Dialysis Performance of the FX CorAL Membrane (PATENCY)

January 2, 2024 updated by: University Hospital, Ghent

Performance of Toxin Removal and Clotting Kinetics in Dialyzers

Anticoagulation remains an important issue in the setting of hemodialysis, and up till now there are some major points on which further research is needed. First, it is important to have a portfolio of the performance of different commercially available dialyzers with respect to fiber clotting. Second, to better estimate the impact of clotting on the overall dialysis performance, clotting kinetics during dialysis should be understood.

The aim of the present project is therefore to quantify the performance of the FX CorAL dialyzer (Fresenius Medical Care, Germany) in settings with reduced anticoagulation, and compare different performance outcomes (percent open fibers, solute removal rates) to those of other commercially available dialyzers. The different outcomes are related to the dialyzer extraction ratio and reduction ratio of small and middle molecules and albumin, the visual scoring of the dialyzer post dialysis, and the anticoagulation properties as assessed by fiber blocking in the dialyzer. Performance parameters at different time steps will allow to better understand clotting kinetics during dialysis.

Study Overview

Detailed Description

The FX CorAL is a polysulfone dialyzer with improved biocompatibility due to the increased polyvinylpyrrolidone (PVP) content on its blood side surface. Herewith, protein adsorption onto the membrane is lower, decreasing initiation of inflammation and suggesting less clotting problems.

The comPERFORM study already revealed the superiority of the FX CorAL 600 dialyzer versus the polysulfone-based Xevonta Hi15 (B. Braun, Germany) and the polyethersulfone-based Elisio 150H (Nipro, Japan) with respect to the clearance of Beta-2-Microglobulin at 60 min after dialysis start. Also the clearance of the larger toxin myoglobin was found superior in the FX CorAL, and removal rates, as calculated over the 240 min post dilution hemodiafiltration sessions, were superior in the FX CorAL for both middle molecules. Furthermore, albumin sieving properties of FX CorAL changed less over time than with the competitors.

Our group previously showed that here is a correlation between myoglobin dialyzer extraction and the percentage of open dialyzer fibers as quantified using the gold standard micro Computed Tomography (CT) scanning technique. However, previous data also suggested that fiber blocking during dialysis is not a linear process, but is rather accelerated during the second half of dialysis. Nevertheless, details about the kinetics of membrane and fiber clotting are still lacking. It can be hypothesized that the membrane pores and fibers in the FX CorAL have an improved tendency to remain open during the entire dialysis session, and that this will result in improved total solute removal, especially for middle molecules. Quantification of dialyzer fiber blocking using micro-CT scanning of dialyzer cross-sections, and this preferably after different durations of dialysis sessions, should be used to test this hypothesis. By including time points for performance measurements, like Extraction Ratio, with blood sampling e.g. at 10, 30, 60 min after dialysis start on the one hand, and at 180 and 240min, it might be possible to better understand clotting kinetics.

Polysulfone and polyethersulfone membranes have previously been associated with hypersensitivity reactions attributed to the use of additives such as PVP to enhance hydrophilic properties. In the FX CorAL however, the PVP is surface-stabilized and does not elute from the inner membrane surface. The Solacea dialyzer (Nipro, Japan), manufactured with an asymmetric triacetate (ATATM) membrane without hydrophilization agents, is also showing a lower risk of hypersensitivity, less decrease in platelets, and high permeability and filtration performance, and might be seen as a competitor for the FX CorAL.

Anticoagulation needs to be well-balanced to avoid an increased risk for bleeding complications on the one hand, and clotting of the extracorporeal circuit resulting in blood loss for the patient on the other hand. A high degree of fiber clotting in the early stage of the dialysis session might also result in decreased solute removal. More biocompatible membranes, such as FX CorAL and Solacea are theoretically less prone to coagulation problems during dialysis and could be used with decreased anticoagulation dose, reducing bleeding problems whilst still maintaining patency. The Solacea dialyzer showed maintained adequacy in conditions where systemic anticoagulation is contraindicated. Up till now, no studies to explore this in the FX CorAL have been performed.

