10-year Follow-up of a Wide Diameter Bone Anchored Hearing Implant
Long-term Follow-up of a Wide Diameter Bone Anchored Hearing Implant: the 10-year Experience on Stability, Survival and Tolerability of the BIA300®
Study Overview
Status
Conditions
Detailed Description
This study is a continuation of two previous conducted trials:
Stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation new BIA300 (test) implant vs previous generation (control) implant:
- CAG5173: 6-month(1) and 3-year(2) data.
- CBAS5562: 5-year data(3).
- 3 weeks loading: stability, long term survival and tolerability of a novel Baha® implant system, a multi-centre investigation 3 week loading vs 6+ week loading (= same group as control group trial 1). 6- month(4) and 3-year(5) data.
In this study the (5-year data for trial 2 and) 10-year data regarding stability, survival and soft tissue tolerability from the above mentioned trials will be evaluated and compared. The previous trials were multicentre, but the current study will be single-centred and therefore only include the participants of the Radboud site. Data will be collected at a visit at (5 years, for trial 2 and) 10 years (+/- 6 months) after implantation.
Up to now, no RCT of these implants (with the loading of 3 weeks after implantation) has assessed stability, survival and tolerability over a 10-year follow-up period.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Gelderland
-
Nijmegen, Gelderland, Netherlands, 6525EX
- Radboudumc
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- To be included in the original trials, patients had to be at least 18 years old.
Exclusion Criteria:
- Exclusion criteria were an inability to follow investigational procedures, any factor at the discretion of the investigator that was considered to contraindicate participation, and any disease or treatment known to compromise the bone quality at the implant site. For patients who had lost or removed the implant, the time to implant loss was recorded. Patients who could not attend the 5 or 10 year follow-up visit for other reasons were included in the implant survival analysis as well; the last available information regarding implant survival was obtained verbally from the patient, from medical records, or from information captured from the original investigations. Furthermore, patients were excluded if they had a bone thickness less than 4 mm at the implant site during implantation.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: TREATMENT
- Allocation: NON_RANDOMIZED
- Interventional Model: PARALLEL
- Masking: NONE
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
---|---|
ACTIVE_COMPARATOR: Control implant loaded at 6 weeks post-surgery
The control implant included the previous generation as-machined titanium Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm conically shaped abutment.
|
The control implant included the previous generation as-machined titanium Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm conically shaped abutment.
|
EXPERIMENTAL: Test implant loaded at 6 weeks post-surgery
The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden).
This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
|
The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden).
This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
In addition to the difference in abutment shape and the wider diameter, the test implant incorporates small-sized threads at the implant neck and the moderately rough TiOblast™ (Dentsply, Mölndal, Sweden) surface on the intraosseous part of the implant6-8.
|
EXPERIMENTAL: Test implant loaded at 3 weeks post-surgery
The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden).
This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
|
The test implant and the implant used for the 3-week loading trial included the wide diameter titanium implant (diameter 4.5mm; length 4mm) with 6mm rounded, apically converging titanium abutment developed by Cochlear Bone Anchored Solutions AB (Mölnlyncke, Sweden).
This system with an additional minor change to the internal abutment connection design was later commercialized under the name Cochlear™ Baha® BIA300 Implant with abutment.
In addition to the difference in abutment shape and the wider diameter, the test implant incorporates small-sized threads at the implant neck and the moderately rough TiOblast™ (Dentsply, Mölndal, Sweden) surface on the intraosseous part of the implant6-8.
Loading the implant at 3 weeks after implantation (instead of 6 weeks after implantation)
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
implant stability
Time Frame: 10 years after implantation
|
Implant Stability Quotiënt (ISQ) will be measured at the 10y-visit.
The resonance frequency analysis renders an ISQ value ranging from 1 to 100.
Measurements shall be performed at the abutment level, as is done in the previous investigations.
The highest and lowest ISQ value out of two perpendicular measurements shall be recorded as ISQ High and ISQ low.
