A Study to Evaluate a Multiple Model Probabilistic Predictive Controller (MMPPC) for Closed Loop Insulin Delivery
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
You are invited to participate in a research study for the development of an artificial pancreas. An artificial pancreas uses a program which takes information from a continuous blood glucose monitor and uses that information to tell an insulin infusion pump how much insulin to deliver. The primary purpose of this study is to gain experience with insulin delivery algorithms or programs program (algorithm) provides the best regulation of glucose levels so that there are no severe low blood glucose reactions and blood glucose levels are generally between 70 to 180 mg/dl. If the system is working properly, you would not need to enter the amount of food you were eating, give an insulin bolus, or change your basal rates while wearing the device. You would need to periodically check to be sure the continuous glucose sensor was functioning properly and you would need to respond to alarms that might occur if your blood glucose was too high, too low, or the glucose sensor or pump were not working well. In addition to this the investigators will also gain experience with insulin delivery algorithms to minimize the number of glucose readings which are above or below target. It is our intention to modify the algorithms during these studies.
In this study the investigators plan to use a commercially available insulin infusion pump (OmniPod) to deliver lispro (Humalog) insulin. The investigators will use Navigator continuous glucose sensors both to monitor glucose levels (sensor 1) and provide the glucose concentration for the closed loop algorithm (sensor 2). The signal from the second Navigator will be sent by serial cable to a computer which will be at the patient's bedside. A control algorithm will reside on the computer, and the amount of insulin to be delivered will be transmitted to the OmniPod Personal Device Manager which will then send a radiofrequency (rf) signal to the Omnipod pump residing on the subject. A health care provider will be in attendance and monitoring discrete blood glucose levels (YSI, GlucoScout, or HemoCue measurements) at least every 30 minutes. While the Navigator and Omnipod are commercially available they will be used in this study as part of an investigational system while you are in the hospital, an investigational system is one that is not approved for use by the FDA.
Undersøgelsestype
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
California
-
Stanford, California, Forenede Stater, 94305
- Stanford University School of Medicine
-
Stanford, California, Forenede Stater, 94305
- CTRU located in Blake Wilbur
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
Tager imod sunde frivillige
Køn, der er berettiget til at studere
Beskrivelse
Inclusion Criteria:
Eligibility
To be eligible for the study, all subjects must meet the following criteria:
- Clinical diagnosis of type 1 diabetes and using daily insulin therapy for at least one year. The diagnosis of type 1 diabetes is based on the investigator's judgment; Cpeptide level and antibody determinations are not needed.
- Age 21 years to less than 45.0 years
- Subject has used a downloadable insulin pump for at least 3 months
- Subject understands the study protocol and agrees to comply with it
- Informed Consent Form signed
- A Personal Home computer with internet access (must have access to a PC for uploading, not a Mac).
Exclusion Criteria:
Exclusion
Subjects who meet any of the following criteria are not eligible for the study:
- The presence of a significant medical disorder that in the judgment of the investigator will affect the wearing of the sensors or the completion of any aspect of the protocol.
The presence of any of the following diseases:
- Asthma if treated with systemic or inhaled corticosteroids in the last 6 months
- Cystic fibrosis
- Other major illness that in the judgment of the investigator might interfere with the completion of the protocol Adequately treated thyroid disease and celiac disease do not exclude subjects from enrollment
- Inpatient psychiatric treatment in the past 6 months
- Current use of oral/inhaled glucocorticoids or other medications, which in the judgment of the investigator would be a contraindication to participation in the study.
- Pregnancy, breast feeding, or intention of becoming pregnant in the next 2 months.
- Weight less than 26 kg
- Renal failure or peritoneal dialysis
- History of heart disease
- The use of beta-blockers
- History of cerebrovascular disease, or non-hypoglycemic seizures, or intolerance of glucagon treatment.
- History of a hypoglycemic seizure within 6 months of enrollment.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Behandling
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
---|---|
Eksperimentel: closed-loop control
Multiple Model Predictive Controller
|
Use of closed loop algorithm in an inpatient environment while closely monitoring blood glucose levels.
Andre navne:
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
---|---|---|
Test the feasibility of using MMPPC controller for closed loop insulin delivery in a monitored inpatient clinical research environment.
