A Clinical Study of the Performance of a Glucose Blinding Protein-Based Continuous Glucose Monitor (GBP CGM) (SMA002)

July 3, 2015 updated by: Stacey Anderson, University of Virginia

A Clinical Study of the Warm-Up, Accuracy, and Lag Performance of a Glucose Blinding Protein-Based Continuous Glucose Monitor (GBP CGM)

The purpose of this study is to assess the accuracy of the Becton Dickenson (BD) Technologies Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) in patients with Type 1 diabetes during low (goal glucose 55 mg/dL), normal (80-140 mg/dL) and high (>180mg/dL) glucose levels over a 24 hour period. This will be achieved by monitoring blood sugar levels when a regular dinner meal is given, when a liquid breakfast meal (BOOST Original containing 41 grams Carbohydrates (CHO), 4 grams fat, 10 grams protein) is given, when subcutaneous insulin is dosed to induce hypoglycemia to a goal of 55 mg/dL, and when a regular lunch meal is given. A Continuous Glucose Monitor (CGM) is an electronic device that measures and displays blood sugar (glucose) levels in the body throughout the day and night. The method being used to detect blood sugar in the investigational Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) is different than the method that is currently in use by commercially available models.

Some sensors cannot tell the difference between glucose (sugar) and other substances such as Tylenol, aspirin or citric acid etc. Because they cannot tell the difference, they may give false readings. The GBP CGM is made to only recognize glucose in the body rather than other substances (e.g., Tylenol, aspirin, citric acid, etc.). As a result, the investigators expect the new GBP CGM to be more accurate at detecting low blood sugar levels than the current devices.

Study Overview

Status

Completed

Intervention / Treatment

Detailed Description

Development of accurate continuous glucose monitoring devices is critically important for the maintenance of strict glycemic control without increasing the risk for hypoglycemia and for the successful implementation of an artificial pancreas. Glucose binding protein-based (GBP) continuous glucose monitors offer several potential advantages over commercially available glucose oxidase based sensors. Unlike glucose oxidase based sensors, GBP sensors do not require establishment of an equilibrium reduction/oxidation reaction and are not as susceptible to interference from other compounds because they are engineered to recognize only glucose. As a result, GBP-based glucose sensors are exceptionally accurate at low blood glucose levels and have a faster warm-up time than glucose oxidase based systems. The principal idea of this proposal is to evaluate the accuracy and performance of a second generation glucose binding protein-based continuous glucose monitor (GBP CGM) developed by BD Technologies in subjects with type 1 diabetes, with emphasis on warm-up period after insertion, accuracy in hypoglycemic and hyperglycemic ranges, and lag time over a 24 hour period of sensor use. In order to accomplish this, the investigators will conduct a clinical trial consisting of 25.5-hour studies involving 15 subjects, each wearing one active and one mock (no GBP) device simultaneously during meals and induced hypoglycemia to observe a wide range of glucose values. The investigators envision that the glucose binding protein-based continuous glucose monitor will demonstrate minimal warm-up time after insertion, overall accuracy comparable to commercial glucose oxidase sensors with excellent performance in the hypoglycemic range and a nominal lag time to sensing a change in blood glucose.

Study Type

Interventional

Enrollment (Actual)

16

Phase

  • Phase 2

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

    • Virginia
      • Charlottesville, Virginia, United States, 22908
        • University of Virginia

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

21 years to 65 years (Adult, Older Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Clinical diagnosis of type 1 diabetes mellitus for ≥1 year. For an individual to be enrolled at least one criterion from each list must be met.
  • Criteria for documented hyperglycemia (at least 1 must be met):

    1. Fasting glucose ≥126 mg/dL - confirmed
    2. Two-hour oral Glucose Tolerance Tests (OGTT) glucose ≥200 mg/dL - confirmed
    3. HbA1c ≥6.5% documented - confirmed
    4. Random glucose ≥200 mg/dL with symptoms
    5. No data at diagnosis is available but the participant has a convincing history of hyperglycemia consistent with diabetes
  • Criteria for requiring insulin at diagnosis (1 must be met):

