An Investigation Into a New Manufacture Technique of Trans-femoral Sockets

November 18, 2020 updated by: Össur Ehf

MSS - Clinical Investigation An Investigation of a New Manufacture Technique of a Trans-femoral Socket and Its Function for Amputees.

The investigation will be of a single group prospective pre/post design with follow up at 6 weeks (6WFU) and 6 months (6MFU); comparing to outcomes at baseline (BL).

Device(s) being tested: Interchangeably: Direct Socket TF; DS-TF; DS Trans-femoral.

The DS-TF is a trans-femoral (TF) set-up of the currently marketed trans-tibial version. It is not marketed in the USA and has not been registered with the FDA.

The comparator is the current socket the subject is using when enrolled in the study.

Subjects recruited: Minimum 50

Inclusion criteria:

50Kg< body weight < 160Kg Cognitive ability to understand all instructions and questionnaires in the study; Patients who have undergone a transfemoral amputation > 1 year post amputation Older than 18 years Willing and able to participate in the study and follow the protocol Circular dimension of 40-65 cm at the crotch Residual limb length at least 20 cm from ischium to distal end Currently using a prosthetic liner Locking users that can successfully be fitted with Iceross® Transfemoral Locking, OR Seal-In users that can successfully be fitted with either Iceross Seal-In® X5 TF or Iceross Transfemoral Seal-In®

Exclusion criteria: 50Kg> body weight > 160Kg Users with cognitive impairment Patients who have undergone a transfemoral amputation <1 year post amputation Younger than 18 years Not a prosthetic ambulator, uses only the prosthesis for cosmetic reasons Circular dimension of less than 40 cm or greater than 65 cm at the crotch Residual limb length less than 20 cm from ischium to distal end Currently not using a prosthetic liner

Study Overview

Status

Completed

Conditions

Intervention / Treatment

Detailed Description

The test will be a single group prospective pre/post design. Amputees are a small proportion of the general population. The population group specified in the inclusion/exclusion criteria is a further subsample of amputees. For practical reasons, i.e. to achieve statistical power, it is therefore more feasible to use within-subject comparison rather than creating study arms to compare. Furthermore, as mobile amputees generally have and use a prosthetic device for their daily activities, within-comparison is feasible comparing to the subjects' previous device.

All investigational activities will be conducted at prosthetic out-patient clinics.

As stated above the primary endpoint is successful use rate, see Table 2, and the secondary endpoints are listed in the table from B-H in that respective order of significance. See previous chapter on objectives and hypothesis and Table 2 for rationale.

Drop-outs and withdrawals will not be replaced.

Equipment required for each subject:

  • Pen/pencil
  • Printed out instruments
  • Investigational device: Materials for making DS-TF socket o Designated locking or seal-in liner to be fitted with the socket

Additionally, each LPI or LCI will have access to:

  • Equipment for DS-TF socket manufacturing
  • iPad with the Össur ProApp installed for data collection The equipment used does not require specific monitoring, maintenance or calibration procedures.

The purpose of the study is to evaluate in a cohort the time it takes to manufacture and delivery (lead time) of a definitive prosthesis (definition of finished prosthetic leg is, a prosthesis that is individually tailored and able to be used both indoors and outdoors, with the maximum weight on the patient equivalent to 160 kg). The existing prosthesis is to be evaluated as well as the new socket. The evaluation of function will take place at baseline, 6 weeks and 6 months after deliverance.

The following outcome instruments will be used:

• At all three study evaluation points; initial fitting, 6 wees and 6 months:

  • OPUS (LEFS, CSD, CSS) https://www.sralab.org/rehabilitation-measures/orthotics-prosthetics-users-survey
  • ProAMP (to test the current function with prosthesis)
  • EQ-5D (for life quality)
  • Socket Comfort Score (SCS)
  • CLASS - socket fit
  • Plus-M
  • ABC
  • TUG

Cost of fitting, to all parties, will be collected and reported. Whatever the outcome of the study will be, it will be very informative for healthcare systems, both payers and providers, and manufacturer of equipment for amputees.

The total time required to implement the full clinical investigation is expected to be approximately 26 months. Each individual subject is expected to participate in the clinical investigation for a minimum of 6 weeks, up to 6 months. The estimated time needed to include the required number (enrolment period) is 20 months.

Study Type

Interventional

Enrollment (Actual)

47

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Alabama
      • Mobile, Alabama, United States, 36606
        • Alabama Artificial Limb Specialists
      • Montgomery, Alabama, United States, 36116
        • Alabama Artificial Limb Specialists
      • Opelika, Alabama, United States, 36801
        • Alabama Artificial Limb Specialists
    • Arkansas
      • Greenwood, Arkansas, United States, 72936
        • Van Der Watt Prosthetics & Orthotics
    • Minnesota
      • Maplewood, Minnesota, United States, 55109
        • Tillges Certified Orthotic Prosthetic Inc
    • New York
      • Albany, New York, United States, 12206
        • Sampson's Prosthetics Lab
      • Schenectady, New York, United States, 12304
        • Sampson's Prosthetics Lab
    • Oregon
      • Portland, Oregon, United States, 97230
        • Eastside Orthotics and Prosthetics
    • Pennsylvania
      • Cranberry Township, Pennsylvania, United States, 16066
        • Union Orthotics & Prosthetics Co
      • E. Rochester, Pennsylvania, United States, 15074
        • Union Orthotics & Prosthetics Co
      • Greensburg, Pennsylvania, United States, 15601
        • Union Orthotics & Prosthetics Co
      • Pittsburgh, Pennsylvania, United States, 15201
        • Union Orthotics & Prosthetics Co
      • Pittsburgh, Pennsylvania, United States, 15236
        • Union Orthotics & Prosthetics Co
    • Tennessee
      • Johnson City, Tennessee, United States, 37604
        • Victory Orthotics and Prosthetics Inc.
      • Kingsport, Tennessee, United States, 37660
        • Victory Orthotics and Prosthetics Inc.
      • Knoxville, Tennessee, United States, 37922
        • Victory Orthotics and Prosthetics Inc.
    • Virginia
      • Roanoke, Virginia, United States, 24012
        • Virginia Prosthetics and Orthotics

