Design Validation of Senofilcon A With New UV-blocking Additive
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Studietype
Inschrijving (Werkelijk)
Fase
- Niet toepasbaar
Contacten en locaties
Studie Locaties
-
-
California
-
Cupertino, California, Verenigde Staten, 95014
- James T. Fujimoto, OD
-
-
Florida
-
Maitland, Florida, Verenigde Staten, 32751
- Maitland Vision Center
-
Tallahassee, Florida, Verenigde Staten, 32308
- Tallahassee Eye Center
-
-
Illinois
-
Bloomington, Illinois, Verenigde Staten, 61701
- Eyecare Associates LLP
-
-
Kansas
-
Pittsburg, Kansas, Verenigde Staten, 66762
- Kannarr Eye Care
-
-
Ohio
-
Granville, Ohio, Verenigde Staten, 43023
- Procare Vision Centers
-
Westerville, Ohio, Verenigde Staten, 43081
- Professional Vision Care, Inc
-
-
Tennessee
-
Memphis, Tennessee, Verenigde Staten, 38111
- Bradley Hines, OD
-
-
Texas
-
Tyler, Texas, Verenigde Staten, 75703
- Frazier Vision, Inc
-
-
Wisconsin
-
New Berlin, Wisconsin, Verenigde Staten, 53151
- Ziegler Leffingwell Eyecare
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
Potential subjects must satisfy all of the following criteria to be enrolled in the study:
- The subject must read, understand, and sign the STATEMENT OF INFORMED CONSENT and receive a fully executed copy of the form.
- Appear able and willing to adhere to the instructions set forth in this clinical protocol.
- Between 18 and 49 (inclusive) years of age at the time of informed consent.
- Eligible presbyopes will be those that wear full distance contact lenses in both eyes, then wear reading glasses over them when needed for near vision.
- The subject is a current spherical soft contact lens wearer in both eyes with a minimum of 6 hours/day and 5 days/week wear time over the last 30 days by self-report.
- Subjects must be able and willing to wear the study lenses at least 6 hours a day, a minimum of 5 days per week.
- Subjects must own a wearable pair of distance spectacles.
- The subject's vertex-corrected spherical equivalent distance refraction must be in the range of -1.00 to -6.00 D (inclusive) in each eye.
- The subject's refractive cylinder must be 0.00 to -1.00 D (inclusive) in each eye.
The subject must have a spherocylindrical best corrected distance Snellen visual acuity of 20/25+3 or better in each eye.
Exclusion Criteria:
Potential subjects who meet any of the following criteria will be excluded from participating in the study:
- Currently pregnant or lactating (subjects who become pregnant during the study will be discontinued).
- Any active or ongoing systemic disease (e.g., Sjögren's Syndrome), infectious disease (e.g., hepatitis, tuberculosis), contagious immunosuppressive diseases (e.g., HIV), autoimmune disease (e.g. rheumatoid arthritis).
- Subjects taking suspect oral medications for less than one year.
- Any prescribed or over the counter (OTC) ocular medication.
- Any known hypersensitivity or allergic reaction to Optifree® PureMoist® multi-purpose care solution or Eye-Cept® rewetting drop solution.
- Toric, extended wear, monovision or multi-focal contact lens correction.
- Any previous or planned (during the course of the study) ocular surgery (e.g., PRK, LASIK, etc.).
- Participation in any contact lens or lens care product clinical trial within 30 days prior to study enrollment.
- Participation in clinical trials involving the Test lens within 3 months prior to study enrollment.
- Employee or immediate family member of an employee of clinical site (e.g., Investigator, Coordinator, Technician).
- Binocular vision abnormality or strabismus by self-report or prior medical history.
- History of recurrent corneal erosions, herpetic keratitis, or pathological dry eye.
- Any active ocular allergies, infections or other ocular abnormalities (entropion, ectropion, extrusions, chalazia, recurrent styes, glaucoma) that the investigator determines may interfere with the outcomes of this study or otherwise contraindicate participation in the study.
- Any Grade 2 or greater slit lamp findings (e.g., edema, corneal neovascularization, corneal staining, tarsal abnormalities, conjunctival injection) on the FDA Slit Lamp Classification Scale.
- Any previous history or signs of a contact lens-related corneal inflammatory event (e.g., past peripheral ulcer or round peripheral scar) within the past 3 years of otherwise successful contact lens wear.
