Standardisering av svelgetrykkmålinger

29. oktober 2026 oppdatert av: University of Wisconsin, Madison

Standardisering av måleparametre for faryngeal høyoppløselig manometri

This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.

Studieoversikt

Status

Har ikke rekruttert ennå

Forhold

Intervensjon / Behandling

Detaljert beskrivelse

The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.

Studietype

Intervensjonell

Registrering (Antatt)

25

Fase

  • Ikke aktuelt

Kontakter og plasseringer

Denne delen inneholder kontaktinformasjon for de som utfører studien, og informasjon om hvor denne studien blir utført.

Studiekontakt

Studiesteder

    • Wisconsin
      • Madison, Wisconsin, Forente stater, 53792
        • University of Wisconsin

Deltakelseskriterier

Forskere ser etter personer som passer til en bestemt beskrivelse, kalt kvalifikasjonskriterier. Noen eksempler på disse kriteriene er en persons generelle helsetilstand eller tidligere behandlinger.

Kvalifikasjonskriterier

Alder som er kvalifisert for studier

18 år til 99 år (Voksen, Eldre voksen)

Tar imot friske frivillige

Ja

Kjønn som er kvalifisert for studier

Alle

Beskrivelse

Inclusion Criteria:

  • without known swallowing disorders

Exclusion Criteria:

  • Known swallowing disorder
  • Allergy to topical anesthetic
  • Allergy to food relevant to study participation (e.g. lactose intolerance)
  • Inability to follow direction or remain calm during the experiment.

Studieplan

Denne delen gir detaljer om studieplanen, inkludert hvordan studien er utformet og hva studien måler.

Hvordan er studiet utformet?

Designdetaljer

  • Primært formål: Diagnostisk
  • Tildeling: Randomisert
  • Intervensjonsmodell: Crossover-oppdrag
  • Masking: Ingen (Open Label)

Våpen og intervensjoner

Deltakergruppe / Arm
Intervensjon / Behandling
Eksperimentell: Laborie then Medtronic

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andre navn:
  • Laborie HRM impedanssystem
  • Laborie Solar™ GI HRM Faststoff, Trålesystem (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andre navn:
  • Medtronic HRM-impedanssystem
  • Medtronic ManoScan™ ESO høypresisjons manometrisystem
  • Medtronic ManoScan360 HRM-kateter
Eksperimentell: Medtronic then Laborie

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andre navn:
  • Laborie HRM impedanssystem
  • Laborie Solar™ GI HRM Faststoff, Trålesystem (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andre navn:
  • Medtronic HRM-impedanssystem
  • Medtronic ManoScan™ ESO høypresisjons manometrisystem
  • Medtronic ManoScan360 HRM-kateter

Hva måler studien?

Primære resultatmål

Resultatmål
Tiltaksbeskrivelse
Tidsramme
Base of tongue pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Hypopharyngeal peak pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pharyngeal Contractile Integral (PhCI)
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure rise time
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Contractile front velocity
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Upper Esophageal Sphincter (UES) resting pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES peak contraction pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation nadir pressure
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation duration
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Nadir impedance
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Bolus presence time
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Flow interval
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure-flow index component
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Timing relationships (onset-to-peak intervals)
Tidsramme: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Construction of Spatial pressure distribution maps
Tidsramme: data collected during 1 study visit (up to 2 hours)

Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure.

data collected during 1 study visit (up to 2 hours)

Samarbeidspartnere og etterforskere

Det er her du vil finne personer og organisasjoner som er involvert i denne studien.

Sponsor

Etterforskere

  • Hovedetterforsker: Timothy McCulloch, MD, University of Wisconsin, Madison

Studierekorddatoer

Disse datoene sporer fremdriften for innsending av studieposter og sammendragsresultater til ClinicalTrials.gov. Studieposter og rapporterte resultater gjennomgås av National Library of Medicine (NLM) for å sikre at de oppfyller spesifikke kvalitetskontrollstandarder før de legges ut på det offentlige nettstedet.

Studer hoveddatoer

Studiestart (Antatt)

1. november 2026

Primær fullføring (Antatt)

1. september 2027

Studiet fullført (Antatt)

1. september 2027

Datoer for studieregistrering

Først innsendt

11. mars 2026

Først innsendt som oppfylte QC-kriteriene

11. mars 2026

Først lagt ut (Faktiske)

16. mars 2026

Oppdateringer av studieposter

Sist oppdatering lagt ut (Faktiske)

2. november 2026

Siste oppdatering sendt inn som oppfylte QC-kriteriene

29. oktober 2026

Sist bekreftet

1. oktober 2026

Mer informasjon

Begreper knyttet til denne studien

Andre studie-ID-numre

  • 2025-1642
  • SMPH | Otolaryngology (Annen identifikator: UW Madison)
  • Protocol Version 4/28/26 (Annen identifikator: UW Madison)

Plan for individuelle deltakerdata (IPD)

Planlegger du å dele individuelle deltakerdata (IPD)?

JA

IPD-deling Støtteinformasjonstype

  • STUDY_PROTOCOL
  • SEVJE
  • ICF

Legemiddel- og utstyrsinformasjon, studiedokumenter

Studerer et amerikansk FDA-regulert medikamentprodukt

Nei

Studerer et amerikansk FDA-regulert enhetsprodukt

Ja

produkt produsert i og eksportert fra USA

Ja

Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .