Comparison of BMD Measurement by DEXA to BeamMed Speed-of-Sound Measurement at Forearm in Patients With Gaucher Disease

November 18, 2014 updated by: Ari Zimran, Shaare Zedek Medical Center

Comparison of BMD Measurements by Dual Energy X-Ray Absorptiometry and Multiple Sites to BeamMed Speed-of-Sound Measurements at the Forearm in Patients With Gaucher Disease

The purpose of this study is to compare the accuracy and comparability and secondarily to assess the values achieved by measurement of the forearm BMD by DXA and SOS by BeamMed, relative to standard DXA evaluations at the FN and LS.

Study Overview

Detailed Description

Dual-energy X-ray absorptiometry (DXA) is the current gold standard for the clinical diagnosis of osteoporosis based on measurement of bone mineral density (BMD) [Baim et al, 2006]. As DXA technology continues to evolve, new instruments and technologies are introduced [Shepherd et al, 2005], making it necessary to document how these advances compare to prior densitometers.

The purpose of this study is to assess bone mineral density (BMD) comparability and precision using a standard Hologic DXA unit and the quantitative ultrasound-driven speed-of-sound (SOS) BeamMed technology. The physical and mechanical properties of bone that are measured by SOS are different than bone density measured by DXA and reflect bone strength and elasticity [Goossens et al, 2008]. The cohorts will be patients with Gaucher disease [Zimran & Elstein, 2010] who currently are requested to undergo BMD evaluation only bi-annually [Mistry et al, 2011], partly because of the radiation of DXA scans (admittedly low relative to standard X-rays and CT). Patients with Gaucher disease are at added risk for osteonecrosis and pathological fractures but to date no biomarker or technology has been able to predict which patients are imminently at risk. Because the option of disease-specific but very costly enzyme replacement therapy (ERT) [Barton et al, 1991] is predicating on the estimated risk of severe disease, predicting skeletal damage (especially osteonecrosis at the femoral neck, FN) because of osteopenia or osteoporosis has clinically relevant ramifications [Mistry et al, 2011]. Additionally, it has been shown that there is a correlation between low BMD at the lumbar spine (LS) and anemia in patients with Gaucher disease [Khan et al, 2012], so that with correction of anemia with ERT it is hoped that there will be amelioration of BMD values, and this too would be worth performing at closer intervals.

The least inconvenient and shortest evaluation for BMD is at the forearm. For approximately a year, all patients at the Gaucher disease have had DXA forearm evaluations when undergoing routine testing. It is the intention of this study to prospectively compare forearm SOS results [Olszynski et al, 2013] with those from the standard DXA equipment with the possibility of using SOS as a quantitative measure of bone integrity on a more regular basis.

The strategy will be to measure forearm BMD by DXA and SOS by BeamMed for comparison of accuracy and comparability and secondarily to assess the values achieved at the forearm relative to standard DXA evaluations at the FN and LS.

Patients who arrive at the Gaucher Clinic for routine monitoring that includes DXA evaluation of BMD at FN, LS, and forearm will be requested to consent to undergo SunBeam evaluation at the forearm as well. The same DXA hardware and software (Hologic, Bedford, MA) will be used for all examinations as have been employed by the Gaucher Clinic since August 2011.

The SOS technology as developed by the BeamMed Company (originally, the Sunlight Ultrasound Technologies, Israel) is based on the physical property that ultrasound waves travel faster through bone than soft tissue. The bone sonometer measures the time that elapses between the axially transmitted sound generated as an inaudible high-frequency pulsed acoustic signal at a center frequency of 1.25 MHz and the first reception of a signal after it has traveled through the selected bone. This is recorded as the SOS in m/sec. The ultrasonic waves are transmitted and received by transducers embedded in the ultrasound probe. As the ultrasound wave travels, the speed, dispersion, and attenuation of the signal is influenced by the density, elasticity, and cohesiveness of the medium, so that the greater the density and elasticity of the medium, the faster the speed of propagation [Ashman et al, 1984].

The device consists of a desktop main unit and 3 probes of different sizes. The small probe (1.4cm x 2.7cm x 11cm), 900-1000 KHz will be used for this study. Calibration is performed each day using a standardized calibrating plastic provided by the company. The SOS is determined as the mean of 3 scans that are found to be statistically similar (coefficient of variation below 1.2%), as calculated by the software provided with the device. Based on these 3 readings, the 95th percentile of the measured SOS is determined and used for statistical analysis.

Study Type

Interventional

Enrollment (Actual)

100

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Jerusalem, Israel
        • Shaare Zedek Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

Yes

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Age: >18
  • Gaucher disease type 1
  • Able to undergo a DEXA test

Exclusion Criteria:

  • Minors

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: SCREENING
  • Allocation: NA
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: SOS forearm test
BeamMed Speed-of-Sound bone strength test at forearm
BeamMed Speed-of-Sound bone strength test at forearm in addition to the DEXA test
Other Names:
  • OMNISENSE 7000S ULTRASOUND BONE SONOMETER

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Time Frame
BMD measurements by DEXA and SOS
Time Frame: 6 months
6 months

Secondary Outcome Measures

Outcome Measure
Time Frame
T-score and Z-score by DEXA and SOS
Time Frame: 6 months
6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

February 1, 2014

Primary Completion (ACTUAL)

September 1, 2014

Study Completion (ACTUAL)

September 1, 2014

Study Registration Dates

First Submitted

January 27, 2014

First Submitted That Met QC Criteria

February 18, 2014

First Posted (ESTIMATE)

February 20, 2014

Study Record Updates

Last Update Posted (ESTIMATE)

November 19, 2014

Last Update Submitted That Met QC Criteria

November 18, 2014

Last Verified

November 1, 2014

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.