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Behandeling
- Toewijzing: Gerandomiseerd
- Interventioneel model: Parallelle opdracht
- Masker: Enkel
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: senofilcon A TEST Lens
Subjects between 18 to 49 years of age that are habitual wearers of spherical soft contact lenses will be randomized into the TEST Lens for the duration of the clinical study.
|
JJVC Investigational Contact Lens
|
|
Actieve vergelijker: senofilcon A CONTROL Lens
Subjects between 18 to 49 years of age that are habitual wearers of spherical soft contact lenses will be randomized into the CONTROL Lens for the duration of the clinical study.
|
Acuvue Oasys
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Overall Comfort Score
Tijdsspanne: 2-Week Follow-up Evlauation
|
Overall comfort was assessed using the Contact Lens User Experience™ (CLUE) questionnaire.
CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patientexperience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65.
Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.
|
2-Week Follow-up Evlauation
|
|
Vision Satisfaction in Bright Lighting
Tijdsspanne: 2-Week Follow-up Evaluation
|
Vision satisfaction in bright light was assessed using the individual item" I was satisfied with the quality of my vision in bright lighting" from the CLUE™ questionnaire.
This item used the response scale, 1: Strongly Disagree, 2: Disagree, 3: Neither Agree nor Disagree, 4: Agree and 5: Strongly Agree.
CLUE is the Contact Lens User Experience™ questionnaire.
CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patient-experience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65.
Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.
The Proportion of responses in each category were reported for each lens type.
|
2-Week Follow-up Evaluation
|
|
Distance Visual Acuity
Tijdsspanne: 2-Week Follow-up Evlauation
|
Distance visual acuity was assessed by eye using LogMAR visual acuity at 4 meters using an ETDRS chart under high illumination high contrast conditions (room illumination > 400 lux and chart luminance 120-200 cd/m2) at the 2-week follow-up for each subject eye.
Lower values of logMAR indicate better vision, where a score of 0.0 equates to 20/20 snellen vision.
The average visual acuity for each lens was reported.
|
2-Week Follow-up Evlauation
|
|
Contact Lens Fitting Acceptance Rate
Tijdsspanne: Up to 2-Week Follow-up
|
Acceptable lens fit will be assessed at all study visits (scheduled and unscheduled) for each subject eye. Fit acceptance rate will be based on the lens fit acceptance of eyes wearing the Test lens only. Fit acceptance is a binary response where Y=1 if lens fit is acceptable and Y=0 otherwise. Unacceptable is defined as unacceptable if any one of the following criteria:
Eyes with multiple unacceptable fitting events will be counted only once. Fit rates of the Control lens will also be collected but are not a primary endpoint. |
Up to 2-Week Follow-up
|
|
Number of Grade 3 or Higher Slit Lamp Findings
Tijdsspanne: Up to 2-Week Follow-up Evlauation
|
Slit Lamp Findings (SLF) were assessed using a biomicroscope and was graded using the FDA grading scale (Grade: 0, 1,2, 3 and 4) with grade 0 represents the absence of findings and 1 to 4 representing successively worse findings (i.e.
Grade 1 = trace, Grade 2 = Mild, Grade 3 = moderate and Grade 4 = severe).
Slit Lamp Findings will be assessed for each subject eye at all study visits (schedule and unscheduled).
SLF is a binary response where Y=1 for at least one Grade 3 or 4 slit lamp finding.
The percentage of eyes with Grade 3 or higher slit lamp findings will be analyzed and will include corneal infiltrates.
Eyes with multiple events will be counted only once.
The number of eyes with SLF with grade 3 or higher by lens was reported.
|
Up to 2-Week Follow-up Evlauation
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Overall Handling Score
Tijdsspanne: 2-Week Follow-up Evlauation
|
Overall handling was assessed using the Contact Lens User Experience™ (CLUE) questionnaire.
CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patientexperience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65.
Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.
|
2-Week Follow-up Evlauation
|
|
Overall Quality of Vision Score
Tijdsspanne: 2-Week Follow-up Evlauation
|
Overall quality of vision was assessed using the Contact Lens User Experience™ (CLUE) questionnaire.
CLUE is a validated patient-reported outcomes (PRO) questionnaire to assess patientexperience attributes of soft contact lenses (comfort, vision, handling, and packaging) in a contact-lens wearing population in the US, ages 18-65.
Derived CLUE scores using Item Response Theory (IRT) follow a normal distribution with a population average score of 60 (SD 20), where higher scores indicate a more favorable/positive response with a range of 0-120.
|
2-Week Follow-up Evlauation
|
Medewerkers en onderzoekers
Studie record data
Bestudeer belangrijke data
Studie start (Werkelijk)
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Werkelijk)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Andere studie-ID-nummers
- CR-6305
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
Informatie over medicijnen en apparaten, studiedocumenten
Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel
Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct
product vervaardigd in en geëxporteerd uit de V.S.
