VAPORHCS/OHSU J: MRI and Devices

April 11, 2024 updated by: Charles Henrikson, MD, Oregon Health and Science University

VAPORHCS/OHSU J: Safety of Clinically Indicated Magnetic Resonance Imaging in Patients With Permanent Pacemakers (PPM) and Implanted Cardioverter Defibrillators (ICDs)

This is a database study of clinically indicated MRIs on patients with permanent pacemakers and implantable cardioverter defibrillators.

Study Overview

Status

Recruiting

Conditions

Intervention / Treatment

Detailed Description

  1. Abstract

    Ferromagnetic Implants such as PPM's and ICD's are traditionally accepted as contraindications to MRI due to safety concerns. The hypothesis is that these procedures can be safely performed when appropriate safety precautions and vetting practices are implemented.

    This is a cohort study. 1000 patients with a clinical need for MRI will be included in the study at OHSU. The type of device and leads will be ascertained and a safety protocol will be strictly adhered to.

  2. Background

    It has been estimated that 50-75% of patients with implanted devices (pacemakers and ICDs) will develop an indication for magnetic resonance imaging (MRI) examination. Thus, there are approximately 200,000 patients annually denied MRI scans because of the presence of PPMs and ICDs in the United States. The current standard of care is not to perform these scans, and instead to utilize other (inferior) imaging, and make clinical decisions without the benefit of MRI imaging. This is the current standard of care of OHSU, where only pacemaker patients who are not pacemaker dependent can undergo MRI scans, per radiology policy. However, this leads to gaps in care for some patients who would clearly benefit from an MRI scan. In some centers, it has been shown that when proper precautions are taken, MRIs can be safely performed on patients with PPMs and ICDs. The published experience from Johns Hopkins reports over 500 scans performed without any adverse outcomes.(1) There are also several smaller studies documenting the safe experience with MRI scanning in device patients (reviewed in (1), table 4)

    The main concerns which have led to restrictions on MRIs in patients with cardiac rhythm devices are the following:(2)

    1. Force and torque-ferromagnetic devices in a magnetic field are subject to static and gradient magnetic field induced force and torque. However, the maximum force that a modern PPM or ICD generator would be subject to is 100g, far under what would be needed to dislodge a device.
    2. Current induction-the gradient magnetic fields in the MRI scanner could induce current in conductors in the field. However, the maximum observed current in vitro has been less than 0.5mA, far less than what is required to capture myocardium.
    3. Heating-Leads and other metallic devices can act as antennae, and may heat in the magnetic field, however, no heating greater than 0.5deg C were observed in vitro.
    4. Inappropriate pacing/shock/inhibition of pacing-Pacemakers and ICDs have the possibility of receiving electromagnetic interference from the MRI and having this lead to inappropriate pacing, inhibition of pacing, ICD shocks, reprogramming, or loss of function.

    While all of the concerns are real and have been reported in earlier generation devices, with proper patient selection and monitoring during the scan, MRIs can be performed safely on most device patients. Johns Hopkins has an experience of over 500 scans without any adverse events.(1) The proposed protocol here follows exactly the Hopkins protocol, and, indeed, the OHSU data will be combined with the Hopkins data for future publications.

    To review the most recent Johns Hopkins publication:(1) 555 MRI studies in 438 device (PPM and ICD) patients were reported. All MRI scans were clinically indicated, and a device clinician, familiar with device interrogation and programming, was physically present for the entire duration of the scan. All patients were continuously monitored via telemetry and pulse oximetry during the MRI. For pacemaker patients, if they are pacemaker dependent they were reprogrammed to a non-sensing mode for the scan. If they were not pacemaker dependent, they were left in their usual mode. For ICD patients, if they were not pacemaker dependent, they were left in their usual mode. However, pacemaker dependent ICD patients were not scanned, as ICDs generally do not have a non-sensing pacing mode. Therapy for tachycardia was disabled in all ICDs for the scan.

    There were three clinical events during scans, all were "power on resets", which means that the device resets to "out of the box" original factory settings. One of these patients had an ICD, felt tugging in his chest during the scan, and the scan was not completed. The other two power on resets were in PPM patients, in whom the scan was completed. Neither were pacemaker dependent, and neither noticed anything abnormal during the scan. All three patients had normally functioning devices following the scan, and at followup.

    In the overall patient population in the Hopkins study, no single patient had a malfunction of a device immediately following the scan. There was a slight change in the electrical parameters of the leads, on average, but this did not result in the need for any intervention on any patient.

  3. Objectives

    The primary objective of this study is to develop a protocol, consistent with DHHS/CMS guidelines and requirements to further document safety in clinically indicated MR imaging in patients with implanted cardioverter defibrillators (ICD's) and pacemakers.

  4. Study Procedures

    MRI's will be conducted on patients who require a clinically indicated MRI only. All studies will be done in the 1.5 Tesla MRI unit. ECG monitoring pads will be placed on the patients for the duration of the study. An external defibrillator and ACLS drugs will be on hand. Heart rate, blood pressure, O2 saturation will also be monitored non-invasively throughout the study. All devices will undergo a complete interrogation prior to imaging. Parameters such as atrial and ventricular pacing thresholds, R and P wave amplitudes, lead impedance, and battery status will be measured and recorded. PPMs will be programmed to an asynchronous mode if dependent and to an inhibited mode in patients without pacemaker dependence.

    The pacemaker function of ICD's in pacemaker independent patients will be programmed to ventricular inhibited (VVI) mode at 50 bpm. Pacemaker dependent patients with ICDs will be excluded.

