The Southern-Norway Post-Stroke Atrial Fibrillation Study (SNAPS)

February 5, 2026 updated by: Sorlandet Hospital HF

This study evaluates whether a procedure using a new wireless heart sensor patch is equal to or better than the standard hospital procedures and equipment at detecting an irregular heartbeat called Atrial Fibrillation (AF) after an ischemic stroke. Atrial fibrillation is a major cause of stroke, but it can be difficult to catch because it often comes and goes.

The study will include approximately 450 adults who have had a stroke or a transient "mini-stroke" (TIA) within the last two weeks. Participants will be assigned by chance (randomized) to one of two groups:

  • Group 1 (Intervention): Participants wear the "ECG247 Smart Heart Sensor." This is a small patch that sticks to the chest and connects to a smartphone. It is worn continuously for up to 14 days, even after leaving the hospital.
  • Group 2 (Standard Care): Participants receive the standard hospital check-up. This typically involves using a "Holter monitor" (a device with wires and electrodes) for a period of about 24 to 48 hours some time after leaving the hospital.

The main goal is to see if the procedure using the patch is equal to the standard procedure in detecting atrial fibrillation in participants. The study will also measure how quickly doctors can start the correct medication and how easy the patients find the devices to use.

Study Overview

Detailed Description

Background and Rationale Atrial fibrillation (AF) is a major risk factor for ischemic stroke, yet it often goes undetected due to its paroxysmal (intermittent) nature. While guidelines recommend prolonged heart rhythm screening after a stroke of undetermined cause, standard hospital practice often relies on short-term telemetry and ambulatory Holter monitoring. These standard methods can be limited by device availability, bulkiness, and short monitoring duration (typically 24-48 hours), potentially leading to under-diagnosis of AF.

This study aims to evaluate the clinical utility and cost-effectiveness of a novel, continuous patch ECG system (ECG247) initiated in the hospital setting compared to standard care.

Study Design This is a prospective, randomized, parallel-group, open-label (no masking) study conducted at a single stroke center. Approximately 450 eligible patients will be enrolled.

Recruitment and Stratification Patients admitted with ischemic stroke or transient ischemic attack (TIA) within the last two weeks are screened for eligibility, aiming for inclusion within 48 hours of admission. Participants are randomized in a 1:1 ratio to either the intervention arm or the standard care arm. Randomization is stratified by age into three groups: 18-64 years, 65-74 years, and ≥75 years.

Intervention Arm (ECG247 Smart Heart Sensor) Patients in this group will have the ECG247 patch applied to the anterior chest wall while in the hospital.

  • Device: The ECG247 is a wireless, water-resistant patch sensor that connects to a dedicated smartphone application via Bluetooth. It transmits data to a secure cloud service for analysis.
  • Duration: The device is worn continuously for up to 14 days. Monitoring begins in-hospital and continues after discharge.
  • Analysis: The system utilizes an algorithm to detect arrhythmia. Episodes of AF >30 seconds are considered positive findings. Study personnel and physicians verify algorithm-detected episodes.
  • Concurrent Care: These patients also receive standard in-house telemetry per hospital protocol while admitted.

Control Arm (Standard Care) Patients in this group receive standard heart rhythm monitoring according to current hospital guidelines.

  • In-hospital: Monitoring via continuous telemetry or intermittent ECGs based on availability and clinical priority.
  • Ambulatory: Upon discharge, patients are referred for ambulatory Holter monitoring. The duration is typically 24-48 hours as decided by the treating cardiologist.
  • Analysis: Holter recordings are analyzed by trained technicians and reviewed by a cardiologist.

Biomarker Analysis Blood samples will be collected from all participants to analyze N-terminal pro-B-type natriuretic peptide (NT-proBNP) levels. The study will assess the association between NT-proBNP levels and the detection of silent AF to evaluate its potential for risk stratification.

Outcomes and Follow-up

  • Primary Outcome: The detection rate of AF (defined as episodes ≥30 seconds).
  • Secondary Outcomes: Time to AF detection, time to initiation of anticoagulation therapy, total monitoring duration, and health economics.
  • User Experience: Patients in the intervention group will complete Patient Reported Experience Measures (PREMs) regarding usability and satisfaction.
  • Long-term Follow-up: Participants will be followed for up to 12 months to assess clinical endpoints, including secondary stroke, TIA, and major bleeding.

