Servo Controlled Oxygen Targeting Study (SCO2T)
Randomised Cross-over Study of Servo-controlled Oxygen Targeting for Premature Infants
Most premature babies require oxygen therapy. There is uncertainty about what oxygen levels are the best. The oxygen levels in the blood are measured using a monitor called a saturation monitor and the oxygen the baby breathes is adjusted to keep the level in a target range. Although there is evidence that lower oxygen levels maybe harmful, it is not known how high they need to be for maximum benefit. Very high levels are also harmful. Saturation monitors are not very good for checking for high oxygen levels. For this a different kind of monitor, called a transcutaneous monitor, is better.
Keeping oxygen levels stable is usually done by nurses adjusting the oxygen levels by hand (manual control). There is also equipment available that can do this automatically (servo control). It is not known which is best.
Studies of automated control have shown that infants spend more time within their intended target oxygen saturation range. These have not included measurements of transcutaneous oxygen.
The investigators aim to show the transcutaneous oxygen levels as well as the oxygen saturation levels when babies have their oxygen adjusted manually or automatically.
Study Overview
Status
Intervention / Treatment
Detailed Description
Presently oxygen is titrated against saturation (SpO2) by manual adjustment. Automated or servo-control systems have been developed that result in tighter control of SpO2 and more time spent in the intended target range. These systems are already in clinical use. Automated systems produce quite large fluctuations in fraction of inspired oxygen (FiO2) in order to keep SpO2 in range. It is possible that this could result in short periods of high or low oxygen tension (PO2) that are undetectable using saturation monitoring. Studies to date have examined the effects of manual and automated (servo) oxygen targeting on SpO2 but not on transcutaneous oxygen tension (TcPO2).
There is a need to determine the achieved SpO2 and TcPO2 distributions associated with the use of manual and automated control as a first step in planning trials comparing these approaches over many weeks. When this is measured over a small number of hours it is not anticipated that this would have an influence on clinical outcome.
This study is a prospective, single centre, randomised crossover trial of automated (servo) control versus manual oxygen titration. Each infant will act as their own control. Infants born at less than 29 weeks gestation, greater than 48 hour of age and receiving supplementary oxygen will be eligible for inclusion.
The study will be undertaken in the Neonatal Unit at the Simpson Centre for Reproductive Health at the Royal Infirmary of Edinburgh.
Total study time is 12 hours for each infant. Infants will be randomised to commence on either automated (servo) control or manual mode. SpO2 (range 90-95%) will be continuously monitored as per normal standard of care. A second pulse oximetry probe will be place for servo control input.
Additional monitoring will be carried out as shown below:
- TcPO2 monitoring
- FiO2 monitoring
- Heart rate monitoring (used to validate SpO2 readings)
- Arterial gas sampling (only if conducted by the direct care team as part of the routine care of the infant; no extra blood samples will be taken as part of the study)
In manual mode, all oxygen adjustments will be made by clinical/nursing staff. In automated mode, oxygen will be adjusted by the respiratory support device.
In automated mode (servo control), oxygen adjustments will be made by one of two devices (depending on the clinical requirements of the baby) - the IntellO2 device (IntellO2, Vapotherm) or Leoni plus CLAC (Closed-Loop Automated oxygen Control) ventilator (Leoni plus, Löwenstein Medical).
SpO2 and TcPO2 readings will be downloaded directly from the multiparameter patient monitor. SpO2 will be measured using a Phillips MX500 multiparameter monitor. TcPO2 will be measured using a SenTec Digital Monitoring System with OxiVent sensor. TcPO2 is calculated by dynamic fluorescence quenching which measures oxygen molecules present in the vicinity of a fluorescent dye incorporated within the sensor surface. The sensor is operated at a constant temperature of 43 degrees Celsius. Control of sensor temperature and application duration are designed to meet all applicable standards and this monitoring device is used routinely in many neonatal units.
Study Type
Enrollment (Actual)
Phase
- Not Applicable
Contacts and Locations
Study Locations
-
-
-
Edinburgh, United Kingdom, EH16 4SA
- The Simpson Centre for Reproductive Health, Royal Infirmary Edinburgh
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Description
Inclusion Criteria:
- Infants born at less than 29 weeks gestation
- Infants greater than 48 hours of age
- Infants who are receiving supplementary oxygen
- Person with parental responsibility able to give consent
Exclusion Criteria:
- Congenital anomalies that would prevent targeting SpO2 to 90-95% (e.g. cardiac defects)
- Clinical condition of an infant would impair accurateTcPO2 measurement (e.g. impaired perfusion or requirement of inotropic or vasopressor support)
Study Plan
How is the study designed?
