Standardizzazione delle Misurazioni della Pressione della Deglutizione
Standardizzazione delle Metriche della Manometria ad Alta Risoluzione Faringea
This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.
Panoramica dello studio
Stato
Stato
Condizioni
Condizioni
Intervento / Trattamento
Intervento / Trattamento
Descrizione dettagliata
The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.
Tipo di studio
Tipo di studio
Iscrizione (Stimato)
Iscrizione
Fase
Fase
- Non applicabile
Contatti e Sedi
Contatto studio
Contatto studio
- Nome: Suzan Abdelhalim, MD
- Numero di telefono: 608-265-2470
- Email: abdelhalim@surgery.wisc.edu
Luoghi di studio
-
-
Wisconsin
-
Madison, Wisconsin, Stati Uniti, 53792
- University of Wisconsin
-
-
Criteri di partecipazione
Criteri di ammissibilità
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
- without known swallowing disorders
Exclusion Criteria:
- Known swallowing disorder
- Allergy to topical anesthetic
- Allergy to food relevant to study participation (e.g. lactose intolerance)
- Inability to follow direction or remain calm during the experiment.
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Diagnostico
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione incrociata
- Mascheramento: Nessuno (etichetta aperta)
Numero di armi
Armi e interventi
Gruppo di partecipanti / ArmGruppo di partecipanti / Arm |
Intervento / TrattamentoIntervento / Trattamento |
|---|---|
|
Sperimentale: Laborie then Medtronic
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Altri nomi:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Altri nomi:
|
|
Sperimentale: Medtronic then Laborie
This study will include a data collection session that involves swallowing tasks during HRM. |
The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.
Altri nomi:
The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.
Altri nomi:
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Base of tongue pressure
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Hypopharyngeal peak pressure
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pharyngeal Contractile Integral (PhCI)
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure rise time
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Contractile front velocity
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Upper Esophageal Sphincter (UES) resting pressure
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES peak contraction pressure
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation nadir pressure
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
UES relaxation duration
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Nadir impedance
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Bolus presence time
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Flow interval
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Pressure-flow index component
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Timing relationships (onset-to-peak intervals)
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison. |
data collected during 1 study visit (up to 2 hours)
|
|
Construction of Spatial pressure distribution maps
Lasso di tempo: data collected during 1 study visit (up to 2 hours)
|
Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure. |
data collected during 1 study visit (up to 2 hours)
|
Collaboratori e investigatori
Sponsor
Sponsor
Investigatori
Investigatori
- Investigatore principale: Timothy McCulloch, MD, University of Wisconsin, Madison
Studiare le date dei record
Studia le date principali
Inizio studio (Stimato)
Inizio studio
Completamento primario (Stimato)
Completamento primario
Completamento dello studio (Stimato)
Completamento dello studio
Date di iscrizione allo studio
Primo inviato
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Primo Inserito
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento pubblicato
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Parole chiave
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
Altri numeri di identificazione dello studio
- 2025-1642
- SMPH | Otolaryngology (Altro identificatore: UW Madison)
- Protocol Version 4/28/26 (Altro identificatore: UW Madison)
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Tipo di informazioni di supporto alla condivisione IPD
- STUDIO_PROTOCOLLO
- LINFA
- ICF
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
prodotto fabbricato ed esportato dagli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .