Padronização das Medições de Pressão de Deglutição

29 de outubro de 2026 atualizado por: University of Wisconsin, Madison

Normalização das Métricas da Manometria de Alta Resolução Faríngea

This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.

Visão geral do estudo

Status

Ainda não está recrutando

Condições

Intervenção / Tratamento

Descrição detalhada

The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.

Tipo de estudo

Intervencional

Inscrição (Estimado)

25

Estágio

  • Não aplicável

Contactos e Locais

Esta seção fornece os detalhes de contato para aqueles que conduzem o estudo e informações sobre onde este estudo está sendo realizado.

Contato de estudo

Locais de estudo

    • Wisconsin
      • Madison, Wisconsin, Estados Unidos, 53792
        • University of Wisconsin

Critérios de participação

Os pesquisadores procuram pessoas que se encaixem em uma determinada descrição, chamada de critérios de elegibilidade. Alguns exemplos desses critérios são a condição geral de saúde de uma pessoa ou tratamentos anteriores.

Critérios de elegibilidade

Idades elegíveis para estudo

18 anos a 99 anos (Adulto, Adulto mais velho)

Aceita Voluntários Saudáveis

Sim

Gêneros Elegíveis para o Estudo

Tudo

Descrição

Inclusion Criteria:

  • without known swallowing disorders

Exclusion Criteria:

  • Known swallowing disorder
  • Allergy to topical anesthetic
  • Allergy to food relevant to study participation (e.g. lactose intolerance)
  • Inability to follow direction or remain calm during the experiment.

Plano de estudo

Esta seção fornece detalhes do plano de estudo, incluindo como o estudo é projetado e o que o estudo está medindo.

Como o estudo é projetado?

Detalhes do projeto

  • Finalidade Principal: Diagnóstico
  • Alocação: Randomizado
  • Modelo Intervencional: Atribuição cruzada
  • Mascaramento: Nenhum (rótulo aberto)

Armas e Intervenções

Grupo de Participantes / Braço
Intervenção / Tratamento
Experimental: Laborie then Medtronic

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Outros nomes:
  • Sistema de impedância HRM Laborie
  • Laborie Solar™ GI HRM de Estado Sólido, Sistema de Carro (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Outros nomes:
  • Sistema de impedância da HRM da Medtronic
  • Sistema de manometria de alta resolução ESO Medtronic ManoScan™
  • Cateter Medtronic ManoScan360 HRM
Experimental: Medtronic then Laborie

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Outros nomes:
  • Sistema de impedância HRM Laborie
  • Laborie Solar™ GI HRM de Estado Sólido, Sistema de Carro (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Outros nomes:
  • Sistema de impedância da HRM da Medtronic
  • Sistema de manometria de alta resolução ESO Medtronic ManoScan™
  • Cateter Medtronic ManoScan360 HRM

O que o estudo está medindo?

Medidas de resultados primários

Medida de resultado
Descrição da medida
Prazo
Base of tongue pressure
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Hypopharyngeal peak pressure
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pharyngeal Contractile Integral (PhCI)
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure rise time
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Contractile front velocity
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Upper Esophageal Sphincter (UES) resting pressure
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES peak contraction pressure
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation nadir pressure
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation duration
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Nadir impedance
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Bolus presence time
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Flow interval
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure-flow index component
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Timing relationships (onset-to-peak intervals)
Prazo: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Construction of Spatial pressure distribution maps
Prazo: data collected during 1 study visit (up to 2 hours)

Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure.

data collected during 1 study visit (up to 2 hours)

Colaboradores e Investigadores

É aqui que você encontrará pessoas e organizações envolvidas com este estudo.

Patrocinador

Investigadores

  • Investigador principal: Timothy McCulloch, MD, University of Wisconsin, Madison

Datas de registro do estudo

Essas datas acompanham o progresso do registro do estudo e os envios de resumo dos resultados para ClinicalTrials.gov. Os registros do estudo e os resultados relatados são revisados ​​pela National Library of Medicine (NLM) para garantir que atendam aos padrões específicos de controle de qualidade antes de serem publicados no site público.

Datas Principais do Estudo

Início do estudo (Estimado)

1 de novembro de 2026

Conclusão Primária (Estimado)

1 de setembro de 2027

Conclusão do estudo (Estimado)

1 de setembro de 2027

Datas de inscrição no estudo

Enviado pela primeira vez

11 de março de 2026

Enviado pela primeira vez que atendeu aos critérios de CQ

11 de março de 2026

Primeira postagem (Real)

16 de março de 2026

Atualizações de registro de estudo

Última Atualização Postada (Real)

2 de novembro de 2026

Última atualização enviada que atendeu aos critérios de controle de qualidade

29 de outubro de 2026

Última verificação

1 de outubro de 2026

Mais Informações

Termos relacionados a este estudo

Outros números de identificação do estudo

  • 2025-1642
  • SMPH | Otolaryngology (Outro identificador: UW Madison)
  • Protocol Version 4/28/26 (Outro identificador: UW Madison)

Plano para dados de participantes individuais (IPD)

Planeja compartilhar dados de participantes individuais (IPD)?

SIM

Tipo de informação de suporte de compartilhamento de IPD

  • PROTOCOLO DE ESTUDO
  • SEIVA
  • CIF

Informações sobre medicamentos e dispositivos, documentos de estudo

Estuda um medicamento regulamentado pela FDA dos EUA

Não

Estuda um produto de dispositivo regulamentado pela FDA dos EUA

Sim

produto fabricado e exportado dos EUA

Sim

Essas informações foram obtidas diretamente do site clinicaltrials.gov sem nenhuma alteração. Se você tiver alguma solicitação para alterar, remover ou atualizar os detalhes do seu estudo, entre em contato com register@clinicaltrials.gov. Assim que uma alteração for implementada em clinicaltrials.gov, ela também será atualizada automaticamente em nosso site .