Standaardiseren van Slikdrukmetingen

29 oktober 2026 bijgewerkt door: University of Wisconsin, Madison

Standaardisatie van metingen bij faryngale hoge-resolutiemanometrie

This study aims to directly compare and contrast swallowing pressure and impedance measurements obtained from two high-resolution manometry (HRM) systems: the Laborie LMT Pharyngeal System and the Medtronic ManoScan ESO HRM System in healthy adult volunteers. The primary objective is to determine how differences in catheter diameter, sensor configuration, and acquisition platforms influence measurement outcomes. Twenty-five healthy adults will participate in a single, approximately 2-hour visit; the study will be open for one year for data collection and two years for data analysis.

Studie Overzicht

Toestand

Nog niet aan het werven

Conditie

Interventie / Behandeling

Gedetailleerde beschrijving

The primary objective is to obtain, compare, and contrast directly the measured swallow between the Laborie Pharyngeal HRM impedance system with the 3.7 mm catheter and the Medtronic HRM impedance system with the 4.2 mm catheter in the same setting with normal healthy volunteers.

Studietype

Ingrijpend

Inschrijving (Geschat)

25

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studiecontact

Studie Locaties

    • Wisconsin
      • Madison, Wisconsin, Verenigde Staten, 53792
        • University of Wisconsin

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

18 jaar tot 99 jaar (Volwassen, Oudere volwassene)

Accepteert gezonde vrijwilligers

Ja

Geslachten die in aanmerking komen voor studie

Allemaal

Beschrijving

Inclusion Criteria:

  • without known swallowing disorders

Exclusion Criteria:

  • Known swallowing disorder
  • Allergy to topical anesthetic
  • Allergy to food relevant to study participation (e.g. lactose intolerance)
  • Inability to follow direction or remain calm during the experiment.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Diagnostisch
  • Toewijzing: Gerandomiseerd
  • Interventioneel model: Crossover-opdracht
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Laborie then Medtronic

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andere namen:
  • Laborie HRM-impedantiesysteem
  • Laborie Solar™ GI HRM Solid-State, Trolley System (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andere namen:
  • Medtronic HRM-impedantiesysteem
  • Medtronic ManoScan™ ESO hoogresolutie manometriesysteem
  • Medtronic ManoScan360 HRM-katheter
Experimenteel: Medtronic then Laborie

This study will include a data collection session that involves swallowing tasks during HRM.

The Solar GI HRM system is intended to record, store, view, and analyze pressure, impedance, EMG, swallow and respiration, and various auxiliary input device data online anywhere in the gastrointestinal tract (pharynx; esophagus; stomach; duodenum; sphincter of Oddi; small bowel; colon and anorectal area, including rectum and pelvic floor) to assist in the diagnosis and evaluation of gastrointestinal and swallowing disorders. Used with 3.7 mm catheter.

Andere namen:
  • Laborie HRM-impedantiesysteem
  • Laborie Solar™ GI HRM Solid-State, Trolley System (G3-9)
  • Laborie Solid State HRIM Catheter (K83659-E-1820-D)

The ManoScan™ ESO high resolution manometry system enables full evaluation of the motor functions of the esophagus. The system provides useful information to support diagnosis of conditions like dysphagia, achalasia, and hiatal hernia. A 36-channel circumferential solid-state catheter with 4.2 mm diameter, 1.0 cm spacing between adjacent sensors.

Andere namen:
  • Medtronic HRM-impedantiesysteem
  • Medtronic ManoScan™ ESO hoogresolutie manometriesysteem
  • Medtronic ManoScan360 HRM-katheter

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Base of tongue pressure
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Hypopharyngeal peak pressure
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pharyngeal Contractile Integral (PhCI)
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure rise time
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Contractile front velocity
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Upper Esophageal Sphincter (UES) resting pressure
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES peak contraction pressure
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation nadir pressure
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
UES relaxation duration
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Nadir impedance
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Bolus presence time
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Flow interval
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Pressure-flow index component
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Timing relationships (onset-to-peak intervals)
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Measured with both Laborie HRM impedance system and Medtronic HRM impedance system for comparison.

data collected during 1 study visit (up to 2 hours)
Construction of Spatial pressure distribution maps
Tijdsspanne: data collected during 1 study visit (up to 2 hours)

Multi-level modeling for repeated measures across bolus types, volumes, and repetitions. Sensor-level and region-level analyses (BOT, hypopharynx, UES). Agreement thresholds and bias definitions added to facilitate interpretability and ensure clarity for cross-platform comparison. Map layers will be summarized as a proxy for this measure.

data collected during 1 study visit (up to 2 hours)

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Sponsor

Onderzoekers

  • Hoofdonderzoeker: Timothy McCulloch, MD, University of Wisconsin, Madison

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Geschat)

1 november 2026

Primaire voltooiing (Geschat)

1 september 2027

Studie voltooiing (Geschat)

1 september 2027

Studieregistratiedata

Eerst ingediend

11 maart 2026

Eerst ingediend dat voldeed aan de QC-criteria

11 maart 2026

Eerst geplaatst (Werkelijk)

16 maart 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

2 november 2026

Laatste update ingediend die voldeed aan QC-criteria

29 oktober 2026

Laatst geverifieerd

1 oktober 2026

Meer informatie

Termen gerelateerd aan deze studie

Andere studie-ID-nummers

  • 2025-1642
  • SMPH | Otolaryngology (Andere identificatie: UW Madison)
  • Protocol Version 4/28/26 (Andere identificatie: UW Madison)

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

JA

IPD delen Ondersteunend informatietype

  • LEERPROTOCOOL
  • SAP
  • ICF

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Ja

product vervaardigd in en geëxporteerd uit de V.S.

Ja

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .