Standardizace měření tlaku při polykání
Standardizace metrik faryngeální vysokorozlišovací manometrie
This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.
Přehled studie
Postavení
Postavení
Podmínky
Podmínky
Intervence / Léčba
Intervence / Léčba
Detailní popis
The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.
Typ studie
Typ studie
Zápis (Odhadovaný)
Zápis
Fáze
Fáze
- Nelze použít
Kontakty a umístění
Studijní kontakt
Studijní kontakt
- Jméno: Suzan Abdelhalim, MD
- Telefonní číslo: 608-265-2470
- E-mail: abdelhalim@surgery.wisc.edu
Studijní místa
-
-
Wisconsin
-
Madison, Wisconsin, Spojené státy, 53792
- University of Wisconsin
-
-
Kritéria účasti
Kritéria způsobilosti
Kritéria způsobilosti
Věk způsobilý ke studiu
Přijímá zdravé dobrovolníky
Pohlaví způsobilá ke studiu
Popis
Inclusion Criteria:
- without known swallowing disorders
Exclusion Criteria:
- Known swallowing disorder
- Allergy to topical anesthetic
- Allergy to food relevant to study participation (e.g. lactose intolerance)
- Inability to follow direction or remain calm during the experiment.
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Diagnostický
- Přidělení: Randomizované
- Intervenční model: Crossover Assignment
- Maskování: Žádné (otevřený štítek)
Počet zbraní
Zbraně a zásahy
Skupina účastníků / ArmSkupina účastníků / Arm |
Intervence / LéčbaIntervence / Léčba |
|---|---|
|
Experimentální: Laborie then Medtronic
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Ostatní jména:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Ostatní jména:
|
|
Experimentální: Medtronic then Laborie
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Ostatní jména:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Ostatní jména:
|
Co je měření studie?
Primární výstupní opatření
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Base of tongue pressure
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Hypopharyngeal peak pressure
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pharyngeal Contractile Integral (PhCI)
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure rise time
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Contractile front velocity
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Upper Esophageal Sphincter (UES) resting pressure
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES peak contraction pressure
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation nadir pressure
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation duration
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Nadir impedance
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Bolus presence time
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Flow interval
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure-flow index component
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Timing relationships (onset-to-peak intervals)
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Construction of Spatial pressure distribution maps
Časové okno: data collected during 1 study visit (up to 2 hours)
|
Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure. |
data collected during 1 study visit (up to 2 hours)
|
Spolupracovníci a vyšetřovatelé
Sponzor
Sponzor
Vyšetřovatelé
Vyšetřovatelé
- Vrchní vyšetřovatel: Timothy McCulloch, MD, University of Wisconsin, Madison
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Odhadovaný)
Začátek studia
Primární dokončení (Odhadovaný)
Primární dokončení
Dokončení studie (Odhadovaný)
Dokončení studie
Termíny zápisu do studia
První předloženo
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
První zveřejněno
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Poslední zveřejněná aktualizace
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
Další relevantní podmínky MeSH
Další identifikační čísla studie
Další identifikační čísla studie
- 2025-1642
- SMPH | Otolaryngology (Jiný identifikátor: UW Madison)
- Protocol Version 4/28/26 (Jiný identifikátor: UW Madison)
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Typ podpůrných informací pro sdílení IPD
- PROTOKOL STUDY
- MÍZA
- ICF
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
produkt vyrobený a vyvážený z USA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .