Optimal Coronary Sinus Lead Implantation Using Intracardiac Impedography and Magnetic Resonance Imaging
Real-Time Intracardiac Impedograms of Left Ventricular Leads to Locate Sites of Latest Mechanical Delay in Cardiac Resynchronization Therapy
Study Overview
Status
Conditions
Intervention / Treatment
Detailed Description
This is a clinical trial using intracardiac impedance signals (impedograms) to assess regional dyssynchrony at various sites of left ventricular (LV) lead placement in humans undergoing CRT device implantation. This study will test the following hypotheses during the funding period:
- LV lead impedograms as an implant tool to place leads at sites of latest mechanical delay are feasible and correlate with other means of assessing dyssynchrony.
- LV lead impedograms vary significantly depending on LV lead location.
- There are several characteristics of LV lead impedograms that correlate with mechanical phenomena of the heart.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
Georgia
-
Atlanta, Georgia, United States, 30322
- Emory University Hospital
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
- Child
- Adult
- Older Adult
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Patients with NYHA class III or IV heart failure
- LVEF ≤ 30%
- QRS duration ≥ 120 ms
Exclusion Criteria:
- Not a candidate for CRT implantation
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
---|---|
Experimental: CRT Candidate
Patients with NYHA Class III or IV heart failure; EF ≤ 30% and QRS duration ≥ 120 ms, who are scheduled for CRT surgery. Intervention: Cardiac Resynchronization Therapy (CRT) implantation |
The impedance measurement is performed during device implantation following CMR. After the RV and LV leads are inserted, secured and tested, they will be connected to the impedance monitor. Impedance recording of at least ten beats will be acquired and stored for future analysis. Each recording will be tagged with the anatomical location of the LV lead. Subsequently, the LV lead will be moved to a different location and the same procedure will be repeated until accessible coronary sinus sites are exhausted. The ultimate LV lead location is determined by the implanting electrophysiologist and is not constrained by the study protocol. Finally, the LV lead will be tested again and the rest of the implantation procedure will proceed as routine. Arm: CRT Candidate |
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Anatomic correlation between largest Ts (see description) and site of longest delay among probed locations in the CMR dyssynchrony map.
Time Frame: Acute intraoperative measurement
|
Parameter will be recorded for at least 10 consecutive heartbeats during sinus and RV paced rhythm Ts = The average time from the sensed RV IEGM to the peak of the LV impedance curve |
Acute intraoperative measurement
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Differential correlation of Ts, Tp, and Td (see description) to the CMR dyssynchrony map.
Time Frame: Acute intraoperative measurement
|
Ts: The average time from the sensed RV IEGM to the peak of the LV impedance curve. Tp: The average time from the paced RV IEGM to the peak of the LV impedance curve. Td: The difference between Ts and Tp (Tp - Ts). |
Acute intraoperative measurement
|
Collaborators and Investigators
Sponsor
Investigators
- Principal Investigator: Michael S. Lloyd, MD, Emory University
Publications and helpful links
General Publications
- Abraham WT, Fisher WG, Smith AL, Delurgio DB, Leon AR, Loh E, Kocovic DZ, Packer M, Clavell AL, Hayes DL, Ellestad M, Trupp RJ, Underwood J, Pickering F, Truex C, McAtee P, Messenger J; MIRACLE Study Group. Multicenter InSync Randomized Clinical Evaluation. Cardiac resynchronization in chronic heart failure. N Engl J Med. 2002 Jun 13;346(24):1845-53. doi: 10.1056/NEJMoa013168.
- Moss AJ, Hall WJ, Cannom DS, Klein H, Brown MW, Daubert JP, Estes NA 3rd, Foster E, Greenberg H, Higgins SL, Pfeffer MA, Solomon SD, Wilber D, Zareba W; MADIT-CRT Trial Investigators. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med. 2009 Oct 1;361(14):1329-38. doi: 10.1056/NEJMoa0906431. Epub 2009 Sep 1.
