Confronto tra pulsossimetria (SpO2) e diversi ossimetri e saturazione arteriosa (SaO2): studio Oxygap2 (Oxygap 2)
The oximeter is used to monitor intensive care patients undergoing oxygen therapy. It indicates pulsed oxygen saturation (SpO2), a reflection of arterial oxygen saturation (SaO2) which enables detection of hypoxemia and hyperoxia, both deleterious state. Current SpO2 recommendations aim to reduce both risk of hypoxemia and hyperoxia. SpO2 is considered the 5th vital sign.
Current recommendations for SpO2 targets do not consider the variability of oximeters used in clinical practice. This variability and lack of specification represent an obstacle to an optimal practice of oxygen therapy.
Thus, this study aims to compare the SpO2 values of different oximeters (General Electric-GE, Medtronic, Masimo and Nonin) used in clinical practice with the SaO2 reference value obtained by an arterial gas in order to specify the precision and the systematic biases of the oximeters studied. This data will also make it possible to refine the recommendations concerning optimal oxygenation
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Tipo di studio
Iscrizione (Stimato)
Contatti e Sedi
Luoghi di studio
-
-
Quebec
-
Montreal, Quebec, Canada, H3T 1C5
- Reclutamento
- CHU Ste-Justine
-
Contatto:
- Michael Sauthier
- Numero di telefono: 514 345-4931
- Email: Michael.Sauthier.med@ssss.gouv.qc.ca
-
Montreal, Quebec, Canada, H2X0A9
- Completato
- CHUM
-
Québec, Quebec, Canada, G1V 4G5
- Reclutamento
- Institut Universitaire de Cardiologie et de Pneumologie de Québec
-
Contatto:
- Francois Lellouche
- Numero di telefono: 3572 1-418-656-8711
- Email: francois.lellouche@criucpq.ulaval.ca
-
-
-
-
-
Les Abymes, Guadalupa, 97139
- Non ancora reclutamento
- CHU de la Guadeloupe
-
Contatto:
- Frederic Martino
- Numero di telefono: : +590 590 89 11 32
- Email: frederic.martino@chu-guadeloupe.fr
-
-
-
-
-
Sousse, Tunisia, 4000
- Non ancora reclutamento
- Hôpital Universitaire Farhat Hached,
-
Contatto:
- Mohamed Boussarsar
- Numero di telefono: +216 73 102 500
- Email: hamadi.boussarsar@gmail.com
-
-
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Metodo di campionamento
Popolazione di studio
All patient in ICU with arterial catheter already installed will be consider
Descrizione
Adult population
Inclusion Criteria:
- ≥ 18 years old (adult population) -
- Patients admitted to the Intensive Care Unit
- Artery catheter already installed
Pediatric population
- Below 18 years old
- Patients admitted to the Intensive Care Unit
- Artery catheter already installed
Exclusion Criteria:
- No or poor signal with the usual pulse oximeter/based on clinician judgment
- High dose of vasopressors or inotropes (epinephrine or norepinephrine ˃ 1mcg/kg/min), shock state (lactates above 3 mmoles/L)
- Pigmented nails or nail polish
- Methemoglobinemia history
- Hemoglobin below 80 g/L
- Patient in isolation (multi-resistant bacteria, C-Difficile, SARS-COV-2…)
- Prone position, Extra Corporel Membrane Oxygenator
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Modelli osservazionali: COHORT
- Prospettive temporali: PROSPECTIVE
Coorti e interventi
Gruppo / Coorte |
Intervento / Trattamento |
|---|---|
|
Pediatric population
Pediatric population < 18 years old (estimated 100 patients) |
Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded |
|
Adult population
Adult population >or= 18 years old (estimated 150 patients) |
Several pulse oximeter will be place on patient find few minutes before arterial blood gases. During an arterial blood sample, all SpO2 will be recorded |
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Difference between GE oximeter SpO2 value and the SaO2 value on arterial blood gases
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Difference between GE oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
|
Difference between Nellcor oximeter SpO2 value and the SaO2 value on arterial blood gases
Lasso di tempo: During procedure (arterial blood gassample collection)
|
Difference between Nellcor oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gassample collection)
|
|
Difference between Masimo oximeter SpO2 value and the SaO2 value on arterial blood gases
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Difference between Masimo oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
|
Difference between Nonin oximeter SpO2 value and the SaO2 value on arterial blood gases
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Difference between Nonin oximeter SpO2 and the SaO2 value |
During procedure (arterial blood gas sample collection)
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Overestimation and underestimation of SaO2
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Percent of pairs with overestimation and underestimation of SaO2 for each oximeter. |
During procedure (arterial blood gas sample collection)
|
|
Hypoxemia detection
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Ability to detect hypoxemia (SpO2 below 90% and PaO2 < 60 mmHg) |
During procedure (arterial blood gas sample collection)
|
Altre misure di risultato
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Hyperoxemia detection
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Ability to detect hyperoxemia (SpO2 above 95% and PaO2 > 100 mmHg) |
During procedure (arterial blood gas sample collection)
|
|
Gaps between the different oximeters
Lasso di tempo: During procedure (arterial blood gas sample collection)
|
Average of the SpO2 value gaps between the different oximeters tested |
During procedure (arterial blood gas sample collection)
|
Collaboratori e investigatori
Sponsor
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Stimato)
Completamento dello studio (Stimato)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- 22373
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .