The Sunrise SpO2 and Pulse Rate Accuracy Validation Study (S2-SOPRAV)

June 22, 2022 updated by: Sunrise

Single-center Study Assessing the SpO2 and Pulse Rate Accuracy of the Sunrise Sensor 2 Compared to Arterial Blood Samples Oxygen Saturation Assessed by CO-oximetry and ECG Respectively, Under Non-motion Conditions in Healthy Adults

The purpose of this study is to evaluate the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter. Testing is conducted under normal office environment conditions. It is expected that the Accuracy Root Mean Square (ARMS) performance of the oximetry system will meet the required specification of ARMS of 3.5% or less allowed for reflectance technology in non-motion conditions for the range of 70-100% SaO2.

A secondary goal is to evaluate the pulse rate performance simultaneously collected over the SpO2 range covered.

Study Overview

Status

Completed

Conditions

Detailed Description

Subjects will be given medical grade mixtures of oxygen and nitrogen to induce stable plateaus across the range of 100% to 70% SaO2. When the reference system's blood saturation stabilizes at an acceptable plateau level, blood sampling can start. Four to eight arterial blood samples will be collected for each plateau, while keeping at least 20 seconds in between each sample and at least 30 seconds once the plateau has been reached. The blood will be immediately analyzed by the reference CO-Oximeter to measure the arterial oxygen saturation (Functional SaO2). SpO2 values from the Sunrise sensor 2 will be collected electronically simultaneously to blood drawn from the indwelling catheter. Data will be collected under non-motion conditions. The pulse rate will be simultaneously collected over the SpO2 range covered. The reference method for the computation of pulse rate accuracy will be an ECG heart rate. The total duration of the test will be about 1 hour per subject. Data analysis results will provide documentation showing SpO2 and pulse rate accuracy performance of the Sunrise sensor 2 as compared to arterial blood samples measured by Reference CO-Oximetry and to ECG heart rate, respectively.

Study Type

Interventional

Enrollment (Actual)

12

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Locations

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

16 years to 48 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Subjects with positive Allen's test
  • Subject must be willing and able to comply with study procedures and duration
  • Subject is a non-smoker or who has not smoked within 2 days prior to the study
  • 15% of the study population or 2 subjects (whichever is larger) with dark pigmented skin should be included in the study

Exclusion Criteria:

  • Subject is considered as being morbidly obese (defined as BMI >39.5)
  • Females who are pregnant, who are trying to get pregnant, or have a urine test positive for pregnancy on the day of the study
  • Subjects with known heart or cardiovascular conditions
  • Subjects with other known health condition (diabetes, thyroid disease, kidney disease/chronic renal impairment, history of seizures (except childhood febrile seizures) or epilepsy, or unexplained syncope, recent history of frequent migraine headaches, recent head injury, cancer/chemotherapy)

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Diagnostic
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: Healthy adult volunteers
All subjects within this single arm of the study will undergo the validation experiment as described in the description
Evaluation of the SpO2 accuracy performance of the Sunrise Sensor 2 placed at the tip of the chin during non-motion conditions over the range of 70-100% by comparison to SaO2 values determined from arterial blood sample specimen analyzed by a CO-oximeter.

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
SpO2 ARMS determination
Time Frame: Within 1 month after subject recruitment
SpO2 accuracy root-mean-square (ARMS) determination for the range of 70- 100%., i.e., the RMS difference between measured values and reference values.
Within 1 month after subject recruitment

Secondary Outcome Measures

Outcome Measure
Measure Description
Time Frame
Pulse rate ARMS determination
Time Frame: Within 1 month after subject recruitment
Pulse rate accuracy root-mean-square (ARMS) determination
Within 1 month after subject recruitment

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Sponsor

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Actual)

May 24, 2022

Primary Completion (Actual)

June 21, 2022

Study Completion (Actual)

June 21, 2022

Study Registration Dates

First Submitted

May 18, 2022

First Submitted That Met QC Criteria

May 18, 2022

First Posted (Actual)

May 23, 2022

Study Record Updates

Last Update Posted (Actual)

June 28, 2022

Last Update Submitted That Met QC Criteria

June 22, 2022

Last Verified

June 1, 2022

More Information

Terms related to this study

Other Study ID Numbers

  • PRO-367

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

No

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

No

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Validation Study

Clinical Trials on Pulse oximeter validation population