Evaluation of a New Blood Glucose Meter System With Capillary and Venous Blood

29 januari 2016 uppdaterad av: Ascensia Diabetes Care

Clinical Evaluation of the Apollo Blood Glucose Monitoring System With Capillary and Venous Blood [Commercial Name is CONTOUR® USB]

The purpose of the study was to evaluate the performance of the meter system in the hands of subjects and healthcare professionals (HCPs). It also evaluated the product user guides in the hands of untrained subjects.

Studieöversikt

Status

Avslutad

Betingelser

Detaljerad beskrivning

The study evaluated the performance of the blood glucose meter system (BGMS) compared to a laboratory glucose method. Subjects and healthcare professionals tested subject capillary blood and healthcare professionals tested subject venous blood. Two meter configurations were evaluated. The study evaluated the acceptability of product labeling in enabling subjects to perform blood glucose testing with the new meter system and for using meter features. Subjects and healthcare professionals provided feedback about the BGMS and its features.

Studietyp

Interventionell

Inskrivning (Faktisk)

102

Fas

  • Inte tillämpbar

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

    • New Jersey
      • Fairfield, New Jersey, Förenta staterna, 07004
        • Consumer Product Testing Co., Inc.

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

18 år till 75 år (Vuxen, Äldre vuxen)

Tar emot friska volontärer

Nej

Kön som är behöriga för studier

Allt

Beskrivning

Inclusion Criteria:

  • Have type 1 or type 2 diabetes
  • Be > 18, < 76 year of age at time of consent, with approximately 50% (+10%)being less than 55 years of age
  • Be willing to complete all study procedures
  • Be routinely testing their blood sugar at home (at least once per day)
  • Be able to speak, read, and understand English and understand the Informed Consent document
  • Be able to read the labeling instructions

Exclusion Criteria:

  • Minors < 18 years of age and adults > 75 years of age
  • Pregnancy
  • Physical (dexterity), visual, or neurological impairments that would make the person unable to perform testing with the BGMS
  • Disorders in the fingertip lancing areas
  • Acute or chronic infections, particularly skin infections
  • Infection with a blood borne pathogen
  • Taking prescription anti-coagulants or having clotting problems that may prolong bleeding. Taking aspirin daily (81 mg or 325 mg) is not reason for exclusion
  • Hemophilia or any other bleeding disorder
  • Having a condition which, in the opinion of the Principal Investigator or designee, would put the person at risk or seriously compromise the integrity of the study
  • Working for a competitive medical device company

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Diagnostisk
  • Tilldelning: N/A
  • Interventionsmodell: Enskild gruppuppgift
  • Maskning: Ingen (Open Label)

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Övrig: Subjects with diabetes
Subjects with diabetes use a new blood glucose monitoring system (BGMS) Apollo Blood Glucose Monitoring System with capillary blood; healthcare professionals use the new BGMS with subject capillary and venous blood.
Subjects with diabetes use a new blood glucose monitoring system with capillary blood; healthcare professionals use the new blood glucose monitoring system with subject capillary and venous blood. All results are compared to a laboratory glucose method - Yellow Springs Instrument (YSI).
Andra namn:
  • CONTOUR® USB

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Number of Capillary and Venous Results Within +/- 15mg/dL or +/- 20% of Laboratory Glucose Method
Tidsram: One hour
Subjects with diabetes and healthcare professionals (HCPs) used a new blood glucose monitoring system (BGMS) with subject capillary blood. HCPs used the new bgms with subject venous blood. All results were compared to a lab glucose method. The BGMS has programmed algorithms to provide results equivalent to either serum/plasma or whole blood glucose methods. Number of results were obtained by combining results from three lots of Contour Blood Glucose strips.
One hour

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Percentage of Participants Rated as <=3 (Labeling Comprehension)
Tidsram: One hour

Study staff rated participants on their success at performing BG testing and Autolog feature after reading product labeling. The rating scale was:

  1. Successful
  2. Successful after being referred to user instructions
  3. Successful with verbal assistance or review of part of user instructions (Similar to review of a specific function during a Customer Service call.)
  4. Unsuccessful (Incorrectly performed part of the testing regimen or required intervention by study staff.)
One hour
Number of Partipants Who Gave These Ratings for Overall Testing Experience With This Meter
Tidsram: One hour
Subjects completed a questionnaire rating their overall experience with the Apollo Blood Glucose Monitoring System (User feedback on the system). The rating scale was 0 (Unacceptable) to 4 (Excellent).
One hour

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Utredare

  • Huvudutredare: Joy Frank, RN, Consumer Product Testing Co., Inc.

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart

1 november 2008

Primärt slutförande (Faktisk)

1 november 2008

Avslutad studie (Faktisk)

1 november 2008

Studieregistreringsdatum

Först inskickad

24 november 2008

Först inskickad som uppfyllde QC-kriterierna

24 november 2008

Första postat (Uppskatta)

25 november 2008

Uppdateringar av studier

Senaste uppdatering publicerad (Uppskatta)

29 februari 2016

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

29 januari 2016

Senast verifierad

1 januari 2016

Mer information

Termer relaterade till denna studie

Andra studie-ID-nummer

  • CTD-2008-17

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .

Kliniska prövningar på Diabetes mellitus

Kliniska prövningar på Apollo Blood Glucose Monitoring System

3
Prenumerera