Study Type

Interventional

Enrollment (Estimated)

450

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Contact Backup

Study Locations

      • Kristiansand, Norway, 4615
        • Recruiting
        • Sørlandet Sykehus HF

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • Patients hospitalized with an ischemic stroke or transient ischemic attack (TIA), including amaurosis fugax, occurring within the last 2 weeks.
  • Initial evaluation of CT and/or CT angiography and/or MRI supports a diagnosis of TIA or ischemic stroke.
  • Available smartphone and access to the ECG247-app to be able to participate in the study.
  • Estimated life span of >6 month
  • Permanent address in Norway
  • Informed Consent, Capable of giving signed informed consent or consent through proxy as described in Appendix which includes compliance with the requirements and restrictions listed in the informed consent form (ICF) and in the study protocol.

Exclusion Criteria:

Participants are excluded from the study if any of the following criteria apply:

Medical Conditions

  • Known AF or atrial flutter prior to inclusion
  • Concomitant use of anticoagulation therapy or established contraindication to its use. To date this includes apixaban, rivaroxaban, edoxaban, dabigatran, warfarin and indirect thrombin inhibitors (except for short term thrombosis prevention).
  • Implanted pacemaker, ICD or loop-recorder
  • >70% stenosis of carotid artery on ipsilateral side to the stroke on CT angiography or ultrasound
  • Pregnancy

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: Randomized
  • Interventional Model: Parallel Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Intervention Arm (ECG247 Smart Heart Sensor)

Patients in this group will have the ECG247 patch applied to the anterior chest wall while in the hospital.

  • Device: The ECG247 is a wireless, water-resistant patch sensor that connects to a dedicated smartphone application via Bluetooth. It transmits data to a secure cloud service for analysis.
  • Duration: The device is worn continuously for up to 14 days. Monitoring begins in-hospital and continues after discharge.
  • Analysis: The system utilizes an algorithm to detect arrhythmia. Episodes of AF >30 seconds are considered positive findings. Study personnel and physicians verify algorithm-detected episodes.
  • Concurrent Care: These patients also receive standard in-house telemetry per hospital protocol while admitted.
This study aims to evaluate the clinical utility and cost-effectiveness of a novel, continuous patch ECG system (ECG247) initiated in the hospital setting compared to standard care.
Other: Control Arm (Standard Care)

Patients in this group receive standard heart rhythm monitoring according to current hospital guidelines.

  • In-hospital: Monitoring via continuous telemetry or intermittent ECGs based on availability and clinical priority.
  • Ambulatory: Upon discharge, patients are referred for ambulatory Holter monitoring. The duration is typically 24-48 hours as decided by the treating cardiologist.
  • Analysis: Holter recordings are analyzed by trained technicians and reviewed by a cardiologist.
Patients in this group receive standard heart rhythm monitoring according to current hospital guidelines.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Detection rate of atrial fibrillation
Time Frame: From enrollment to the end of heart rate monitoring. Will also be registered on follow up after 12 months.
The detection rate of AF (defined as episodes ≥30 seconds). Analysis will be performed on an intention-to-treat basis, and AF detection will be primarily assessed for its non-inferiority and secondarily for difference in proportions. An interim analysis for efficacy measures will be performed using the O'Brien-Flemming correction for the primary endpoint AF detection rate after 72 and 144 participants in each arm.
From enrollment to the end of heart rate monitoring. Will also be registered on follow up after 12 months.
Time from enrollment to treatment initiation with anticoagulant medicine
Time Frame: From enrollment until end of follow up (12 months after enrollment).
Days from enrollment until AF detection results in change of prophylactic treatment, defined as prescribed anticoagulant medication collected by the patient, retrieved from e-prescription data accessed through the Summary Care Record (Kjernejournal).
From enrollment until end of follow up (12 months after enrollment).
Time from enrollment to detection of first AF episode
Time Frame: From enrollment up to 1 year
Time to detection of first episode of atrial fibrillation after hospital admission
From enrollment up to 1 year
Total duration of heart rate monitoring
Time Frame: From enrollment up to 1 year
Total duration of heart rate monitoring, also including an assessment of device signal quality for assessment
From enrollment up to 1 year
Numbers of participants with secondary stroke/TIA and/or major bleeding
Time Frame: From enrollment to one-year follow up, assessed at 1-year follow up.
Assessing secondary stroke and major bleeding at 1 year of follow-up
From enrollment to one-year follow up, assessed at 1-year follow up.

