Clinical Evaluation of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use Device and the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test Device
This study is designed as an international prospective, multicentric, clinical study to investigate the performance and usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Professional Use and Self- Test devices for the qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in human nasopharyngeal (NP) and mid-turbinate nasal swabs, respectively.
This study is part of the performance evaluation to support the CE conformity assessment procedures.
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
In total, a minimum of 1531 male and female subjects will be prospectively enrolled at multiple clinical sites. A minimum of 100 lay users will also be asked to interpret randomly assigned mock devices.
After obtaining written informed consent, a study-specific subject identification number (ID) will be assigned to the subject. All eligible subjects will be assigned to both Professional Use arm and the Self-Test arm of the study. Subject demographics and a brief medical history will be collected.
Lay users (self-tester or caregiver) will first collect one mid-turbinate nasal swab from both nostrils, and perform and interpret the Panbio™ Rapid Panel ST device, under the observation of a trained observer. All procedures for testing and result interpretation, including sample collection and extraction, will be conducted following the IFU provided in the kit. Each Panbio™ Rapid Panel ST result will be interpreted firstly by the lay user and followed immediately by the observer who will also photograph the test at the time of interpretation. Each lay user will fill out a user evaluation questionnaire, and a study staff member will complete an observer questionnaire for each lay user. A minimum of 100 lay users will also read mock device tests presented by a study staff member. Mock devices (simulated to represent testing results) will be used for usability testing (result interpretation).
A study staff member will then collect two NP swabs (one swab from each nostril) from all subjects. The first NP swab sample will be used to conduct a Panbio™ Rapid Panel PU test, following the IFU. The test result will be photographed at the time of result interpretation. Each study staff member who performs a Panbio™ Rapid Panel PU test will fill out a user evaluation questionnaire once, to assess the usability of the device.
The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual.
Each subject's demographic information, education, symptomology data, test results and questionnaires, where relevant, will be recorded on source documentation and transferred to an Electronic Data Capture system.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Contact
- Name: Simon Kordowich
- Phone Number: 0000000
- Email: simon.kordowich@abbott.com
Study Locations
-
-
South Carolina
-
Easley, South Carolina, United States, 29640
- UCCT @ AFC Urgent Care- Easley
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
- Adult
- Older Adult
Accepts Healthy Volunteers
Description
Inclusion Criteria:
Subjects from all age groups, able and willing to provide written informed consent or assent, suspected of having a respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms, and presenting at least two of the following symptoms: fever, headache, tiredness, dry cough, sore throat, runny or stuffy nose, muscle aches, loss of smell, loss of taste or shortness of breath, will be prospectively enrolled.
Exclusion Criteria:
- Subject has had a nasal or a NP swab taken within the last 4 hours and is not available for further testing after 4 hours has elapsed.
- Subject has active nose bleeds or acute facial injuries/trauma.
- Subject has received a nasal vaccine (i.e., FluMist®) within the previous fourteen (14) days.
- Subject is currently taking or has taken an antiviral medication- e.g., Amantadine, Rimantadine, Relenza® (Zanamivir), Tamiflu® (Oseltamivir Phosphate), and Flumadine®-for influenza or COVID-19 within the previous thirty (30) days.
- Subject is currently enrolled in a study to evaluate an investigational drug.
- Self-tester or caregiver has prior medical or laboratory training.
- Self-tester or caregiver has a visual impairment that cannot be restored using glasses or contact lenses or lay user is unable to read the Instructions for Use.
- Subject has a condition deemed unfit to safely perform the test.
- Subject is unwilling or unable to provide informed consent.
- Vulnerable populations as deemed inappropriate for the study by the site's PI.
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
---|---|
Experimental: Nasal sampling/Testing (Self-Test) + Nasopharyngeal sampling/Testing (Professional Use)
Lay users (self-tester or caregiver) will be provided with a Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test kit. Each lay user will collect one mid-turbinate nasal swab from both nostrils, perform and interpret the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test. All procedures for testing and result interpretation, including sample collection and extraction will be conducted by the lay user following the Instructions for Use provided in the kit. The first NP swab sample will be used to conduct a Panbio™ COVID- 19/FluA&B Rapid Panel Professional Use test. The second NP swab sample will be eluted in Universal Transport Medium (UTM) provided by Abbott/the Core Laboratory, labelled and stored, according to the laboratory manual. UTM samples will be shipped to the Core Laboratory for testing with RT-PCR protocols for Flu A, Flu B and SARS-CoV-2, according to the laboratory manual. |
The Panbio™ COVID-19/Flu A&B Rapid Panel Professional use device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in nasopharyngeal swabs, collected by a trained health worker. The Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device is a visual lateral flow assay, for rapid immunochromatographic qualitative detection of COVID-19 antigen, Influenza A antigen and Influenza B antigen in self-collected mid-turbinate nasal swabs. The test is designed to be used by lay users
Other Names:
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Sensitivity and specificity of the Panbio™ COVID-19/Flu A&B Rapid Panel device.
