Automated Measurement of Optic Nerve Sheath Diameter

December 2, 2020 updated by: Venkatakrishna Rajajee, University of Michigan

Accuracy of Optic Nerve Sheath Diameter Measurement From Ultrasound Videos Using an Image Analysis Algorithm Compared to Expert Measurement in Patients With Acute Brain Injury

Optic Nerve Ultrasound (ONUS) is a promising non-invasive tool for the detection of raised Intracranial Pressure (ICP). Variability in the optimal Optic Nerve Sheath Diameter (ONSD) threshold corresponding to elevated ICP in multiple studies limits the value of ONUS in clinical practice. The investigators goal is to develop and validate an automated image analysis algorithm for standardization of ONSD measurement from ultrasound videos. Patients with acute brain injury requiring invasive ICP monitoring will undergo bedside ONUS, with blinded ONSD measurement by an expert investigator. The image analysis algorithm will then be used to measure ONSD and accuracy determined compared to the "reference standard" expert measurement.

Study Overview

Detailed Description

BACKGROUND:

A promising tool under investigation for the non-invasive estimation of ICP is Optic Nerve Ultrasound (ONUS). Raised ICP results in distension of the optic nerve sheath (ONS), a continuation of the dura mater. Ocular imaging, performed by clinicians using point-of-care ultrasound machines, can detect ONS distension behind the eye. A study conducted at the University of Michigan identified an ONS Diameter (ONSD) cutoff of >0.51cm as having 98% sensitivity and 91% specificity for the detection of intracranial hypertension, defined as ICP>25mmHg. The significant variation seen in the optimal ONSD threshold for identification of high ICP across several studies, however, greatly limits the practical application of this technique at the bedside. Much of this variation in the optimal ONSD threshold is likely related to technique, with variation in the margins used to define the ONS on acquired ultrasound images by different operators. The ONS, visible as a linear hypodense structure behind the eye can vary in its visualized dimension based on the angle and plane of insonation. Automated image analysis may permit standardization of ONSD measurement and thereby minimize interobserver variability.

SPECIFIC AIM:

The investigators goal is to develop a computer image-analysis algorithm to standardize measurement of the ONSD from ultrasound videos, and to validate against the reference standard of expert manual measurement of ONSD.

METHODS:

Development of image-analysis algorithm: Videos in DICOM format of prior ONUS studies performed for clinical purposes in the neurointensive care unit will be deidentified and used for initial development and modification of the automated image analysis algorithm, prior to prospective enrollment of subjects for determination of accuracy.

Under IRB approval, the investigators will perform Optic Nerve Ultrasound (ONUS) on eligible subjects admitted to the ICU following informed consent of the patient or appropriate surrogate.

Optic Nerve Ultrasound and ONSD measurement: These patients will undergo ONUS in the ICU. ONUS is performed with the patient's eye closed, and with a linear array transducer placed on the upper margin of the orbit to obtain a sonographic video clip of the eye, followed by bedside measurement of ONSD 3mm behind the level of the posterior scleral border. Imaging will be performed for both eyes for each patient. The expert investigator performing the ONUS study and manual ONSD measurement will be blinded to the patient's ICP by turning the monitor away from the sonographer from the time of entry into the patient's room until exit. The corresponding ICP from the invasive monitor will be separately documented by the bedside nurse. The ONUS video corresponding to the highest measured ONSD will be submitted in DICOM format for automated image analysis.

Equipment: Sonosite M-Turbo point-of-care ultrasound machine and an L25 linear array transducer with an ophthalmic preset.

Statistical Analysis:

  • Simple descriptive statistics including means and standard deviations
  • Simple Pearson correlation to allow for visual inspection across a range of values
  • Tukey mean-difference plot (Bland - Altman plot) will be used to assess agreement between the two methods, the ultrasound video analysis of optic nerve sheath diameter and ONSD ultrasound interpretation by a clinician. Limits of agreement will be calculated with standard errors and 95% confidence intervals.

Study Type

Observational

Enrollment (Actual)

75

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Michigan
      • Ann Arbor, Michigan, United States, 48109
        • University of Michigan Health System

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years and older (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Sampling Method

Non-Probability Sample

Study Population

Patients with acute brain injury requiring invasive ICP monitoring

Description

Inclusion Criteria:

  • Admitted to an ICU in the University of Michigan with Acute Brain Injury
  • Age> 18 years
  • Invasive ICP monitor (external ventricular drain or intraparenchymal catheter) already in place for a clinical indication.

Exclusion Criteria:

  • Pre-existing ocular pathology other than refractive error
  • Traumatic globe injury
  • Age < 18 years
  • Known pregnancy or subsequently discovered pregnancy after admission Prisoners

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Observational Models: Cohort
  • Time Perspectives: Prospective

Cohorts and Interventions

Group / Cohort
Intervention / Treatment
Optic nerve ultrasound
All patients will undergo Optic Nerve Ultrasound (ONUS), with measurement of Optic Nerve Sheath Diameter (ONSD) by both an expert investigator as well as by the automated image analysis algorithm.
All patients will undergo Optic Nerve Ultrasound (ONUS), with measurement of Optic Nerve Sheath Diameter (ONSD) by both an expert investigator as well as by the automated image analysis algorithm.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Expert measurement of Optic Nerve Sheath Diameter
Time Frame: Day of enrollment
The primary endpoint for determination of accuracy of automated image analysis measurement of Optic Nerve Sheath Diameter (ONSD) will be ONSD measurement performed by an expert investigator with >5 years experience with ONUS and ONSD measurement.
Day of enrollment

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Collaborators

Investigators

  • Principal Investigator: Venkatakrishna (Krishna) Rajajee, MD, University of Michigan

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

July 1, 2015

Primary Completion (Actual)

December 31, 2019

Study Completion (Actual)

December 31, 2019

Study Registration Dates

First Submitted

May 26, 2016

First Submitted That Met QC Criteria

May 31, 2016

First Posted (Estimate)

June 1, 2016

Study Record Updates

Last Update Posted (Actual)

December 4, 2020

Last Update Submitted That Met QC Criteria

December 2, 2020

Last Verified

December 1, 2020

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

Undecided

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Traumatic Brain Injury

Clinical Trials on Optic nerve ultrasound

3
Subscribe