Esta página se tradujo automáticamente y no se garantiza la precisión de la traducción. por favor refiérase a versión inglesa para un texto fuente.

Prospective Trial for Clinical Validation of "QOCA Disposable BLE Thermometer "

21 de julio de 2022 actualizado por: Chih-Jung Chen, Chang Gung Memorial Hospital
The purpose of this trial is to evaluate the accuracy and safety of the "QOCA Disposable Body Temperature Patch" (Q-temp-w1).

Descripción general del estudio

Estado

Reclutamiento

Condiciones

Descripción detallada

Q-temp-w1 is a contact type body temperature patch developed by Quanta Computer Co., Ltd. It is mainly used to measure the body temperature of the human body. It can be used for general body temperature measurement, home temperature measurement records and medical tracking. Both the European Union and the US FDA state that the clinical test methods of the temperature patch must comply with ISO 80601-2-56. Therefore, this test plan will follow the ISO 80601-2-56 criteria to evaluate the accuracy of the temperature patch and follow it. The sampling age range and the fever/non-fever ratio are stratified, and clinical efficacy analysis is performed based on the measurement results. The main evaluation indicators include clinical bias, limits of agreement, and clinical repeatability.

Tipo de estudio

Intervencionista

Inscripción (Anticipado)

133

Fase

  • No aplica

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Estudio Contacto

Copia de seguridad de contactos de estudio

  • Nombre: Tien-Yun Huang, Masters
  • Número de teléfono: +886975056034
  • Correo electrónico: ivyhuang@cgmh.org.tw

Ubicaciones de estudio

      • Taoyuan, Taiwán, 333
        • Reclutamiento
        • ChangGungMH
        • Contacto:
          • Tien-Yun Huang, Masters
          • Número de teléfono: +886975056034
          • Correo electrónico: ivyhuang@cgmh.org.tw
        • Contacto:
        • Investigador principal:
          • Chin-Jung Chen, MD
        • Sub-Investigador:
          • Ching-Tai Huang, MD
        • Sub-Investigador:
          • Ming-Chou Chiang, MD

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

1 día a 100 años (Niño, Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Todos

Descripción

Inclusion Criteria:

  • 1. Male or female aged between 0-100 years.
  • 2. Patient agrees to participate in the trial and sign informed consent.

Exclusion Criteria:

  • 1. Those who have any broken skin or skin diseases on the part of the body which is the test site of the BLE thermometer, Including the underarms and the back of the body.
  • 2. Those whose test site of the BLE thermometer are known to be contaminated by uncle arable body fluids, or lead to the inaccurate of the BLE thermometer.
  • 3. Those who have restlessness or other conditions that make them unable to tolerate the temperature measurement procedure.

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Otro
  • Asignación: N / A
  • Modelo Intervencionista: Asignación de un solo grupo
  • Enmascaramiento: Ninguno (etiqueta abierta)

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: Single group

This study is specially designed according to its age range and customized/non-customized range, and meets the requirements of ISO 80601-2-56 test population (table). Comply with the age range and population requirements of ISO 80601-2-56, as well as clinical and subject trials, including news to the population over five years old, and hopefuls must account for at least 30% of the total and less than 50% of the total.

Each subject uses Q-temp-w1 to measure axillary temperature and obtains 3 temperature values, and at the same time uses a reference body temperature patch (Omron thermometer MC-171W) to measure the other side axillary temperature, and 1 is measured Temperature value data. Perform clinical efficacy analysis based on the measurement results. The main evaluation indicators of the trial include clinical bias, limits of agreement, and clinical repeatability.

Q-temp-w1 is a contact type body temperature patch developed by Quanta Computer Co., Ltd. It is mainly used to measure the body temperature of the human body. It can be used for general body temperature measurement, home temperature measurement records and medical tracking.
Omron thermometer(MC-171W) is mainly used to measure the body temperature of the human body. The reference temperature patch used in this study is the OMRON thermometer, model: MC-171W.

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Clinical bias
Periodo de tiempo: 10 min
Use the first value of the 3 test patch body temperature data to calculate and analyze the paired data of the test subject with the reference body temperature data.
10 min

Medidas de resultado secundarias

Medida de resultado
Medida Descripción
Periodo de tiempo
Safety /Adverse Event Outcome Measure
Periodo de tiempo: 20 min / 24 hour (until recovered if it's SAE)
Number of Participants with Serious and Non-Serious Adverse Events
20 min / 24 hour (until recovered if it's SAE)

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Colaboradores

Investigadores

  • Investigador principal: Chin-Jung Chen, MD, Chang Gung Medical fundation

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

24 de marzo de 2021

Finalización primaria (Anticipado)

23 de septiembre de 2022

Finalización del estudio (Anticipado)

23 de septiembre de 2022

Fechas de registro del estudio

Enviado por primera vez

13 de agosto de 2021

Primero enviado que cumplió con los criterios de control de calidad

13 de agosto de 2021

Publicado por primera vez (Actual)

20 de agosto de 2021

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

25 de julio de 2022

Última actualización enviada que cumplió con los criterios de control de calidad

21 de julio de 2022

Última verificación

1 de julio de 2022

Más información

Términos relacionados con este estudio

Otros números de identificación del estudio

  • QOCA-01

Plan de datos de participantes individuales (IPD)

¿Planea compartir datos de participantes individuales (IPD)?

No

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Ensayos clínicos sobre Q-temp-w1

3
Suscribir