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .
Klinische onderzoeken op Gezichtsscherpte
-
VisiblyVoltooidDigitaal Acuity-productVerenigde Staten
-
Burzynski Research InstituteVoltooidVisual Pathway GlioomVerenigde Staten
-
National Cancer Institute (NCI)Actief, niet wervendNeurofibromatose type 1 | Laaggradig glioom | Visual Pathway GlioomVerenigde Staten, Canada, Puerto Rico
-
National Cancer Institute (NCI)Actief, niet wervendLaaggradig glioom | Terugkerend visueel padglioom | Refractair visueel pad Glioom | Recidiverend pilocytisch astrocytoom bij kinderen | Terugkerende neurofibromatose type 1 | Refractaire neurofibromatose type 1Verenigde Staten
-
National Cancer Institute (NCI)VoltooidNeurofibromatose type 1 | Recidiverend primair neoplasma van het centrale zenuwstelsel | Refractair primair centraal zenuwstelsel neoplasma | Terugkerend hersenstamglioom bij kinderen | Terugkerende visuele glioom in de kindertijdVerenigde Staten
-
National Cancer Institute (NCI)VoltooidRefractair lymfoom | Refractair maligne solide neoplasma | Recidiverend lymfoom | Recidiverend maligne solide neoplasma | Recidiverend primair neoplasma van het centrale zenuwstelsel | Refractair primair centraal zenuwstelsel neoplasma | Hersenstam neoplasma | Neoplasma van de pijnappelklier | Terugkerend... en andere voorwaardenVerenigde Staten
-
National Cancer Institute (NCI)VoltooidNiet-gespecificeerde solide kindertumor, protocolspecifiek | Terugkerend zacht weefsel sarcoom bij kinderen | Terugkerend hersenstamglioom bij kinderen | Terugkerende visuele glioom in de kindertijd | Kiemceltumor van het centrale zenuwstelsel bij kinderen | Choriocarcinoom van het centrale... en andere voorwaardenVerenigde Staten
-
National Cancer Institute (NCI)VoltooidRecidiverend medulloblastoom bij kinderen | Recidiverend ependymoom bij kinderen | Atypische teratoïde / rhabdoïde tumor bij kinderen | Meningioom graad I bij kinderen | Meningioom graad II bij kinderen | Graad III meningeoom bij kinderen | Infratentoriaal ependymoom bij kinderen | Supratentoriaal... en andere voorwaardenVerenigde Staten
-
Children's Oncology GroupNational Cancer Institute (NCI)VoltooidRecidiverend medulloblastoom bij kinderen | Recidiverend ependymoom bij kinderen | Atypische teratoïde / rhabdoïde tumor bij kinderen | Ependymoom | Craniopharyngioma bij kinderen | Meningioom graad I bij kinderen | Meningioom graad II bij kinderen | Graad III meningeoom bij kinderen | Infratentoriaal... en andere voorwaardenVerenigde Staten, Canada, Australië, Nieuw-Zeeland, Puerto Rico, Zwitserland
-
Children's Oncology GroupNational Cancer Institute (NCI)VoltooidNiet-gespecificeerde solide kindertumor, protocolspecifiek | Recidiverend medulloblastoom bij kinderen | Recidiverend ependymoom bij kinderen | Recidiverend neuroblastoom | Recidiverend osteosarcoom | Terugkerende kinderrabdomyosarcoom | Eerder behandeld rabdomyosarcoom bij kinderen | Recidiverend... en andere voorwaardenVerenigde Staten
Klinische onderzoeken op senofilcon A TEST Lens
-
Johnson & Johnson Vision Care, Inc.Voltooid
-
Johnson & Johnson Vision Care, Inc.VoltooidCorrectie van brekingsfout | Verzwakking van fel lichtVerenigde Staten
-
Johnson & Johnson Vision Care, Inc.VoltooidOculaire fysiologieVerenigde Staten
-
CooperVision, Inc.VoltooidAstigmatisme BilateraalCanada
-
Johnson & Johnson Vision Care, Inc.Voltooid
-
CooperVision, Inc.Voltooid
-
Johnson & Johnson Vision Care, Inc.VoltooidVervallen contactlens slijtageVerenigde Staten
-
Johnson & Johnson Vision Care, Inc.VoltooidAmetropieVerenigde Staten
-
Johnson & Johnson Vision Care, Inc.VoltooidKlinische evaluatie van goedgekeurde en onderzoekscontactlenzen met behulp van een nieuw UV-additiefGezichtsscherpteVerenigde Staten
-
Johnson & Johnson Vision Care, Inc.Voltooid