Tidsramme: 36 Hour Admission
|
Safety: 1) No reference glucose values <50 mg/dl, and no more than 30 minutes with reference glucose values between 50-60 mg/dl per day based on linear interpolation between values 3) No reference glucose value >250 mg/dl outside of the 3 hours following a meal.
|
36 Hour Admission
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
---|---|---|
Efficacy
Tidsramme: 30 hour admission
|
The percent of time spent between 70 mg/dl and 180 mg/dl, mean and standard deviation of measurements.
|
30 hour admission
|
Samarbejdspartnere og efterforskere
Sponsor
Datoer for undersøgelser
Studer store datoer
Studiestart
Primær færdiggørelse (Faktiske)
Studieafslutning (Faktiske)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Skøn)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Skøn)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- SU-09222010-6910
- SPO: 47012 (Andet bevillings-/finansieringsnummer: JDRF 22-2009-812)
- eProtocol: 8566 (Andet bevillings-/finansieringsnummer: JDRF 22-2009-812)
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Type 1 diabetes mellitus
-
University of Colorado, DenverMassachusetts General Hospital; Beta Bionics, Inc.AfsluttetDiabetes mellitus, type 1 | Type 1 diabetes | Diabetes type 1 | Type 1 diabetes mellitus | Autoimmun diabetes | Diabetes mellitus, insulinafhængig | Juvenil-Debut Diabetes | Diabetes, autoimmun | Insulinafhængig diabetes mellitus 1 | Diabetes mellitus, insulinafhængig, 1 | Diabetes mellitus, skør | Diabetes Mellitus... og andre forholdForenede Stater
-
University of California, San FranciscoJuvenile Diabetes Research FoundationAfsluttetType 1 diabetes mellitus | Diabetes mellitus, type I | Insulinafhængig diabetes mellitus 1 | Diabetes mellitus, insulinafhængig, 1 | IDDMForenede Stater, Australien
-
Capillary Biomedical, Inc.AfsluttetDiabetes mellitus, type 1 | Type 1 diabetes | Type 1 diabetes mellitus | Diabetes mellitus, insulinafhængig, 1Australien
-
Spiden AGDCB Research AGRekrutteringType 1 diabetes mellitus | Type 1 diabetes mellitus med hypoglykæmi | Type 1 diabetes mellitus med hyperglykæmiSchweiz
-
Capillary Biomedical, Inc.AfsluttetType 1 diabetes | Type 1 diabetes mellitus | Diabetes mellitus, type I | Diabetes mellitus, insulinafhængig, 1 | IDDMØstrig
-
Instytut Diabetologii Sp. z o.o.National Center for Research and Development, Poland; Nalecz Institute...UkendtType 1 diabetes mellitus med hyperglykæmi | Type 1 diabetes mellitus med hypoglykæmiPolen
-
National Institute of Allergy and Infectious Diseases...PPD; Rho Federal Systems Division, Inc.; Immune Tolerance Network (ITN)AfsluttetType 1 diabetes mellitus | T1DM | T1D | Nyopstået type 1-diabetes mellitusForenede Stater, Australien
-
National Institute of Diabetes and Digestive and...AfsluttetDiabetes mellitus type 1Forenede Stater
-
Stem Cells ArabiaAfsluttetGlukosemetabolismeforstyrrelser | Metaboliske sygdomme | Sygdomme i immunsystemet | Autoimmune sygdomme | Sygdomme i det endokrine system | Diabetes mellitus type 1Jordan
-
Shanghai Changzheng HospitalRekrutteringSkør type 1 diabetes mellitusKina
Kliniske forsøg med Multiple Model Predictive Controller
-
University of ParmaAfsluttetMotorisk aktivitet | Hemiplegisk cerebral pareseItalien
-
University of VirginiaJuvenile Diabetes Research FoundationAfsluttetType 1 diabetesForenede Stater
-
University of VirginiaNational Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)Afsluttet
-
donald whitingBoston Scientific CorporationRekruttering
-
Sansum Diabetes Research InstituteUniversity of California, Santa BarbaraAfsluttetType 1 diabetes mellitusForenede Stater