    In addition, there is a theoretical concern that patients who have capped or abandoned leads in their chest may be at risk for heating. While there is little or no objective data to support this concern, given this theoretical concern, patients in this category will only be admitted to the study if the referring physician deems the MRI critical and after the PI has carefully considered the risk/benefit ration for each patient who falls in to this category.

    Similarly, there historically has also been a concern that newly implanted leads (<4 weeks) are at risk for lead to dislodgement d/t possible torque from the MRI environment. This too is only a theoretical concern and there is little or no scientific evidence to support this claim. Nevertheless, patients with newly implanted pacemaker/ICD leads will be considered by the PI for protocol inclusion only if the referring physician deems the MRI to be critical and after considering the risk/benefit to each individual patient.

    After the appropriate MRI protocol for each MRI patient's unique condition has been completed, the device will be re-programmed to its original settings and completely interrogated to detect any changes in device performance.

    The duration of the study extends to the post procedure interrogation at 1-6 weeks post MRI. IF the PPM or ICD reveals any malfunction post imaging (not seen yet), the patient will be followed by the electrophysiology service and appropriate follow-up management will be arranged.

  5. Study Statistics

    1. Primary outcome variable. Patient safety and device malfunction during or after MRI
    2. Secondary outcome variables. None
    3. Statistical plan including sample size justification and interim data analysis. We hope to enroll 1000 patients over 4 years to further evaluate safety.
    4. Early stopping rules. N/A
  6. Risks

    1. Medical risks, listing all procedures, their major and minor risks and expected frequency.

      There are no studies showing any health hazards associated with magnetic field exposure. PPM and ICD implants subjected to the strong electromagnetic field of MR imaging are thought to be susceptible to movement and torque. Other traditional concerns include malfunction due to demagnetization or magnetic activation leading to inappropriate pacing and shocks and artificial sensing of electromagnetic interference from MRI leading to inappropriate inhibition and/or activation of antitachycardia therapy, device malfunction and induced arrhythmias, as detailed in the background section above. Heating of the device and surrounding tissues leading to fibrosis and increased capture thresholds have also been a theoretical concern. However, recent work has revealed the safety of PPM's and certain ICD's in the MR environment given appropriate precautions(1, 3, 4).

    2. Steps taken to minimize the risks. Appropriate measures discussed in the protocol section will be taken to minimize risks associated with device exposure to MRI. Consistent with standard MRI policy, patients in the first trimester of pregnancy and patients with history of allergic reactions to gadolinium will undergo imaging without any contrast media.
    3. Plan for reporting unanticipated problems or study deviations. Patient data will be monitored on a per patient basis. Pre-vs. post MRI device testing data will be compared and any significant differences immediately reported to the IRB.
  7. Benefits a. Description of the probable benefits for the participant and for society. The patient will benefit from a standardized protocol to minimize the risk of an MR scan deemed necessary for his/her care which would otherwise be denied or performed without any organized framework. Society will benefit because of the increasing number of implanted devices and expanding use of MRI.

8 Costs

a. Detail costs of study procedure(s) or drug (s) or substance(s) to participants and identify who will pay for them.

The patient and his/her insurance will be responsible for the cost of the clinically indicated MRI.

Study Type

Interventional

Enrollment (Estimated)

10000

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Study Locations

    • Oregon
      • Portland, Oregon, United States, 97239
        • Recruiting
        • Oregon Health and Scince University
        • Contact:
          • Charles Henrikson, MD
      • Portland, Oregon, United States, 97239
        • Recruiting
        • Portland Veterans Administration Medical Center
        • Contact:
          • Peter Jessel, MD

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

9 years to 99 years (Child, Adult, Older Adult)

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

- Patients with an absolute clinical need for MR imaging and PPM (year 1996 model or later) or ICD's (year 2000 or later)

Exclusion Criteria:

  • Patients with pacemaker models before 1996 and ICD models before year 2000.
  • Pacemaker dependent patients with ICD.
  • Patients who weighing less than 80 lbs.
  • Patients who complete the MRI standard screening form and are deemed inappropriate for MRI for any reason
  • Pregnant patients will be excluded in their first trimester and will not get gadolinium at any time during their pregnancy
  • Patient with abandoned/capped leads or leads implanted <4 weeks unless the procedure is deemed critical by the referring physician and approved by the PI

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Other: the study
Patients with a PPM or ICD getting a clinically indicated MRI
MRI

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
MRI Safety during MRI scan in patients with PPMs/ICDs.
Time Frame: During MRI scan.
Percent of patients successfully completing MRI scan without complication per interrogation of their device and clinical assessment.
During MRI scan.
MRI Safety following MRI scan in patients with PPMs/ICDs.
Time Frame: 1-6 week follow up in device clinic
Percent of patients successfully following up after MRI scan without complication per interrogation of their device and clinical assessment.
1-6 week follow up in device clinic

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Charles Henrikson, MD, Oregon Health and Science University

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

July 19, 2012

Primary Completion (Estimated)

December 31, 2030

Study Completion (Estimated)

December 31, 2030

Study Registration Dates

First Submitted

March 13, 2019

First Submitted That Met QC Criteria

April 11, 2019

First Posted (Actual)

April 16, 2019

Study Record Updates

Last Update Posted (Estimated)

April 15, 2024

Last Update Submitted That Met QC Criteria

April 11, 2024

Last Verified

April 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • 8423

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.