Time Frame: "up to 24 months"
|
To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID-19/Flu A&B Rapid Panel in fresh collected mid-turbinate nasal swab specimens from individuals suspected of respiratory viral infection consistent with COVID-19 and/or Flu by their healthcare provider within the first five days of the onset of symptoms in a Near Patient Testing setting, compared to Flu and SARS-CoV-2 RT-PCR tested using NP swab samples. Concordance analyisis of investigational device and reference method. Acceptance criterion: Sensitivity >80%, Specificity >95% To assess the clinical performance (sensitivity/specificity) of the Panbio™ COVID- 19/Flu A&B Rapid Panel Self-Test device, as performed by lay users (self-testers or caregivers) using fresh collected mid-turbinate nasal self- collected samples, when compared to Flu and SARS-CoV-2 RT- PCR using NP swab samples. |
"up to 24 months"
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Concordance of self-test and professional use device
Time Frame: "up to 24 months"
|
To assess concordance between the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device using mid-turbinate nasal self-collected samples and the Professional Use device using NP samples collected by a trained health worker.
|
"up to 24 months"
|
Usability of the professional use device
Time Frame: "up to 24 months"
|
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel professional use device when used by trained healthcare workers.
Usability is assessed by a questionnaire.
|
"up to 24 months"
|
Usability of the self test device
Time Frame: "up to 24 months"
|
To assess the usability of the Panbio™ COVID-19/Flu A&B Rapid Panel Self-Test device, as performed by lay users with no laboratory or clinical experience.
Usability testing will also comprise of the interpretation of mock devices.
User and observer questionnaires will also be utilized to assess usability.
|
"up to 24 months"
|
Collaborators and Investigators
Sponsor
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Estimated)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- SDRD-H-029-P
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on COVID-19
-
University of Roma La SapienzaQueen Mary University of London; Università degli studi di Roma Foro Italico; Bios Prevention SrlCompletedPost Acute Sequelae of COVID-19 | Post COVID-19 Condition | Long-COVID | Chronic COVID-19 SyndromeItaly
-
Yang I. PachankisActive, not recruitingCOVID-19 Respiratory Infection | COVID-19 Stress Syndrome | COVID-19 Vaccine Adverse Reaction | COVID-19-Associated Thromboembolism | COVID-19 Post-Intensive Care Syndrome | COVID-19-Associated StrokeChina
-
Massachusetts General HospitalRecruitingPost Acute COVID-19 Syndrome | Long COVID | Post Acute Sequelae of COVID-19 | Long COVID-19United States
-
Indonesia UniversityRecruitingPost-COVID-19 Syndrome | Long COVID | Post COVID-19 Condition | Post-COVID Syndrome | Long COVID-19Indonesia
-
Erasmus Medical CenterDa Vinci Clinic; HGC RijswijkNot yet recruitingPost-COVID-19 Syndrome | Long COVID | Long Covid19 | Post COVID-19 Condition | Post-COVID Syndrome | Post COVID-19 Condition, Unspecified | Post-COVID ConditionNetherlands
-
Dr. Soetomo General HospitalIndonesia-MoH; Universitas Airlangga; Biotis Pharmaceuticals, IndonesiaRecruitingCOVID-19 Pandemic | COVID-19 Vaccines | COVID-19 Virus DiseaseIndonesia
-
University of Witten/HerdeckeInstitut für Rehabilitationsforschung NorderneyCompletedPost-COVID-19 Syndrome | Long-COVID-19 SyndromeGermany
-
Jonathann Kuo, MDActive, not recruitingSARS-CoV2 Infection | Post-COVID-19 Syndrome | Dysautonomia | Post Acute COVID-19 Syndrome | Long COVID | Long Covid19 | COVID-19 Recurrent | Post-Acute COVID-19 | Post-Acute COVID-19 Infection | Post Acute Sequelae of COVID-19 | Dysautonomia Like Disorder | Dysautonomia Orthostatic Hypotension Syndrome | Post... and other conditionsUnited States
-
University Hospital, Ioannina1st Division of Internal Medicine, University Hospital of IoanninaRecruitingCOVID-19 Pneumonia | COVID-19 Respiratory Infection | COVID-19 Pandemic | COVID-19 Acute Respiratory Distress Syndrome | COVID-19-Associated Pneumonia | COVID 19 Associated Coagulopathy | COVID-19 (Coronavirus Disease 2019) | COVID-19-Associated ThromboembolismGreece
Clinical Trials on Panbio™
-
Abbott Rapid DxCompletedCOVID-19 | Influenza A | Influenza BUnited States
-
Abbott Rapid DxCompletedCOVID-19 | Influenza A | Influenza Type BUnited States
-
Abbott Rapid DxRecruitingCOVID-19 | Influenza A | Influenza Type BUnited States
-
Foundation for Innovative New Diagnostics, SwitzerlandBotswana Harvard AIDS Institute PartnershipCompletedCorona Virus InfectionBotswana
-
Murdoch Childrens Research InstituteRhinomed Pty LtdCompletedCOVID-19 | SARS CoV 2 InfectionAustralia
-
National Institute of Respiratory Diseases, MexicoCompleted
-
McGill University Health Centre/Research Institute...University of Cape Town Lung InstituteRecruiting
-
Abbott Rapid DxCompletedCOVID-19 | Influenza A | Influenza Type BUnited States
-
University of British ColumbiaProvidence Health & Services; Vancouver Airport Authority; WestJet Airlines; KLM...Completed
-
BaroNova, Inc.Completed