- Chung ES, Leon AR, Tavazzi L, Sun JP, Nihoyannopoulos P, Merlino J, Abraham WT, Ghio S, Leclercq C, Bax JJ, Yu CM, Gorcsan J 3rd, St John Sutton M, De Sutter J, Murillo J. Results of the Predictors of Response to CRT (PROSPECT) trial. Circulation. 2008 May 20;117(20):2608-16. doi: 10.1161/CIRCULATIONAHA.107.743120. Epub 2008 May 5.
- Osswald S, Cron T, Gradel C, Hilti P, Lippert M, Strobel J, Schaldach M, Buser P, Pfisterer M. Closed-loop stimulation using intracardiac impedance as a sensor principle: correlation of right ventricular dP/dtmax and intracardiac impedance during dobutamine stress test. Pacing Clin Electrophysiol. 2000 Oct;23(10 Pt 1):1502-8. doi: 10.1046/j.1460-9592.2000.01502.x.
- Jessup M. MADIT-CRT--breathtaking or time to catch our breath? N Engl J Med. 2009 Oct 1;361(14):1394-6. doi: 10.1056/NEJMe0907335. Epub 2009 Sep 1. No abstract available.
- Couri DM, Mankad S. Cardiac resynchronization therapy. Curr Treat Options Cardiovasc Med. 2008 Dec;10(6):538-48. doi: 10.1007/s11936-008-0046-5.
- Lloyd MS, Heeke S, Lerakis S, Langberg JJ. Reverse polarity pacing: the hemodynamic benefit of anodal currents at lead tips for cardiac resynchronization therapy. J Cardiovasc Electrophysiol. 2007 Nov;18(11):1167-71. doi: 10.1111/j.1540-8167.2007.00943.x.
- Mounsey JP, Knisley SB. Anodal capture, cathodal capture, and left ventricular cardiac excitation. J Cardiovasc Electrophysiol. 2009 Jun;20(6):650-2. doi: 10.1111/j.1540-8167.2008.01418.x. Epub 2009 Jan 9. No abstract available.
- Theis C, Bavikati VV, Langberg JJ, Lloyd MS. The relationship of bipolar left ventricular pacing stimulus intensity to cardiac depolarization and repolarization in humans with cardiac resynchronization devices. J Cardiovasc Electrophysiol. 2009 Jun;20(6):645-9. doi: 10.1111/j.1540-8167.2008.01378.x. Epub 2009 Dec 15.
- Becker M, Franke A, Breithardt OA, Ocklenburg C, Kaminski T, Kramann R, Knackstedt C, Stellbrink C, Hanrath P, Schauerte P, Hoffmann R. Impact of left ventricular lead position on the efficacy of cardiac resynchronisation therapy: a two-dimensional strain echocardiography study. Heart. 2007 Oct;93(10):1197-203. doi: 10.1136/hrt.2006.095612. Epub 2007 Feb 19.
- Becker M, Kramann R, Franke A, Breithardt OA, Heussen N, Knackstedt C, Stellbrink C, Schauerte P, Kelm M, Hoffmann R. Impact of left ventricular lead position in cardiac resynchronization therapy on left ventricular remodelling. A circumferential strain analysis based on 2D echocardiography. Eur Heart J. 2007 May;28(10):1211-20. doi: 10.1093/eurheartj/ehm034. Epub 2007 Apr 10.
- Koos R, Neizel M, Schummers G, Krombach GA, Stanzel S, Gunther RW, Kelm M, Kuhl HP. Feasibility and initial experience of assessment of mechanical dyssynchrony using cardiovascular magnetic resonance and semi-automatic border detection. J Cardiovasc Magn Reson. 2008 Nov 4;10(1):49. doi: 10.1186/1532-429X-10-49.
- Tsao J, Kozerke S, Boesiger P, Pruessmann KP. Optimizing spatiotemporal sampling for k-t BLAST and k-t SENSE: application to high-resolution real-time cardiac steady-state free precession. Magn Reson Med. 2005 Jun;53(6):1372-82. doi: 10.1002/mrm.20483.
- RUSHMER RF, CRYSTAL DK, WAGNER C, ELLIS RM. Intracardiac impedance plethysmography. Am J Physiol. 1953 Jul;174(1):171-4. doi: 10.1152/ajplegacy.1953.174.1.171. No abstract available.
