Aortic Valve and Root Measurements Under Real-Time 3-Dimensional Visualization During Angiography

2017년 7월 13일 업데이트: NYU Langone Health

The American College of Cardiology/American Heart Association guidelines place symptomatic severe Aortic Stenosis as a class I indication for aortic valve replacement. With the recently approved Edwards-Sapien TAVR device and the ongoing investigations using the CoreValve TAVR device, patients ineligible or at high risk for open-heart surgery are now eligible to undergo TAVR.

Patients selected for TAVR undergo an EKG-gated cardiac Multislice CTA to evaluate aortic valve anatomy and aortic root dimensions for device sizing, as well as coronary angiography to define coronary anatomy. Both tests utilize contrast media to visualize anatomy, which may result in contrast-induced nephropathy in anywhere from 7.5% to more than 50% of patients depending on associated clinical risk factors.

There is a need to consolidate this pre-operative testing whenever possible, and with real-time 3-dimensional visualization of aortic valve and root anatomy using DynaCT cardiac acquisition in the cath lab angiography suite during the coronary angiography, there may be a benefit with reduced contrast load (20 to 35cc for DynaCT, 100cc for CTA).

We would like to make a comparison of aortic valve and root measurements using CTA and DynaCT to affirm the accuracy of DynaCT vs the CTA gold standard.

연구 개요

상세 설명

Calcific or degenerative aortic valve disease is one of the most common valvular diseases among the elderly and, with life expectancy continuing to increase, a major source of burden on current healthcare. Although patients with aortic stenosis (AS) can remain asymptomatic for a prolonged period, they do eventually develop symptoms, including angina, exertional syncope and heart failure, which is associated with a 5-year, 3-year, and 1-year, respectively, 50% mortality. The American College of Cardiology/American Heart Association guidelines place symptomatic severe AS as a class I indication for aortic valve replacement. Currently, surgical replacement of the aortic valve is the standard of care. However, with the recently approved Edwards-Sapient TAVR device and the ongoing investigations using the CoreValve TAVR device, patients ineligible or at high risk for open-heart surgery are now eligible to undergo TAVR.

Currently, patients selected for TAVR undergo an EKG-gated cardiac Multislice CTA to evaluate aortic valve anatomy and aortic root dimensions for device sizing. Patients also undergo coronary angiography to define coronary anatomy. Both tests utilize contrast media to visualize anatomy, which may result in contrast-induced nephropathy in anywhere from 7.5% to more than 50% of patients depending on associated clinical risk factors. Contrast-induced nephropathy can, in turn, lead to dialysis and/or increased mortality.

There is a need to consolidate this pre-operative testing whenever possible, and with real-time 3-dimensional visualization of aortic valve and root anatomy using DynaCT cardiac acquisition in the cath lab angiography suite during the coronary angiography, there may be a benefit with reduced contrast load (20 to 35cc for DynaCT, 100cc for CTA). DynaCT conducts fully automatic aortic root segmentation and overlays 3-dimensional structures onto 2-dimensional fluoroscopy in a few seconds (~5 seconds, 60 frames/second rotational angiography). The software automatically recognizes and represents the aortic root, detects and indicates coronary ostia and lowest cusp points of aortic valve (nadir points), and reconstructs commissure lines of aortic valve. Due to cardiac motion, 3-dimensional images without motion correction or cardiac gating will look blurred. Therefore, DynaCT reconstruction motion correction software will be used. The DynaCT motion correction software is a new algorithm that will correct for this motion and apply it during reconstruction- allowing the user to spin without using rapid pacing. However, first, a comparison of aortic valve and root measurements using CTA and DynaCT must be made to affirm the accuracy of DynaCT vs the CTA gold standard.

Study Design This is a pilot study evaluating patients undergoing TAVR at the New York University (NYU) Langone Medical Center. Patients will undergo the standard CTA protocol and invasive coronary angiography performed as part of the pre-operative assessment for TAVR. However, patients in this study will also undergo DynaCT during coronary angiography, utilizing 1 acquisition sweep and 20 to 35cc more of contrast media. Measurements of the major aortic annulus diameter, orthogonal minor aortic annulus diameter, aortic annulus perimeter, maximum ascending aorta diameter at 40mm above the annulus, sinus of Valsalva diameters, sinus of Valsalva heights, and aortic root angulation will be made using both the CTA and DynaCT protocols by a radiologist blinded to patient identity after the study has reached its target recruitment. Based on these measurements, a trained interventional cardiologist will select the appropriate TAVR size in a similarly blinded fashion after the study has reached its target recruitment (i.e. the data collected using DynaCT will not be used in the clinical decision making process). Data will be compared from both modalities using appropriate tests of reproducibility and correlation.

Study Technique CTA will be performed per standard methods as part of standard pre-TAVR evaluation.

Coronary angiography will be performed via femoral artery access (minimum 6Fr) as per standard pre-TAVR evaluation. After coronary angiography is performed, a 6Fr pigtail catheter will be placed in the aorta and 40cc of 50% contrast media diluted normal saline will be delivered using standard automated injection during dynaCT cardiac image acquisition. Intravenous delivery of contrast was considered as alternative.

As per standard care, intravenous access will be obtained and participants will be hydrated during the CTA, coronary angiography, and dynaCT procedures.

