PerQseal® Elite Venous Clinical Study (ELITE-Venous) (ELITE-Venous)

August 17, 2023 updated by: Vivasure Medical Limited

Safety and Performance Study for Venous Large Hole Vascular Closure Device - ELITE Study

The objective of this study is to evaluate the safety and effectiveness of the PerQseal Elite vascular closure system when used to achieve haemostasis of common femoral venotomies created by 14 to 22F sheaths (venotomy up to 26F) in patients undergoing percutaneous catheter-based interventional procedures, with sufficient rigor to provide robust scientific evidence for the demonstration of clinical safety and efficacy of the PerQseal Elite closure system to support a CE-mark and a PMA submission.

Study Overview

Status

Not yet recruiting

Intervention / Treatment

Detailed Description

This study is a prospective, multicentre, single-arm study to investigate the safety and efficacy of the PerQseal Elite in up to 97 patients at up to 12 European investigational sites. The study shall not be blinded prior to, during or post the procedure. Patients undergoing a percutaneous catheter-based venous interventional procedure (e.g., mitral valve repair, leadless pacemaker implantation, circulatory assist or ECMO via large bore femoral vein catheter) requiring a venotomy created by 14 to 22F sheaths (venotomies up to 26F), via the common femoral vein will be screened against the study inclusion/exclusion criteria.

All subjects shall have a 30-minute, 60-minute, pre-discharge (~48 hours), 30-Days ± 7-Days and 180-Days ± 30-Days follow-up assessments.

All safety data from the study will be assessed by the Data Safety Monitoring Committee on a continuous basis.

Study Type

Interventional

Enrollment (Estimated)

90

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

  • Adult
  • Older Adult

Accepts Healthy Volunteers

Yes

Description

Inclusion Criteria:

  1. Age ≥ 19 years.
  2. Clinically indicated for a percutaneous venous interventional catheter-based procedure, e.g., mitral valve edge to edge repair, left atrial appendage device closure, leadless pacemaker implantation, patent foramen ovale closure or use of a circulatory assist device or extracorporeal oxygenation using a common femoral venotomy created by a 14 to 22F sheath (venotomy up to 26F).
  3. Subject is willing and able to provide appropriate study-specific informed consent, follow protocol procedures, and comply with follow-up visit requirements.
  4. Females who are not pregnant or lactating and not planning to become pregnant for the duration of the study.

Baseline Exclusion Criteria:

  1. Evidence of current systemic bacterial or cutaneous infection, including groin infection,
  2. Known bleeding diathesis, unstable INR, definite or potential coagulopathy, platelet count < 100,000/μl or subjects on long term anticoagulants with an INR > 2.2 within 12 hours prior to index procedure,
  3. Significant anaemia (example: haemoglobin < 8 g/dL or haematocrit < 22%), within 24 hours prior to index procedure,
  4. Known type II heparin-induced thrombocytopenia,
  5. Documented right ventricular dysfunction < 13%,
  6. Ipsilateral or contralateral lower extremity amputation,
  7. Previous groin surgery within the region of the ipsilateral vessel access,
  8. Common femoral vein diameter stenosis > 50% or previous bypass surgery/stent placement in the common femoral vein of target leg,
  9. Known existing nerve damage in the target leg,
  10. Nephrotic syndrome or renal insufficiency (Glomerular Filtration Rate ≤ 30 ml/min or baseline serum creatinine > 2.5 mg/dl) or on renal replacement therapy or albumin level < 3.5 g/dl or symptoms of pitting oedema,
  11. Systolic pulmonary artery pressure > 60 mmHg
  12. Known allergy to any of the materials used in the PerQseal Elite device (refer to Investigator's Brochure),
  13. Subject is unsuitable for surgical repair of the target leg access site,
  14. Subject has undergone a percutaneous procedure greater than 8F sheath in the target vessel, within the 90-days prior to index procedure,
  15. Subject has undergone a percutaneous procedure of 8F sheath or less using an absorbable intravascular closure device for haemostasis, in the target vessel, within the 90-days prior to index procedure,
  16. Subject has undergone a percutaneous procedure of 8F sheath or less using a suture mediated closure device or manual/mechanical pressure for haemostasis in the target vessel, within the 30-days prior to index procedure,
  17. History of DVT or pulmonary embolism or venous thrombotic event,
  18. Splenectomy or suffering psoriasis or paraesthesia of the ipsilateral leg,
  19. Further planned endovascular/catheter-based procedure in the target leg in the 30-days following the index procedure,
  20. Subject is enrolled in another investigational medical device or drug study,
  21. Subject has been previously enrolled in this clinical study,
  22. Subject is unable to maintain ipsilateral leg still during the index procedure (restless leg syndrome),
  23. Current COVID-19 infection (with or without symptoms), recent positive test for COVID-19 within two weeks, or recent exposure to a person with COVID-19 infection within two weeks.