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (ADULT, OLDER_ADULT)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Cognitive ability to understand all instructions and questionnaires in the study.
  • Patients who have undergone a trans-femoral amputation > 1 year post amputation.
  • Willing and able to participate in the study and follow the protocol.
  • Circular dimension of 40-65 cm at the crotch.
  • Residual limb length at least 20 cm from ischium to distal end.

Exclusion Criteria:

  • 50Kg > body weight > 160Kg
  • Users with cognitive impairment
  • Not a prosthetic ambulator, uses only the prosthesis for cosmetic reasons

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: TREATMENT
  • Allocation: NON_RANDOMIZED
  • Interventional Model: SINGLE_GROUP
  • Masking: NONE

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
ACTIVE_COMPARATOR: Clinical need
Participants enrolled that require a new socket.

The device is Class I, a low-risk product, and is a further development of a well-established technology. Össur Direct Socket (DS) technology has been on the market since 1996 (ICEX first generation) for the manufacture of leg prosthesis. This prosthetic socket for trans-femoral amputees (referred hereby to as DS-TF) has the same basic functional design as the Direct Socket (MSS) for trans-tibial amputees (referred hereby to as DS-TT) that is available on the market today in addition to the newly added silicon brim.

A prosthetic socket is a non-invasive, non-sterile, single user reusable device, which is used as part of prosthetic system. The aspect of the prosthesis that is in direct physical contact with the amputee is usually not the socket, to which a prosthesis is connected to, but the liner that serves as an interface between the amputee and the rest of the prosthesis. An amputee typically wears a prosthesis and thereby utilizes a socket, for up to 18 hours a day.

Other Names:
  • DS-TF
  • DS Trans-femoral
EXPERIMENTAL: No clincal need
Participants enrolled that do not require a new socket

The device is Class I, a low-risk product, and is a further development of a well-established technology. Össur Direct Socket (DS) technology has been on the market since 1996 (ICEX first generation) for the manufacture of leg prosthesis. This prosthetic socket for trans-femoral amputees (referred hereby to as DS-TF) has the same basic functional design as the Direct Socket (MSS) for trans-tibial amputees (referred hereby to as DS-TT) that is available on the market today in addition to the newly added silicon brim.

A prosthetic socket is a non-invasive, non-sterile, single user reusable device, which is used as part of prosthetic system. The aspect of the prosthesis that is in direct physical contact with the amputee is usually not the socket, to which a prosthesis is connected to, but the liner that serves as an interface between the amputee and the rest of the prosthesis. An amputee typically wears a prosthesis and thereby utilizes a socket, for up to 18 hours a day.

Other Names:
  • DS-TF
  • DS Trans-femoral

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
OPUS
Time Frame: 6 months
Patient Self report. Paper based. Evaluation of service and device.
6 months
CLASS
Time Frame: 6 months
Amputee Self-report: socket quality evaluation (comfort, pain, stability, aesthetics); collected with ProApp.
6 months
AmpPro
Time Frame: 6 months
CPO evaluation of amputee, objective scoring: activity/mobility; collected with ProApp.
6 months
PLUS-M
Time Frame: 6 months
Amputee Self-report: activity/mobility; collected with ProApp.
6 months
EQ-5D
Time Frame: 6 months
Amputee self report: QoL, paper based.
6 months
ABC
Time Frame: 6 months
Amputee self report: Activity-Balance Confidence scale; collected with ProApp.
6 months
TUG
Time Frame: 6 months
Activity: Timed up and go; collected with ProApp.
6 months
CPO usability
Time Frame: 6 months
In-house generated instrument; paper based
6 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Observation log
Time Frame: Baseline only
Timing and documentation on how the DS-TF is fitted.
Baseline only

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Investigators

  • Principal Investigator: Justin Pratt, CPO, Össur Ehf

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (ACTUAL)

April 30, 2018

Primary Completion (ACTUAL)

June 15, 2020

Study Completion (ACTUAL)

July 1, 2020

Study Registration Dates

First Submitted

March 16, 2020

First Submitted That Met QC Criteria

March 16, 2020

First Posted (ACTUAL)

March 18, 2020

Study Record Updates

Last Update Posted (ACTUAL)

November 20, 2020

Last Update Submitted That Met QC Criteria

November 18, 2020

Last Verified

November 1, 2020

More Information

Terms related to this study

Other Study ID Numbers

  • CIP2017122866

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.