- Salo RW, Wallner TG, Pederson BD. Measurement of ventricular volume by intracardiac impedance: theoretical and empirical approaches. IEEE Trans Biomed Eng. 1986 Feb;33(2):189-95. doi: 10.1109/TBME.1986.325890. No abstract available.
- Kaye G, Edgar D, Mudawi T, Lippert M, Czygan G. Can transventricular intracardiac impedance measurement discriminate haemodynamically unstable ventricular arrhythmias in human? Europace. 2007 Feb;9(2):122-6. doi: 10.1093/europace/eul150.
- Fornwalt BK, Gonzales PC, Delfino JG, Eisner R, Leon AR, Oshinski JN. Quantification of left ventricular internal flow from cardiac magnetic resonance images in patients with dyssynchronous heart failure. J Magn Reson Imaging. 2008 Aug;28(2):375-81. doi: 10.1002/jmri.21446.
- Fornwalt BK, Arita T, Bhasin M, Voulgaris G, Merlino JD, Leon AR, Fyfe DA, Oshinski JN. Cross-correlation quantification of dyssynchrony: a new method for quantifying the synchrony of contraction and relaxation in the heart. J Am Soc Echocardiogr. 2007 Dec;20(12):1330-1337.e1. doi: 10.1016/j.echo.2007.04.030. Epub 2007 Jul 23.
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- IRB00036655
- ACTSI-EP-001 (Other Identifier: Other)
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Heart Failure, Systolic
-
Central Hospital, Nancy, FranceUnknown
-
Zensun Sci. & Tech. Co., Ltd.Active, not recruitingChronic Systolic Heart FailureChina
-
Zensun Sci. & Tech. Co., Ltd.TerminatedChronic Systolic Heart FailureChina
-
Jiangsu HengRui Medicine Co., Ltd.UnknownChronic Systolic Heart FailureChina
-
Hospital Moinhos de VentoTakedaUnknownHeart Failure, Systolic | Systolic Heart Failure | Cardiac FailureBrazil
-
Merck KGaA, Darmstadt, GermanyMerck Ltd.Completed
-
University of NebraskaEli Lilly and Company; Daiichi Sankyo, Inc.Completed
-
Abbott Medical DevicesCompletedSystolic Heart FailureAustralia, Hong Kong, Japan
-
University of CincinnatiCompletedSystolic Heart FailureUnited States
-
GlaxoSmithKlineCompletedHeart Failure, CongestiveKorea, Republic of
Clinical Trials on Cardiac Resynchronization Therapy (CRT) implantation
-
Abbott Medical DevicesCompletedHeart Failure | Cardiac Resynchronization TherapyAustralia
-
Larisa TereshchenkoCompletedHeart Failure | Cardiomyopathy | Left Bundle Branch BlockUnited States
-
Dr. Daniel P Morin, MD MPH FHRSCompletedPulmonary HypertensionUnited States
-
Oslo University HospitalSouth-Eastern Norway Regional Health Authority; Norwegian Health AssociationRecruitingHeart Diseases | Cardiovascular Diseases | Heart Failure | Cardiomyopathies | Pacemaker DDDNorway
-
Biotronik SE & Co. KGCompletedHeart Failure | Cardiac Resynchronization TherapyCzechia, Hungary, Germany, Austria
-
Maastricht University Medical CenterUnknownHeart Failure | Left Bundle-Branch Block | Cardiac Resynchronization TherapyNetherlands
-
Biotronik SE & Co. KGRecruitingHeart FailureSpain, Israel, Switzerland, France, Germany, Czechia, Portugal, Japan, Latvia, Hungary, Australia, Poland, South Africa, Taiwan, United Kingdom, Austria, Italy, Belgium, Singapore, Slovakia
-
Medtronic Cardiac Rhythm and Heart FailureMedtronicCompleted
-
Medtronic Cardiac Rhythm and Heart FailureCompletedHeart Diseases | Atrioventricular BlockUnited States, Canada