    1. Participant required insulin at diagnosis and continually thereafter
    2. Participant did not start insulin at diagnosis but upon investigator review likely needed insulin (significant hyperglycemia that did not respond to oral agents) and did require insulin eventually and used continually
    3. Participant did not start insulin at diagnosis but continued to be hyperglycemic, had positive islet cell antibodies - consistent with latent autoimmune diabetes in adults (LADA) and did require insulin eventually and used continually
  • Use of an insulin pump to treat his or her diabetes for at least six months prior to the study.
  • Actively using a bolus calculator function with the current insulin pump with pre-defined parameters for glucose goal(s), carbohydrate ratio(s), and insulin sensitivity factor(s).
  • Signed informed consent
  • Age ≥21 and <65 years old
  • Body mass index between 19 and 30 kg/m2, inclusive
  • HbA1c ≤11%

Exclusion Criteria:

  • Uncontrolled arterial hypertension (diastolic blood pressure >90 mm Hg and/or systolic blood pressure >160 mm Hg)
  • Impaired hepatic function measured as alanine aminotransferase or aspartate aminotransferase ≥three times the upper reference limit
  • Impaired renal function measured as creatinine >1.2 times above the upper limit of normal
  • Diabetic ketoacidosis in the past 6 months
  • Severe hypoglycemia resulting in a seizure or loss of consciousness in the 12 months prior to enrollment
  • Conditions which may increase the risk of induced hypoglycemia such as known coronary artery disease, congestive heart failure, history of any cardiac disorder or arrhythmia, history of a cerebrovascular event, history of migraines, seizure disorder, syncope, adrenal insufficiency, or neurological disease.
  • Current use of medications containing >4000 mg acetaminophen per day.
  • Current use of L-Monoamine oxidases (MAO) inhibitors.
  • Known microvascular (diabetic) complications (other than diabetic non-proliferative retinopathy),such as history of laser coagulation, proliferative diabetic retinopathy, known diabetic nephropathy (other than microalbuminuria with normal creatinine) or neuropathy requiring treatment
  • Known allergy to eggs
  • Pregnancy, breast-feeding or intention of becoming pregnant
  • Current or recent alcohol or drug abuse by patient history.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Supportive Care
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: GBP-CGM
All participants will wear one active GBP-CGM and one inactive GBP-CGM
Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5.
Other Names:
  • Glucose binding protein-based continuous glucose monitor

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Absolute Relative Difference (ARD)
Time Frame: 25.5 hours

ARD=100*(G_sensor-G_reference)/G_reference

Calculated for when patient's G_ref was Normal (70-180 mg/dl), Hyperglycemic (>180 mg/dl) and Hypoglycemic (<70 mg/dl)

The study data includes 208 paired sensor-YSI plasma glucose readings (G_reference) for each GBP CGM sensor (G_sensor) inserted for 24 hours during hyperglycemic and hypoglycemic challenge conditions. Data pairs will permit the detailed evaluation of sensor performance parameters, including static accuracy metrics such as median and mean absolute deviations and median and mean absolute relative deviation and Point CG-EGA, as well as dynamic parameters, such as warm-up time, trend accuracy (Rate CG-EGA), and sensor lag.

25.5 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Investigators

  • Principal Investigator: Stacey Anderson, MD, University of Virginia, Center for Diabetes Technology

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start

November 1, 2011

Primary Completion (Actual)

July 1, 2012

Study Completion (Actual)

July 1, 2012

Study Registration Dates

First Submitted

November 7, 2011

First Submitted That Met QC Criteria

November 8, 2011

First Posted (Estimate)

November 10, 2011

Study Record Updates

Last Update Posted (Estimate)

July 31, 2015

Last Update Submitted That Met QC Criteria

July 3, 2015

Last Verified

July 1, 2015

More Information

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Type 1 Diabetes Mellitus