Evaluation of Commonly Used Blood Glucose Meters in Vietnam

January 18, 2024 updated by: Huan Nguyen Pham, Hung Vuong Hospital

Evaluation of Commonly Used Blood Glucose Meters in Vietnam and Effect of Hematocrit on Device's Accuracy

Rationale: Most BGM (blood glucose monitoring) devices in Vietnam are manufactured for home use only. However, they are commonly used in healthcare facility, primarily at point of care. We want to assess the validity of these devices by using CLSI (Clinical and Laboratory Standards Institute) POCT12-A3 to assess the accuracy.

Objective: To validate the accuracy and effect of hematocrit on these devices, along with the precision profile using CLSI EP15A3

Study design: This is a cross-sectional study.

Study population: Because we need a wide range of blood glucose level. We chose pregnant women who underwent 75g glucose tolerance test.

Study Overview

Status

Completed

Detailed Description

Most blood glucose monitors circulating the market are manufactured after ISO 15197:2013 standard, which is intended for self-measurement. Medical organizations need to verify blood glucose monitors if they are designed for prescription point-of-care use or validate blood glucose monitors if they are designed for over-the-counter use.

Pregnant women are selected after consent, their blood are then drawn by phlebotomist into an EDTA (ethylenediamine tetraacetic acid) with NaF (sodium flouride) blood tube. Samples are run on alinity h hematology analyzer, manufactured by Abbott, to measure hematocrit. The analyzers are verified by laboratory procedure, recommended by CLSI (Clinical and Laboratory Standards Institute) method and quality control is checked before in accordance to the laboratory procedure. After that, blood tube is mixed thoroughly before being pipetted. Each BGM device will measure blood glucose level of the pipetted whole blood. Finally, samples will be run on a cobas c 502 analyzer to measure plasma glucose level, which is a reference method for determining the accuracy of the BGM devices.

Statistical analysis: Meter performance is acceptable for use in hospitals when 95% of the individual results from the POC glucose meter system agree within 0.67 mmol/L (12 mg/dL) of the laboratory analyzer values at glucose concentrations below 5.55 mmol/L (100 mg/dL) and within 12.5% of the laboratory analyzer values at glucose concentrations at or above 5.55 mmol/L (100 mg/dL). In addition, the sum of the number of individual results with 1) errors that exceed 0.83 mmol/L (15 mg/dL) at glucose concentrations below 4.2 mmol/L (75 mg/dL) and 2) errors that exceed 20% at glucose concentrations at or above 4.2 mmol/L (75 mg/dL) should not exceed 2% of all results. An ordinary linear regression is performed to see how well the glucose level of BGM devices correlates with the reference method. Bland-Altman difference plots are used to describe the difference along with the hematocrit of each sample.

Study Type

Observational

Enrollment (Actual)

135

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

      • Ho Chi Minh, Vietnam, 700000
        • Hung Vuong Hospital

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult

Accepts Healthy Volunteers

Yes

Sampling Method

Non-Probability Sample

Study Population

Pregnant women at obstetrics clinics at Hung Vuong Hospital

Description

Inclusion Criteria:

  • Pregnant women undergo 75g glucose tolerance test

Exclusion Criteria:

  • Usage of anticoagulant drug
  • History of bleeding disorders.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Hematocrit
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Hematocrit in percentage, measured on hematology analyzer alinity h
Each participant's blood samples are run immediately after phlebotomy
Reference Glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on reference glucose analyzer cobas c 502
Each participant's blood samples are run immediately after phlebotomy
ACCU-CHEK Inform II glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK Inform II, Serial number UU14463019
Each participant's blood samples are run immediately after phlebotomy
ACCU-CHEK performa glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ACCU-CHEK performa, Serial number 68827245972
Each participant's blood samples are run immediately after phlebotomy
ONETOUCH VerioVue glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device ONETOUCH VerioVue, Serial number YEPMT06H
Each participant's blood samples are run immediately after phlebotomy
Contour Plus glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device Contour Plus, Serial number 6128289
Each participant's blood samples are run immediately after phlebotomy
EasyGlucose glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device EasyGlucose, Serial number G16D21I0700816
Each participant's blood samples are run immediately after phlebotomy
VivaChek Ino glucose concentration
Time Frame: Each participant's blood samples are run immediately after phlebotomy
Glucose concentration, measured on assessed blood glucose monitoring device VivaChek Ino, Serial number 311Z0021DFA
Each participant's blood samples are run immediately after phlebotomy

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Huan N Pham, M.D., Hung Vuong Hospital

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

June 1, 2022

Primary Completion (Actual)

June 16, 2022

Study Completion (Actual)

October 20, 2022

Study Registration Dates

First Submitted

January 18, 2024

First Submitted That Met QC Criteria

January 18, 2024

First Posted (Actual)

January 29, 2024

Study Record Updates

Last Update Posted (Actual)

January 29, 2024

Last Update Submitted That Met QC Criteria

January 18, 2024

Last Verified

January 1, 2024

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

YES

IPD Plan Description

Data will be shared upon reasonable request

IPD Sharing Time Frame

3 months after publication

IPD Sharing Supporting Information Type

  • STUDY_PROTOCOL
  • SAP

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Diabetes

3
Subscribe