Selecting the Best Ventilator Hyperinflation Settings (VHI1)
Selecting the Best Ventilator Hyperinflation Settings Based on Physiologic Markers: Randomized Controlled Study
Panoramica dello studio
Stato
Condizioni
Intervento / Trattamento
Descrizione dettagliata
Background: Ventilator Hyperinflation (VHI) has been shown to be effective in improving respiratory mechanics, secretion removal, and gas exchange in mechanically ventilated patients; however, there are no recommendations on the best ventilator settings to perform the technique. Thus, the aim of this study was to compare six modes of VHI, concerning physiological markers of efficacy and safety criteria, in order to support the optimal VHI settings selection for mechanically ventilated patients.
Methods: In a crossover study, every included mechanically ventilated patient underwent six modes of VHI in a randomized order: Volume Control Continuous Mandatory Ventilation (VC-CMV) with inspiratory flow = 20Lpm (VC-CMV20), VC-CMV with inspiratory flow = 50Lpm (VC-CMV50), Pressure Control Continuous Mandatory Ventilation (PC-CMV) with inspiratory time = 1s. (PC-CMV1), PC-CMV with inspiratory time = 3s. (PC-CMV3), Pressure Support Ventilation (PSV) with cycling off = 10% of peak inspiratory flow (PSV10), and PSV with cycling off = 25% of peak inspiratory flow (PSV25). The maximum expansion (tidal volume), expiratory flow bias criteria (inspiratory and expiratory flow patterns), over-distension (alveolar pressure), asynchronies and hemodynamic variables (mean arterial pressure and heart rate) were assessed during the interventions.
Tipo di studio
Iscrizione (Effettivo)
Fase
- Non applicabile
Criteri di partecipazione
Criteri di ammissibilità
Età idonea allo studio
Accetta volontari sani
Sessi ammissibili allo studio
Descrizione
Inclusion Criteria:
- Patients under mechanical ventilation for more than 48h
Exclusion Criteria:
- mucus hypersecretion (defined as the need for suctioning < 2-h intervals),
- absence of respiratory drive,
- atelectasis,
- severe bronchospasm,
- positive end expiratory pressure > 10cmH2O,
- PaO2-FiO2 relationship < 150,
- mean arterial pressure < 60mmHg,
- inotrope requirement equivalent to >15 ml/h total of adrenaline and noradrenalin,
- intracranial pressure > 20mmHg
Piano di studio
Come è strutturato lo studio?
Dettagli di progettazione
- Scopo principale: Trattamento
- Assegnazione: Randomizzato
- Modello interventistico: Assegnazione incrociata
- Mascheramento: Separare
Armi e interventi
Gruppo di partecipanti / Arm |
Intervento / Trattamento |
|---|---|
|
Nessun intervento: BASELINE
The subjects were kept in their current ventilatory mode.
|
|
|
Sperimentale: VC-CMV20
Application of a ventilator hyperinflation intervention with Volume Control Continuous Mandatory Ventilation (VC-CMV) with an inspiratory flow of 20Lpm.
|
Application of a ventilator hyperinflation intervention with Volume Control Continuous Mandatory Ventilation (VC-CMV).
The inspiratory flow was set at 20Lpm and the tidal volume was increased in steps of 200mL until the peak airway pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
|
Sperimentale: VC-CMV50
Application of a ventilator hyperinflation intervention with Volume Control Continuous Mandatory Ventilation (VC-CMV) with an inspiratory flow of 50Lpm.
|
Application of a ventilator hyperinflation intervention with Volume Control Continuous Mandatory Ventilation (VC-CMV).
The inspiratory flow was set at 50Lpm and the tidal volume was increased in steps of 200mL until the peak airway pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
|
Sperimentale: PC-CMV1
Application of a ventilator hyperinflation intervention with Pressure Control Continuous Mandatory Ventilation (PC-CMV1) with an inspiratory time of 1 second.
|
Application of a ventilator hyperinflation intervention with Pressure Control Continuous Mandatory Ventilation (PC-CMV1).
The inspiratory time was set at 1 second and the pressure control was increased until a peak pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
|
Sperimentale: PC-CMV3
Application of a ventilator hyperinflation intervention with Pressure Control Continuous Mandatory Ventilation (PC-CMV1) with an inspiratory time of 3 seconds.
|
Application of a ventilator hyperinflation intervention with Pressure Control Continuous Mandatory Ventilation (PC-CMV1).
The inspiratory time was set at 3 seconds and the pressure control was increased until a peak pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
|
Sperimentale: PSV10
Application of a ventilator hyperinflation intervention with Pressure Support Ventilation (PSV) with a cycling off of 10% of peak inspiratory flow.
|
Application of a ventilator hyperinflation intervention with Pressure Support Ventilation (PSV).
The cycling off was set at 10% of peak inspiratory flow and the pressure support was increased until a peak pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
|
Sperimentale: PSV25
Application of a ventilator hyperinflation intervention with Pressure Support Ventilation (PSV) with a cycling off of 25% of peak inspiratory flow.
|
Application of a ventilator hyperinflation intervention with Pressure Support Ventilation (PSV).
The cycling off was set at 25% of peak inspiratory flow and the pressure support was increased until a peak pressure of 40cmH2O was achieved.
After achieving the target pressure, this ventilatory regimen lasted 15 minutes.
Positive end expiratory pressure and the inspired oxygen fraction were not modified.
|
Cosa sta misurando lo studio?
Misure di risultato primarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Peak inspiratory to expiratory flow ratio
Lasso di tempo: Ten minutes after the onset of intervention.
|
Dichotomous variable, defined as achieving a peak inspiratory flow rate (PIFR) less than 90% of the peak expiratory flow rate (PEFR)
|
Ten minutes after the onset of intervention.
|
|
Peak expiratory flow higher than 40 Lpm
Lasso di tempo: Ten minutes after the onset of intervention.
|
Dichotomous variable, defined as achieving a PEFR higher than 40 l/min
|
Ten minutes after the onset of intervention.
|
|
Difference between peak inspiratory and expiratory flows.
Lasso di tempo: Ten minutes after the onset of intervention.
|
Dichotomous variable, defined as achieving a difference higher than 17Lpm.
|
Ten minutes after the onset of intervention.
|
|
Pulmonary expansion
Lasso di tempo: Ten minutes after the onset of intervention.
|
Percentage of tidal volume above the normal tidal volume (estimated as 6mL/kg).
|
Ten minutes after the onset of intervention.
|
Misure di risultato secondarie
Misura del risultato |
Misura Descrizione |
Lasso di tempo |
|---|---|---|
|
Mean arterial pressure
Lasso di tempo: Ten minutes after the onset of intervention.
|
Mean arterial pressure verified using the multi-parameter monitor.
|
Ten minutes after the onset of intervention.
|
|
Heart Rate
Lasso di tempo: Ten minutes after the onset of intervention.
|
Heart rate verified using the multi-parameter monitor.
|
Ten minutes after the onset of intervention.
|
Collaboratori e investigatori
Collaboratori
Investigatori
- Cattedra di studio: FERNANDO S GUIMARAES, PhD, Centro Universitário Augusto Motta
Pubblicazioni e link utili
Pubblicazioni generali
- Berney S, Denehy L. A comparison of the effects of manual and ventilator hyperinflation on static lung compliance and sputum production in intubated and ventilated intensive care patients. Physiother Res Int. 2002;7(2):100-8. doi: 10.1002/pri.246.
- Lemes DA, Zin WA, Guimaraes FS. Hyperinflation using pressure support ventilation improves secretion clearance and respiratory mechanics in ventilated patients with pulmonary infection: a randomised crossover trial. Aust J Physiother. 2009;55(4):249-54. doi: 10.1016/s0004-9514(09)70004-2.
- Thomas PJ. The effect of mechanical ventilator settings during ventilator hyperinflation techniques: a bench-top analysis. Anaesth Intensive Care. 2015 Jan;43(1):81-7. doi: 10.1177/0310057X1504300112.
- Ntoumenopoulos G, Shannon H, Main E. Do commonly used ventilator settings for mechanically ventilated adults have the potential to embed secretions or promote clearance? Respir Care. 2011 Dec;56(12):1887-92. doi: 10.4187/respcare.01229. Epub 2011 Jun 17.
- Anderson A, Alexanders J, Sinani C, Hayes S, Fogarty M. Effects of ventilator vs manual hyperinflation in adults receiving mechanical ventilation: a systematic review of randomised clinical trials. Physiotherapy. 2015 Jun;101(2):103-10. doi: 10.1016/j.physio.2014.07.006. Epub 2014 Oct 6.
- Davies JD, Senussi MH, Mireles-Cabodevila E. Should A Tidal Volume of 6 mL/kg Be Used in All Patients? Respir Care. 2016 Jun;61(6):774-90. doi: 10.4187/respcare.04651.