연구 유형

중재적

등록 (실제)

25

단계

  • 해당 없음

연락처 및 위치

이 섹션에서는 연구를 수행하는 사람들의 연락처 정보와 이 연구가 수행되는 장소에 대한 정보를 제공합니다.

연구 장소

    • New York
      • New York, New York, 미국, 10016
        • NYU Langone Medical Center

참여기준

연구원은 적격성 기준이라는 특정 설명에 맞는 사람을 찾습니다. 이러한 기준의 몇 가지 예는 개인의 일반적인 건강 상태 또는 이전 치료입니다.

자격 기준

공부할 수 있는 나이

18년 이상 (성인, 고령자)

건강한 자원 봉사자를 받아들입니다

아니

연구 대상 성별

모두

설명

Inclusion Criteria:

  • Subjects must be more than 18 years of age with a plan to undergo TAVR.

Exclusion Criteria:

  • Subjects will be excluded if they meet one of the following criteria:

    1. estimated glomerular filtration rate <40 mL/min/1.73m2;
    2. highly irregular heart rates, frequent extra systoles, additional extra pulsations, not in sinus heart rhythm;
    3. allergy to contrast media;
    4. pregnant; or
    5. Unable to consent.

공부 계획

이 섹션에서는 연구 설계 방법과 연구가 측정하는 내용을 포함하여 연구 계획에 대한 세부 정보를 제공합니다.

연구는 어떻게 설계됩니까?

디자인 세부사항

  • 주 목적: 특수 증상
  • 할당: 해당 없음
  • 중재 모델: 단일 그룹 할당
  • 마스킹: 없음(오픈 라벨)

무기와 개입

참가자 그룹 / 팔
개입 / 치료
실험적: Real-time 3-dimensional DynaCT
Patients will undergo the standard CTA protocol and invasive coronary angiography performed as part of the pre-operative assessment for TAVR. Patients in this study will also undergo DynaCT during coronary angiography, utilizing 1 acquisition sweep and 20 to 35cc more of contrast media. Measurements of the major aortic annulus diameter, orthogonal minor aortic annulus diameter, aortic annulus perimeter, maximum ascending aorta diameter at 40mm above the annulus, sinus of Valsalva diameters, sinus of Valsalva heights, and aortic root angulation will be made using both the CTA and DynaCT protocols by a radiologist blinded to patient identity. Based on these measurements, a trained interventional cardiologist will select the appropriate TAVR size for the patient.
After coronary angiography is performed, a 6Fr pigtail catheter will be placed in the aorta and 40cc of 50% contrast media diluted normal saline will be delivered using standard automated injection during dynaCT cardiac image acquisition. Intravenous delivery of contrast as alternative.

연구는 무엇을 측정합니까?

주요 결과 측정

결과 측정
측정값 설명
기간
Correlation of aortic annulus measurement measured on invasive 3D dyna CT (experimental) and on non-invasive CT scan (standard of care)
기간: 1 day
Each patient will undergo non-invasive CT scan for aortic annulus measurement (in order to choose appropriate TAVR size) as per standard of care. They will then undergo invasive dyna CT for aortic annulus measurement during their clinically-indicated invasive coronary angiogram. The primary outcome is correlation between the aortic annulus dimension measurement made on dyna CT and non-invasive CT.
1 day

2차 결과 측정

결과 측정
측정값 설명
기간
TAVR size
기간: 1 day
A trained interventional cardiologist will evaluate the measurements made from each CTA abd DynaCT image and select the appropriate TAVR size in a blinded fashion.
1 day

공동 작업자 및 조사자

여기에서 이 연구와 관련된 사람과 조직을 찾을 수 있습니다.

스폰서

연구 기록 날짜

이 날짜는 ClinicalTrials.gov에 대한 연구 기록 및 요약 결과 제출의 진행 상황을 추적합니다. 연구 기록 및 보고된 결과는 공개 웹사이트에 게시되기 전에 특정 품질 관리 기준을 충족하는지 확인하기 위해 국립 의학 도서관(NLM)에서 검토합니다.

연구 주요 날짜

연구 시작 (실제)

2014년 1월 6일

기본 완료 (실제)

2016년 6월 22일

연구 완료 (실제)

2017년 6월 13일

연구 등록 날짜

최초 제출

2014년 2월 19일

QC 기준을 충족하는 최초 제출

2014년 2월 26일

처음 게시됨 (추정)

2014년 2월 27일

연구 기록 업데이트

마지막 업데이트 게시됨 (실제)

2017년 7월 18일

QC 기준을 충족하는 마지막 업데이트 제출

2017년 7월 13일

마지막으로 확인됨

2017년 7월 1일

추가 정보

이 연구와 관련된 용어

약물 및 장치 정보, 연구 문서

미국 FDA 규제 의약품 연구

아니

미국 FDA 규제 기기 제품 연구

아니

미국에서 제조되어 미국에서 수출되는 제품

아니

이 정보는 변경 없이 clinicaltrials.gov 웹사이트에서 직접 가져온 것입니다. 귀하의 연구 세부 정보를 변경, 제거 또는 업데이트하도록 요청하는 경우 register@clinicaltrials.gov. 문의하십시오. 변경 사항이 clinicaltrials.gov에 구현되는 즉시 저희 웹사이트에도 자동으로 업데이트됩니다. .

관상동맥 질환에 대한 임상 시험

Real-time 3-dimensional DynaCT에 대한 임상 시험

3
구독하다