Safety and Efficacy of Medtronic Extended Mio 30 Infusion Set

September 15, 2022 updated by: Prof. Amir Tirosh, Sheba Medical Center

The purpose of this study is to evaluate the Extended Mio 30 infusion set (EM30IS) 7-day functioning compared to the regular Mio 30 infusion set (M30IS) 3-day functioning. The study will be carried out in adult patients with type 1 diabetes.

Secondary objective, to evaluate the differences in glucose control between the infusion set currently used by participants (Medtronic Quick set infusion sets) and the Medtronic Mio 30 Infusion Set.

Study Overview

Status

Completed

Detailed Description

This is a 1-center, prospective, open-label, cross-over controlled study with type 1 patients with diabetes on insulin pump therapy with Continuous Glucose Monitoring (CGM).

A total of up to 40 subjects will be enrolled in order to have 20 subjects meeting eligibility criteria. Eligible patients meeting the enrollment and randomization criteria will enter a run-in period of up to 1 month for device adaptation. After the run- in phase, The participants will enter the first study part; participants will be randomized into 2 arms, each arm will undergo 2 phases. In each phase a different type of infusion set will be evaluated. Each evaluation will be for 1 month. The order will be determined by the randomization process:

Arm A: Quick set infusion set- M30IS. First phase Quick set, second phase Mio 30.

Arm B: M30IS- Quick set infusion set. First phase Mio 30, second phase Quick set infusion set.

Trial Arm A: Subjects will be randomized to a group who will be using the Quick set infusion set for the initial Phase. All patients will be trained on the use of the Quick set infusion set by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another Quick set infusion set).

After 1 month, the patients will return to a visit, return all the extracted catheters sets and will switch to the M30IS, entering Phase 2. All patients will then be retrained on the use of the M30IS by site staff and demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS). 1 month, the patients will return all the extracted catheters sets.

Trial Arm B: Subjects will be randomized to a group who will be using the Mio 30 infusion set (M30IS) for the second Phase. All patients will be retrained on the use of the M30IS by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS).

After 1 month, the patients will return to a visit, return all the extracted catheters sets and will switch to the Quick set infusion set, entering Phase 2. All patients will be trained at this visit on the use of the Quick set infusion set by site staff and demonstrate proficiency in the use of the Quick set infusion set. All subjects will be instructed to change sets every 3 days or at set failure (replace with another Quick set infusion set). 1 month, the patients will return all the extracted catheters sets

After completing the first study part, participants will be randomized again into 2 arms, each arm will undergo 2 phases. In each phase a different type of infusion set will be evaluated. Each evaluation will be for 4 sets (up to 28 days). The order will be determined by the randomization process:

  • Arm A: M30IS - EM30IS. First phase Mio 30, second phase Extended Mio 30.
  • Arm B: EM30IS - M30IS. First phase Extended Mio 30, second phase Mio 30. Trial Arm A: Subjects will be randomized to a group who will be using the Mio 30 infusion set (M30IS) for the Phase 1. All patients will be retrained on the use of the M30IS by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS).

At day 12 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the Extended Mio 30 infusion set (EM30IS), entering Phase 2. All patients will be trained at this visit on the use of the EM30IS by site staff and demonstrate proficiency in the use of the EM30IS. All subjects will be instructed to change sets every 7 days or at set failure (replace with another EM30IS). After 28 days or after using 4 sets, the patients will return all the extracted catheters sets.

Trial Arm B: Subjects will be randomized to a group who will be using the EM30IS for the initial Phase. All patients will be trained on the use of the EM30IS by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 7 days or at set failure (replace with another EM30IS).

At day 28 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the M30IS, entering Phase 2. All patients will then be retrained on the use of the M30IS by site staff and demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS). After 12 days or after using 4 sets, the patients will all the extracted catheters sets.

Each subject will wear MiniMed™ 780G/670G insulin system. Each subject will be given for the first study part: 11 infusion sets each to wear, each infusion set for at least the labelled use time (3 days for M30IS or 3 days for Quick set infusin set), or until infusion set failure if this occurs before 3 days.for the second part 4 infusion sets each to wear, each infusion set for at least the labelled use time (3 days for M30IS or 7 days for EM30IS), or until infusion set failure if this occurs before 3 days.

Study Type

Interventional

Enrollment (Actual)

20

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Ramat gan, Israel, 52662
        • Sheba Medical Center
      • Tel-Hashomer, Israel
        • Sheba Medical Center

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 80 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Clinical diagnosis of type 1 diabetes for at least one year
  • Age 18 to 80 years
  • Not currently known to be pregnant, nor planning pregnancy during the study.
  • Willingness to follow the protocol and sign the informed consent

Exclusion Criteria:

  • A known medical condition that in the judgment of the investigator might interfere with the completion of the protocol.
  • Pregnant or lactating females

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: A - M30IS vs. EM30IS. First phase Mio 30, second phase Extended Mio 30.

Subjects will be randomized to a group who will be using the Mio 30 infusion set (M30IS) for the Phase 1. All patients will be retrained on the use of the M30IS by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS).

At day 12 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the Extended Mio 30 infusion set (EM30IS), entering Phase 2. All patients will be trained at this visit on the use of the EM30IS by site staff and demonstrate proficiency in the use of the EM30IS. All subjects will be instructed to change sets every 7 days or at set failure (replace with another EM30IS). After 28 days or after using 4 sets, the patients will return all the extracted catheters sets.

Safety and Efficacy of Medtronic Extended Mio 30 Infusion Set
Other Names:
  • Medtronic Mio 30 Infusion Set
  • Medtronic Quick set infusion set
Experimental: B - EM30IS vs. M30IS. First phase Extended Mio 30, second phase Mio 30

Subjects will be randomized to a group who will be using the EM30IS for the initial Phase. All patients will be trained on the use of the EM30IS by site staff and will be asked to demonstrate proficiency. All subjects will be instructed to change sets every 7 days or at set failure (replace with another EM30IS).

At day 28 or after using 4 sets (set used defined as a set that was used for more than 6 hours), the patients will return to a visit, return all the extracted catheters sets and will switch to the M30IS, entering Phase 2. All patients will then be retrained on the use of the M30IS by site staff and demonstrate proficiency. All subjects will be instructed to change sets every 3 days or at set failure (replace with another M30IS). After 12 days or after using 4 sets, the patients will all the extracted catheters sets.

Safety and Efficacy of Medtronic Extended Mio 30 Infusion Set
Other Names:
  • Medtronic Mio 30 Infusion Set
  • Medtronic Quick set infusion set

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Percentage of sets that failed due to unexplained hyperglycemia
Time Frame: 6 months
The primary outcome is to have no more than 16% of infusion set failures due to "unexplained hyperglycemia"
6 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

August 4, 2022

Primary Completion (Actual)

August 4, 2022

Study Completion (Actual)

August 4, 2022

Study Registration Dates

First Submitted

September 8, 2022

First Submitted That Met QC Criteria

September 15, 2022

First Posted (Actual)

September 16, 2022

Study Record Updates

Last Update Posted (Actual)

September 16, 2022

Last Update Submitted That Met QC Criteria

September 15, 2022

Last Verified

September 1, 2022

More Information

Terms related to this study

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes Mellitus, Type 1

Clinical Trials on Medtronic Extended Mio 30 Infusion Set

3