Bio Impedance-assisted Monitoring of Chronic Hemodialysis Patients

October 13, 2019 updated by: Adi Leiba, Assuta Ashdod Hospital

Regional Electric Bio Impedance-assisted Monitoring of Cardiac, Vascular Resistance and Fluid Balance, in Chronic Hemodialysis Patients

Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30 patients, who are prone to develop hypotensive episode during dialysis. Sitting blood pressures will be measured immediately prior to each hemodynamic measurement: before initiation of dialysis, every each hour and in the beginning of hypotension episode, just before the end and 10 min after the end of the treatment.

Gender, age, height, weight, electrode location and blood pressure data will be entered into the device. The device will measure and calculate hemodynamic parameters on each heart beat during 60 s and provides the averaged parameters.

Technology for hemodynamic measurements:

The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug Administration indication for use of the device states 'NICaS is intended to monitor and display hemodynamic parameters in males and females with known or suspected cardiac disorders needing cardiac assessment'.

SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral ankle above the posterior tibial pulse (Figure 1).

Figure 1 : Sensor location

SV is calculated by Frinerman's formula:

SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the arterial system as a result of intraarterial expansion during systole, R is basal resistance, q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected for gender and age, KWis the correction of weight according to ideal values, HF is a hydration factor that takes into account the ratio between R and body mass index (BMI), which is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b is the diastolic time interval. SV is automatically calculated every 20 s and is the average of three measurements obtained consecutively during 60 s of monitoring. The SV index is calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an oscillometric method, sitting systolic and diastolic blood pressure measurements were made automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3], cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn - s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.

As the device measures pulsatile flow and is blinded to constant flow, fluid removal during dialysis has no impact on measurement accuracy. This was recently validated by correlating SV to ECG measurements during hemodialysis treatments. Good correlation was maintained during treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis); curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the normal population are depicted by a dotted octagon.

Study Overview

Status

Unknown

Intervention / Treatment

Detailed Description

Hemodynamic trends will be assessed using the device, in 100 dialysis sessions in 30 patients, who are prone to develop hypotensive episode during dialysis. Sitting blood pressures will be measured immediately prior to each hemodynamic measurement: before initiation of dialysis, every each hour and in the beginning of hypotension episode, just before the end and 10 min after the end of the treatment.

Gender, age, height, weight, electrode location and blood pressure data will be entered into the device. The device will measure and calculate hemodynamic parameters on each heart beat during 60 s and provides the averaged parameters.

Technology for hemodynamic measurements:

The device (NICaS, NI Medical) is a noninvasive regional bioimpedance cardiac measurement and analysis system (FDA 510k clearance no. K080941, 12 June 2009). The US Food and Drug Administration indication for use of the device states 'NICaS is intended to monitor and display hemodynamic parameters in males and females with known or suspected cardiac disorders needing cardiac assessment'.

SV will be measured by applying an alternating electrical current of 1.4mA at 30 kHz frequency through the patient's body via two pairs of tetrapolar sensors, one pair placed on the wrist of the nonaccess arm above the radial pulse and the other pair on the contralateral ankle above the posterior tibial pulse (Figure 1).

Figure 1 : Sensor location

SV is calculated by Frinerman's formula:

SV¼(dR/R) - q - (L2/Ri) - (ab)/b - KW - HF [2-4], where dR is the impedance change in the arterial system as a result of intraarterial expansion during systole, R is basal resistance, q is blood electrical resistance, L is the patient's height, Ri is basal resistance corrected for gender and age, KWis the correction of weight according to ideal values, HF is a hydration factor that takes into account the ratio between R and body mass index (BMI), which is correlated to body water volume, ab is the electrocardiogram (ECG) R-R wave interval and b is the diastolic time interval. SV is automatically calculated every 20 s and is the average of three measurements obtained consecutively during 60 s of monitoring. The SV index is calculated as SV/body surface area using the Du Bois formula [11]. Heart rate is calculated from a one channel ECG and cardiac (output) index¼SV index - heart rate/1000. Using an oscillometric method, sitting systolic and diastolic blood pressure measurements were made automatically by the dialysis machine. Mean arterial pressure [2 - (diastolicsystolic)/3], cardiac power index [CPI; mean arterial pressure (MAP) -cardiac index - 0.0022 w/m2; normal range 0.45-0.85w/m2] [12, 13] and total peripheral resistance (MAP/ cardiac index - 80 dyn - s/cm5 - m2; normal range 1600-3000 dyn - s/cm5- m2) [13] will be calculated.

As the device measures pulsatile flow and is blinded to constant flow, fluid removal during dialysis has no impact on measurement accuracy. This was recently validated by correlating SV to ECG measurements during hemodialysis treatments. Good correlation was maintained during treatment. Further, NICaS performance immunity to fluid reduction was demonstrated by the maintenance of correlation to ECG results throughout dialysis treatments [9]. The results are drawn on hemodynamic graphs showing the MAP (y-axis) as a function of cardiac index (x-axis); curves of total peripheral resistance index (TPRI) and CPI are displayed. Ranges for the normal population are depicted by a dotted octagon.

Study Type

Interventional

Enrollment (Anticipated)

30

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

      • Ashdod, Israel, 7747629
        • Recruiting
        • Nephrology and Hypertention Institute, hemodialysis unit
        • Contact:

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 85 years (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

-Chronic hemodialysis patients

Exclusion Criteria:

  • Age limits

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: PREVENTION
  • Allocation: RANDOMIZED
  • Interventional Model: PARALLEL
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
EXPERIMENTAL: CHRONIC HEMODIALYSIS PATIENTS, NICAS-ASSISTED MONITORING
Chronic hemodialysis patients, NICAS-assisted monitored
Electric regional bio impedance device
NO_INTERVENTION: Control hemodialysis patients
Chronic hemodialysis patients, conventional clinical care and monitoring.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Hemodynamic adverse events
Time Frame: One year
Extreme blood pressure decrease, hospitalization, death
One year

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Adi Leiba, Prof, ASSUTA HOSPITAL ASHDOD

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ACTUAL)

November 18, 2018

Primary Completion (ANTICIPATED)

December 31, 2019

Study Completion (ANTICIPATED)

December 31, 2020

Study Registration Dates

First Submitted

October 13, 2019

First Submitted That Met QC Criteria

October 13, 2019

First Posted (ACTUAL)

October 16, 2019

Study Record Updates

Last Update Posted (ACTUAL)

October 16, 2019

Last Update Submitted That Met QC Criteria

October 13, 2019

Last Verified

October 1, 2019

More Information

Terms related to this study

Other Study ID Numbers

  • 0069-18

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Chronic Kidney Diseases

Clinical Trials on NICAS

3
Subscribe