The main questions the investigators want to answer with this study are:

  1. How is fiber patency post dialysis in the FX CorAL versus competitors?
  2. Can the FX CorAL be safely and adequately used in settings with reduced anticoagulation?
  3. How is the time-dependent evolution of dialyzer extraction and clearance of small and middle molecules changing during the course of dialysis?
  4. How is the time-dependent evolution of albumin loss?
  5. How is the progression of membrane blockage and fiber blockage during the course of dialysis, i.e. during the last hour of a 4 hours dialysis session?

Study Type

Interventional

Enrollment (Actual)

10

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Gent, Belgium, 9000
        • Ghent University Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

No

Description

Inclusion Criteria:

  • stable chronic hemodialysis patient
  • well functioning vascular access

Exclusion Criteria:

  • known coagulation disorder
  • active inflammation
  • malignancy

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: Randomized
  • Interventional Model: Crossover Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Hemodiafiltration with FX CorAL 800 dialyzer
  • Patients are dialyzed three consecutive dialysis sessions with an FX CorAL 800 dialyzer, starting on Saturday.
  • On Tuesday and Thursday, patients are dialyzed with only a quarter of their regular anticoagulation dose during 4 hours on Tuesday, and 3 hours on Thursday.
  • Blood is sampled from the arterial and venous blood line at different time points to determine concentrations of different uremic toxins and albumin
Patients are dialyzed with only one quarter of their regular anticoagulation dose during the sessions on Tuesday and Thursday
Blood is sampled at start and at 10, 30, 60, 180, and 240 min after dialysis start for the 4 hours dialysis session, and at start and 10 and 180 min after dialysis start for the 3 hours dialysis session.
Patients have a washout with their regular dialysis duration on Saturday, a study session of 4 hours dialysis on Tuesday, and a study session of 3 hours dialysis on Thursday
Patients are dialyzed with one of the four dialyzers during the entire week, starting at Saturday (i.a. Saturday, Tuesday and Thursday). Block randomization is performed per two weeks. In week 1 and 2, patients are dialyzed with the FX CorAL 800 and the FX 800 Cordiax. In week 3 and 4 of the study, patients are dialyzed with the Solacea 19H and the Xevonta Hi20.
Experimental: Hemodiafiltration with FX 800 Cordiax dialyzer
  • Patients are dialyzed three consecutive dialysis sessions with an FX 800 Cordiax dialyzer, starting on Saturday.
  • On Tuesday and Thursday, patients are dialyzed with only a quarter of their regular anticoagulation dose during 4 hours on Tuesday, and 3 hours on Thursday.
  • Blood is sampled from the arterial and venous blood line at different time points to determine concentrations of different uremic toxins and albumin
Patients are dialyzed with only one quarter of their regular anticoagulation dose during the sessions on Tuesday and Thursday
Blood is sampled at start and at 10, 30, 60, 180, and 240 min after dialysis start for the 4 hours dialysis session, and at start and 10 and 180 min after dialysis start for the 3 hours dialysis session.
Patients have a washout with their regular dialysis duration on Saturday, a study session of 4 hours dialysis on Tuesday, and a study session of 3 hours dialysis on Thursday
Patients are dialyzed with one of the four dialyzers during the entire week, starting at Saturday (i.a. Saturday, Tuesday and Thursday). Block randomization is performed per two weeks. In week 1 and 2, patients are dialyzed with the FX CorAL 800 and the FX 800 Cordiax. In week 3 and 4 of the study, patients are dialyzed with the Solacea 19H and the Xevonta Hi20.
Experimental: Hemodiafiltration with Solacea 19H dialyzer
  • Patients are dialyzed three consecutive dialysis sessions with a Solacea 19H dialyzer, starting on Saturday.
  • On Tuesday and Thursday, patients are dialyzed with only a quarter of their regular anticoagulation dose during 4 hours on Tuesday, and 3 hours on Thursday.
