzor
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete
Elsődleges befejezés (Tényleges)
A tanulmány befejezése (Tényleges)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Becslés)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
További vonatkozó MeSH feltételek
Egyéb vizsgálati azonosító számok
- A4009
- CDM00060272 (Egyéb azonosító: BSC Protocol Number)
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .
Klinikai vizsgálatok a Vercise DBS settings
-
Deep Brain Innovations LLCMegszűntParkinson kórEgyesült Államok
-
Chinese PLA General HospitalMég nincs toborzás
-
University Medical Center GroningenIsmeretlen
-
Academisch Medisch Centrum - Universiteit van Amsterdam...Radboud University Medical Center; University Medical Center Groningen; Maastricht... és más munkatársakToborzásKözponti idegrendszeri betegségek | Esszenciális Tremor | Mély agyi stimuláció | Mágneses rezonancia | Diffúziós tenzor képalkotásHollandia
-
Jaimie M. HendersonHarvard Medical School (HMS and HSDM); The Cleveland Clinic; Weill Medical College... és más munkatársakBefejezve
-
University of Southern CaliforniaBefejezveParkinson kórEgyesült Államok
-
University of British ColumbiaMég nincs toborzás
-
St. George's Hospital, LondonUniversity of OxfordIsmeretlenParkinson kór
-
Ruijin HospitalNational Natural Science Foundation of China; Icahn School of Medicine at Mount... és más munkatársakVisszavontObszesszív-kompulzív zavarKína