Design Details
- Primary Purpose: Treatment
- Allocation: Randomized
- Interventional Model: Crossover Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
---|---|
Experimental: Servo control
Automated control of oxygen. The oxygen saturation target range will be set to 93%. Automated oxygen control can be over-ridden by manual adjustment of oxygen at any time if this is considered necessary to optimise control of oxygenation according to current clinical targets. |
FiO2 adjustments will be made by one of two respiratory devices depending on the clinical requirements and current ventilatory therapy of the infant
|
No Intervention: Manual control
Standard practice.
Oxygen adjustments will be made by clinical/nursing staff to maintain a target oxygen range of 90%-95%.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Incidence of hyperoxia and hypoxia on transcutaneous monitoring
Time Frame: 12 hours
|
To discover the percentage time spent within a TcPO2 range of 50mmHg (6.7kPa) - 80mmHg (10.7kPa) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
---|---|---|
Transcutaneous oxygen variability
Time Frame: 12 hours
|
To discover the variability in TcPO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Incidence of hyperoxia and hypoxia on saturation monitoring
Time Frame: 12 hours
|
To discover the percentage time spent within target SpO2 range of 90-95% when infants are targeted using automated (servo) versus manual control.
|
12 hours
|
Saturation variability
Time Frame: 12 hours
|
To discover the variability in SpO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Fraction of inspired oxygen variability
Time Frame: 12 hours
|
To discover the variability in FiO2 (measured by standard deviation) when infants are targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Pooled frequency histogram of TcPO2
Time Frame: 12 hours
|
To generate a pooled frequency histogram of percentage time at a TcPO2 of below 30mmHg, 30-39.9mmHg,
40-49.9mmHg,
50-59.9mmHg,
60-69.9mmHg,
70-79.9mmHg,
and 80mmHg and above for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Pooled frequency histogram of SpO2
Time Frame: 12 hours
|
To generate a pooled frequency histogram of percentage time at each SpO2 point between 80 - 100% for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Pooled frequency histogram of FiO2
Time Frame: 12 hours
|
To generate a pooled frequency histogram of the cumulative frequency at a FiO2 of 0.21-0.3,
0.31-0.4,
0.41-0.5,
0.51-0.6,
0.61-0.7,
0.81-0.9
and 0.91-1.0
for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Desaturations
Time Frame: 12 hours
|
To discover the frequency, duration and depth of desaturations and the area (change in PO2 versus time) above and below the set PO2 threshold for infants targeted to an SpO2 range of 90-95% using automated (servo) versus manual control.
|
12 hours
|
Collaborators and Investigators
Sponsor
Collaborators
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- AC19110
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Hypoxia
-
Western University, CanadaUnknownBrain Hypoxia IschemiaCanada
-
Yale UniversityTianjin Chest HospitalCompletedBrain Ischemia Hypoxia | Muscle; Ischemic | Muscle HypoxiaChina
-
Fliegerärztliches InstitutCompleted
-
University of Texas at AustinCompletedIntermittent HypoxiaUnited States
-
Owlet Baby Care, Inc.Completed
-
Robert L. OwensCompletedIntermittent HypoxiaUnited States
-
University of Texas at AustinCompleted
-
Harbin Medical UniversityUnknown
-
University Hospital TuebingenUnknown
-
Montreal Heart InstituteUniversité de MontréalCompleted
Clinical Trials on Servo control
-
University of OuluCompleted
-
Stanford UniversityInspiration HealthcareCompleted
-
University of OuluCompleted
-
ResMedCRI-The Clinical Research Institute GmbH; University Hospital RegensburgCompletedObstructive Sleep Apnea | Central Sleep Apnea | Mixed Sleep Apnea | Complex Sleep ApneaSwitzerland, Spain, Denmark, Portugal, France, Germany
-
University of EdinburghNHS LothianTerminatedHypoxia | Obstetric Labor, Premature | Hyperoxia | Premature Infant | Oxygen TherapyUnited Kingdom
-
Maquet Critical Care ABNAMSA; Paediatric and Neonatal Mechanical Ventilation (PNV) ConsultingNot yet recruitingElective HFOV for Respiratory Failure in Neonates/Infants | Rescue HFOV in Neonates and/Infants With Refractory Respiratory Failure Under Conventional Therapy
-
University Hospital PadovaCompletedNeonatal HypothermiaItaly
-
Société Française de Recherche et de Médecine du...Active, not recruitingCentral Sleep Apnea Syndrome in Patients Whose EF is Above 45 PercentFrance
-
University Hospital, MontpellierCompletedComplex Sleep Apnea Syndrome | Chronic Heart Failure and | Obstructive Sleep Apnea Syndrome and | Idiopathic Central Sleep Apnea Syndrome | Idiopathic Induced Periodic Breathing | Central Sleep Apnea SyndromeFrance
-
Veterans Medical Research FoundationResMed FoundationUnknownHeart Failure | Sleep Apnea, ObstructiveUnited States