    Procedural Exclusion Criteria:

  24. Anatomically the common femoral vein is substantially posterior to the femoral artery, confirmed by Duplex ultrasound,
  25. Initial common femoral venous access achieved other than with the use of an ultrasound guided access approach,
  26. If the venous access is, or suspected of being, via an artery,
  27. Difficult dilation during initial target femoral vein access (e.g., that damages or kinks dilators) while step-dilating up to the large-bore device,
  28. During venous puncture, the target femoral vein suspected to have experienced a posterior venous wall needle puncture or underwent > one needle puncture during the primary procedure, with a needle larger than a micropuncture needle (> 21 gauge or > 0.819 mm diameter). (Note: not an exclusion if micropuncture technique under ultrasound guidance used for femoral vein access [Ref. 30, 31]),
  29. Subject has a tissue tract (i.e., estimated distance from skin entry point to venous anterior surface at venotomy) expected to be greater than 8 cm,
  30. Significant blood loss requiring transfusion of blood products during primary procedure or within 30-days prior to index procedure,
  31. Activated clotting time (ACT) > 250 seconds immediately prior to sheath removal or if ACT measurements are expected to be > 250 seconds for more than 24 hours after index procedure,
  32. Target puncture site is in a vascular graft,
  33. Target venotomy greater than 26F,
  34. Target femoral vein diameter is less than 7 mm, based on Duplex ultrasound or angiography,
  35. Evidence of venous diameter stenosis > 20% within 15 mm proximal or distal to venotomy site based on Duplex ultrasound or angiography,
  36. Target venotomy has been made in the superficial femoral vein, profunda femoris vein (deep femoral vein), great saphenous vein or located less than 15 mm distal (cranial) to the bifurcation of either the profunda femoris or great saphenous vein and common femoral vein,
  37. Target venotomy located behind (posterior) or above (cranial) to the inguinal ligament based upon bony landmarks (above femoral head on A-P projection),
  38. Subjects with an acute haematoma > 4 cm diameter, arteriovenous fistula, pseudoaneurysm or intraluminal thrombosis at the target access site identified intra-procedurally,
  39. Evidence of bleeding around the primary procedure sheath (VARC type 1/BARC type 2 or higher),
  40. Intra-procedural (angiographic or Duplex ultrasound) evidence of venous laceration, dissection or stenosis within the femoral vein that would preclude use of the PerQseal Elite device,

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Treatment
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Investigational Device Arm (PerQseal Elite- Venous)
Large hole percutaneous venous closure device - PerQseal Elite
The term PerQseal Elite used singularly refers to the closure system comprising the closure device and Introducer. These individually are referred to as: "PerQseal Elite VCD" or "PerQseal Elite closure device" and/or "PerQseal Elite Introducer".
Other Names:
  • Percutaneous ELITE Venous Closure Device

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Primary Safety Endpoint- Combined rate of both Major and Minor access site complications
Time Frame: through 30 Days
the combined rate of both Major and Minor access site complications attributable to the PerQseal Elite device§ through 30-days, as adjudicated by the Clinical Events Committee (CEC) as study device related.
through 30 Days
Primary Effectiveness Endpoint - Time to Haemostasis
Time Frame: 20 Minutes
Time to Haemostasis (TTH) defined as elapsed time in minutes from PerQseal Elite (Introducer-sheath and Delivery-device) removal to first observed cessation of common femoral vein (CFV) bleeding (clinically acceptable cessation of venous bleeding), excluding cutaneous or subcutaneous oozing, and in the absence of a developing haematoma. This is assessed at the end of the index procedure prior leaving the operating room/catheterisation laboratory.
20 Minutes

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Secondary Effectiveness Endpoint - Total Closure Time
Time Frame: 40 Minutes
Total Closure Time is defined as: Time to Haemostasis (TTH) plus Time to Device Deployment (TTDD).
40 Minutes
Secondary Effectiveness Endpoint - PerQseal Elite Technical Success Rate
Time Frame: 10 Minutes
PerQseal Elite Technical Success Rate is defined as the number of PerQseal Elite devices that are deployed and achieve haemostasis (i.e., cessation of venous bleeding excluding cutaneous or subcutaneous oozing) without need for more than 10 minutes of firm, adjunctive manual compression or alternative treatment (other than adjunctive endovascular ballooning) at target access site, divided by the total number of PerQseal Elite devices where deployment was attempted.
10 Minutes
Secondary Effectiveness Endpoint - PerQseal Elite Treatment Success Rate
Time Frame: through 30 days
PerQseal Elite Treatment Success Rate is defined as the number of subjects who meet PerQseal Elite Technical Success without experiencing a device related Access Site Complication (based on CEC adjudication) through 30-days, divided by the total number of subjects where PerQseal Elite device deployment was attempted.
through 30 days
Secondary Effectiveness Endpoint -Time to Device Deployment (TTDD)
Time Frame: 20 Minutes
Time to Device Deployment is defined as the time from insertion of the PerQseal Elite delivery device into the PerQseal Elite Introducer sheath to complete removal of the PerQseal Elite (delivery device and introducer-sheath) from the subject following PerQseal Elite deployment.
20 Minutes
Secondary Effectiveness Endpoint - Time to Ambulation
Time Frame: through 24 hours
Time to Ambulation is defined as: the elapsed time from PerQseal Elite removal from the subject post deployment until the subject can walk 20 feet/6 meters without venous re-bleeding at the target access site. This endpoint will be evaluated only for subjects who are ambulatory (e.g., not confined to wheelchair or bed) at the time of enrolment.
through 24 hours

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Estimated)

October 1, 2023

Primary Completion (Estimated)

April 1, 2024

Study Completion (Estimated)

October 1, 2024

Study Registration Dates

First Submitted

August 17, 2023

First Submitted That Met QC Criteria

August 17, 2023

First Posted (Actual)

August 23, 2023

Study Record Updates

Last Update Posted (Actual)

August 23, 2023

Last Update Submitted That Met QC Criteria

August 17, 2023

Last Verified

August 1, 2023

More Information

Terms related to this study

Other Study ID Numbers

  • P934-00

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Large Hole Percutaneous Venous Closure

Clinical Trials on PerQseal ELITE

3
Subscribe