The measurement should preferably be done by the same Osstell instrument (Osstell, Gothenburg, Sweden) that was used in the previous completed study.
SmartPeg Type 43 and 55 shall be used during the measurement (Osstell, Gothenburg, Sweden).
The procedures for obtaining ISQ values is described in the manufacturer's instruction that will be handed out to the investigators.
|
10 years after implantation
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
implant survival
Time Frame: 10 years after implantation
|
All patients will be asked if they have experienced any implant osseointegration problems.
The time from implantation until implant loss or removal will be noted.
In case of implant removal, reason for removal shall be recorded.
|
10 years after implantation
|
soft tissue reaction
Time Frame: 10 years after implantation
|
Soft tissue reactions will be measured using the Holgers' classification & IPS-score. The Holgers' classification is designed to capture signs and symptoms of inflammation or infection at the site of implantation (scale 0-4). The IPS-score is a complete standardized assessment scale for skin complications for both percutaneous and transcutaneous implants for bone conduction devices (BCDs) with a proposed treatment advice based on the outcome. The IPS score has three subscales: inflammation (0-4); pain (0-2), and skin height (0-2). |
10 years after implantation
|
device use
Time Frame: 10 years after surgery
|
To evaluate the long-term device use of a Baha®.
|
10 years after surgery
|
Collaborators and Investigators
Collaborators
Study record dates
Study Major Dates
Study Start (ACTUAL)
Primary Completion (ACTUAL)
Study Completion (ACTUAL)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (ACTUAL)
Study Record Updates
Last Update Posted (ACTUAL)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- CAG5173/CBAS5562 10y
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hearing Loss
-
MED-EL Elektromedizinische Geräte GesmbHCompletedHearing Loss | Hearing Loss, Sensorineural | Hearing Loss, Bilateral | Hearing Loss, Conductive | Hearing Loss, Unilateral | Hearing Loss, MixedAustria, Germany, United Kingdom
-
University of California, San FranciscoPatient-Centered Outcomes Research InstituteRecruitingHearing Loss | Hearing Loss, Sensorineural | Hearing Loss, Bilateral | Hearing Loss, Conductive | Hearing Loss, Noise-Induced | Hearing Loss, Unilateral | Hearing Loss, Mixed | Hearing Disorders in ChildrenUnited States
-
Oticon MedicalRecruitingEar Diseases | Hearing Loss, Conductive | Hearing Loss Mixed | Hearing Disability | Conductive Hearing Loss | Conductive Hearing Loss, Bilateral | Conductive Hearing Loss, UnilateralUnited Kingdom
-
Oticon MedicalNot yet recruitingSensorineural Hearing Loss, Bilateral | Sensorineural Hearing Loss, Severe | Sensorineural Hearing Loss, Profound
-
Manchester University NHS Foundation TrustUniversity of ManchesterCompletedCochlear Hearing Loss | Sensorineural Hearing Loss, BilateralUnited Kingdom
-
Oticon MedicalCompletedConductive Hearing Loss | Conductive and Sensori-neural Hearing Loss in the Same Ear | Unilateral, Profound Sensori-neural Hearing LossUnited States
-
Frequency TherapeuticsCompletedHearing Loss, Sensorineural | Presbycusis | Noise Induced Hearing Loss | Sudden Hearing LossUnited States
-
Frequency TherapeuticsCompletedHearing Loss, Sensorineural | Noise Induced Hearing Loss | Sudden Hearing LossUnited States
-
Otologic Pharmaceutics, Inc.CompletedHearing Loss | Sensorineural Hearing Loss | Noise-Induced Hearing LossUnited States
-
Otologics LLCUnknownMixed Conductive and Sensorineural Hearing Loss, Bilateral
Clinical Trials on Baha® flange fixture (diameter 3.75mm; length 4mm) with 6mm abutment
-
Clark StanfordWithdrawn