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Explore the practical utility in regards to validity of results of AF screening after a stroke with a patch ECG
Time Frame: After 1 year of recruitment
Inter-rater validity of ECG results between the ECG247 algorithm, cardiologists and neurologists
After 1 year of recruitment
Explore the practical utility of time consumed using AF screening after a stroke with a patch ECG
Time Frame: After 1 year of recruitment
A suitable number of observations will be performed to evaluate the time needed to complete the patch ECG attachment and inform the patient, interpret the results and follow-up of results
After 1 year of recruitment
Explore the practical utility and reliability of AF screening after a stroke with a patch ECG
Time Frame: Assessed at 1 year after enrollment start
Reliability of the monitoring device in crude versus interpretable monitoring duration in hours. Reliability will be assessed by the study personnel when creating the patch ECG result record. The whole pulse rate strip will be analyzed to identify periods of 1>h with a lack of signal or reduced signal quality. The total monitoring duration registered by the algorithm in hours versus manual interpretation with readable signal in hours will be evaluated as a percentage: Useable monitoring time/Total monitoring time x 100% = Percentage of useful monitoring.
Assessed at 1 year after enrollment start
Explore the practical utility and patient satisfaction of AF screening after a stroke with a patch ECG
Time Frame: Through study completion
Patient Reported Experience Measures (PREMs) to all patients in interventional arm within a month of discharge and within 1 month of Holter completion for control arm, registration of technical problems, allergic reactions to patch, photo of patch placement and registration of patients not considered eligible to the study due to lack of smartphone and other causes not listed in the exclusion/inclusion criteria. PREMS will be reported by a 7 point Likert scale, with higher values indicating a high level of satisfaction.
Through study completion
Diagnostic quality of in-house patch ECG compared to telemetry
Time Frame: From start of enrollment until discharge of last included patient

Sensitivity and specificity of AF detection in patch ECG compared to concomitant telemetry.

Assessment of detection of other dangerous arrhythmias in patch ECG compared to telemetry.

Quality of signal from telemetry compared to patch ECG. In the subgroup of patients where patch ECG and telemetry are worn simultaneously, the results of both methods will be cross-validated against each other

From start of enrollment until discharge of last included patient
Cost effectiveness of patch ECG compared to standard care
Time Frame: From enrollment until end of follow up.
Cost-effectiveness analysis of patch ECG and standard care with the endpoint being reduced secondary stroke
From enrollment until end of follow up.
Assess predictors of silent AF
Time Frame: From enrollment until end of patient follow-up

Association between clinical, radiological and laboratory factors (including) NT-pro-BNP level and AF NT-proBNP will be assessed upon inclusion and can be analyzed from blood samples drawn on day 1 after admission.

MRI results Echocardiography results CHADSVASC, age, gender, medication used, previous diseases, BMI

From enrollment until end of patient follow-up

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

January 10, 2026

Primary Completion (Estimated)

December 1, 2028

Study Completion (Estimated)

December 1, 2029

Study Registration Dates

First Submitted

January 25, 2026

First Submitted That Met QC Criteria

February 5, 2026

First Posted (Actual)

February 12, 2026

Study Record Updates

Last Update Posted (Actual)

February 12, 2026

Last Update Submitted That Met QC Criteria

February 5, 2026

Last Verified

December 1, 2025

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

UNDECIDED

IPD Plan Description

Currently undecided regarding the sharing of IPD. No established formal cooperation agreements with external research groups to date. Any future decision to share data will depend on the establishment of such collaboration agreements in adherence to data protection regulations.

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

product manufactured in and exported from the U.S.

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.