  • Blood is sampled from the arterial and venous blood line at different time points to determine concentrations of different uremic toxins and albumin
Patients are dialyzed with only one quarter of their regular anticoagulation dose during the sessions on Tuesday and Thursday
Blood is sampled at start and at 10, 30, 60, 180, and 240 min after dialysis start for the 4 hours dialysis session, and at start and 10 and 180 min after dialysis start for the 3 hours dialysis session.
Patients have a washout with their regular dialysis duration on Saturday, a study session of 4 hours dialysis on Tuesday, and a study session of 3 hours dialysis on Thursday
Patients are dialyzed with one of the four dialyzers during the entire week, starting at Saturday (i.a. Saturday, Tuesday and Thursday). Block randomization is performed per two weeks. In week 1 and 2, patients are dialyzed with the FX CorAL 800 and the FX 800 Cordiax. In week 3 and 4 of the study, patients are dialyzed with the Solacea 19H and the Xevonta Hi20.
Experimental: Hemodiafiltration with Xevonta Hi-20
  • Patients are dialyzed three consecutive dialysis sessions with a Xevonta Hi-20 dialyzer, starting on Saturday.
  • On Tuesday and Thursday, patients are dialyzed with only a quarter of their regular anticoagulation dose during 4 hours on Tuesday, and 3 hours on Thursday.
  • Blood is sampled from the arterial and venous blood line at different time points to determine concentrations of different uremic toxins and albumin
Patients are dialyzed with only one quarter of their regular anticoagulation dose during the sessions on Tuesday and Thursday
Blood is sampled at start and at 10, 30, 60, 180, and 240 min after dialysis start for the 4 hours dialysis session, and at start and 10 and 180 min after dialysis start for the 3 hours dialysis session.
Patients have a washout with their regular dialysis duration on Saturday, a study session of 4 hours dialysis on Tuesday, and a study session of 3 hours dialysis on Thursday
Patients are dialyzed with one of the four dialyzers during the entire week, starting at Saturday (i.a. Saturday, Tuesday and Thursday). Block randomization is performed per two weeks. In week 1 and 2, patients are dialyzed with the FX CorAL 800 and the FX 800 Cordiax. In week 3 and 4 of the study, patients are dialyzed with the Solacea 19H and the Xevonta Hi20.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Fiber patency (i.e. number of open dialyzer fibers) in FX CorAL versus the competing dialyzers
Time Frame: 4 months
Patients are dialyzed with four different dialyzers: FX CorAL 800, FX 800 Cordiax, Solacea 19H and Xevonta 20Hi
4 months
Calculation of dialyzer extraction in FX CorAL dialyzer in the setting of reduced anticoagulation
Time Frame: 4 months
During the study dialysis sessions, patients only receive one quarter of their regular anticoagulation dose
4 months

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Time evolution of dialyzer extraction of uremic toxins and albumin
Time Frame: 4 months
Blood is sampled at different time points to calculate extraction of different uremic toxins and albumin during dialysis
4 months
Time evolution of fiber patency during dialysis
Time Frame: 4 months
Fiber patency is determined by micro Computed Tomography (CT) scanning after the dialysis session of 4 and 3 hours, respectively. By comparing the patency for those two sessions, we will learn about the progression of fiber clotting during 3 and 4 hours dialysis
4 months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Wim Van Biesen, PhD, University Hospital, Ghent

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

October 2, 2023

Primary Completion (Actual)

October 26, 2023

Study Completion (Estimated)

August 31, 2024

Study Registration Dates

First Submitted

December 7, 2023

First Submitted That Met QC Criteria

January 2, 2024

First Posted (Actual)

January 12, 2024

Study Record Updates

Last Update Posted (Actual)

January 12, 2024

Last Update Submitted That Met QC Criteria

January 2, 2024

Last Verified

January 1, 2024

More Information

Terms related to this study

Other Study ID Numbers

  • UGhent

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.