- de Wit M. Monitoring of patient-ventilator interaction at the bedside. Respir Care. 2011 Jan;56(1):61-72. doi: 10.4187/respcare.01077.
Collegamenti utili
Studiare le date dei record
Studia le date principali
Inizio studio (Effettivo)
Completamento primario (Effettivo)
Completamento dello studio (Effettivo)
Date di iscrizione allo studio
Primo inviato
Primo inviato che soddisfa i criteri di controllo qualità
Primo Inserito (Effettivo)
Aggiornamenti dei record di studio
Ultimo aggiornamento pubblicato (Effettivo)
Ultimo aggiornamento inviato che soddisfa i criteri QC
Ultimo verificato
Maggiori informazioni
Termini relativi a questo studio
Termini MeSH pertinenti aggiuntivi
Altri numeri di identificazione dello studio
- VHI1
Piano per i dati dei singoli partecipanti (IPD)
Hai intenzione di condividere i dati dei singoli partecipanti (IPD)?
Informazioni su farmaci e dispositivi, documenti di studio
Studia un prodotto farmaceutico regolamentato dalla FDA degli Stati Uniti
Studia un dispositivo regolamentato dalla FDA degli Stati Uniti
Queste informazioni sono state recuperate direttamente dal sito web clinicaltrials.gov senza alcuna modifica. In caso di richieste di modifica, rimozione o aggiornamento dei dettagli dello studio, contattare register@clinicaltrials.gov. Non appena verrà implementata una modifica su clinicaltrials.gov, questa verrà aggiornata automaticamente anche sul nostro sito web .
Prove cliniche su Disturbi respiratori
-
Charite University, Berlin, GermanyReclutamentoSchizofrenia Spectrum Disorders (SSD)Germania
-
University of Trás-os-Montes and Alto DouroCentro Hospitalar De Trás-Os-Montes E Alto Douro, E.P.E.ReclutamentoDisturbi psicotici | Grave malattia mentale | Schizofrenia Spectrum Disorders (SSD)Portogallo
-
Central Institute of Mental Health, MannheimReclutamentoSchizofrenia Spectrum Disorders (SSD)Germania
-
Tianjin Medical University General HospitalReclutamentoNeuromielite Optica Spectrum Disorders (NMOSD)Cina
-
Tianjin Medical University General HospitalNon ancora reclutamentoNeuromielite ottica (NMO) | Neuromielite Optica Spectrum Disorders (NMOSD)
-
Huashan HospitalNon ancora reclutamentoNeuromielite Optica Spectrum Disorders (NMOSD)Cina
-
Huashan HospitalCompletato
-
Huashan HospitalShanghai AbelZeta Ltd.Non ancora reclutamentoEncefalite autoimmune | Sclerosi multipla (SM) | Sindrome della persona rigida | Neuromielite Optica Spectrum Disorders (NMOSD)
-
RenJi HospitalAbelZeta Pharma Inc.ReclutamentoMiastenia grave | Lupus eritematoso sistemico (LES) | Sclerosi multipla (SM) | Sclerosi sistemica (SSc) | Miopatia necrotizzante immunomediata (IMNM) | Neuromielite Optica Spectrum Disorders (NMOSD)Cina
Prove cliniche su VC-CMV20
-
National Defense Medical Center, TaiwanCompletatoDolore, Acuto | Comportamento infantile | Stress fisiologico | Latte maternoTaiwan
-
ViaCyteCalifornia Institute for Regenerative Medicine (CIRM)TerminatoDiabete mellito di tipo 1Canada, Stati Uniti
-
ViaCyteTerminato
-
Queen's University, BelfastAWARE NI; Northern Ireland Public Health Research NetworkCompletato
-
Sunnybrook Health Sciences CentreCanadian Initiative for Outcomes in Rheumatology CareTerminato
-
VA Office of Research and DevelopmentNon ancora reclutamentoBroncopneumopatia cronica ostruttivaStati Uniti
-
ViaCyteCompletatoDiabete mellito di tipo 1Canada
-
ViaCyteVertex Pharmaceuticals Incorporated; California Institute for Regenerative Medicine... e altri collaboratoriCompletatoDiabete mellito di tipo 1 con ipoglicemiaStati Uniti, Canada, Belgio
-
University Medical Center GroningenTerminato
-
International Centre for Diarrhoeal Disease Research...ReclutamentoProva di non inferioritàBangladesh