Accuracy and Precision Comparison of Temptraq

25. August 2017 aktualisiert von: Sandra Siedlecki, The Cleveland Clinic

Accuracy and Precision Test of TempTraq® Compared to Pulmonary Artery Catheter for Monitoring Temperature in Adults in Intensive Care Unit

Purpose The purpose of this feasibility study is to assess the accuracy and precision of the Temp-Traq thermometer for monitoring body temperature in adult patients under three conditions (hypothermia, normothermia, and hyperthermia).

Specific Aims

  1. To assess the accuracy of the Temp-Traq thermometer as compared to a gold standard (Core temp measured by Pulmonary Artery Catheter)
  2. To assess the precision of the Temp-Traq thermometer over repeated measures
  3. To determine if accuracy &/or precision is consistent in three conditions (hypothermia, normothermia, or hyperthermia)

Studienübersicht

Status

Unbekannt

Bedingungen

Intervention / Behandlung

Detaillierte Beschreibung

Design A repeated measures within-group comparative design will be used for this study.

Sample/Setting To address the specific aims of this study the investigators will identify patients (N = 40) who have a pulmonary artery (PA) catheter in place (patients in intensive care units). Letters of support are attached from the unit managers where the study will take place (see attached).

Procedure Once a patient is identified, the researcher will record basic demographic data (see data collection form attached) and the axillary area will be assessed for visible signs of any skin conditions. If there are no visible signs of a skin condition the researcher will apply the Temp-Traq thermometer to the right or left axilla of the subject and record the location of the placement. The researcher will record both the PA and Temp-Traq recordings of subject's temperature (taken at the same time) on the data collection form (see attachment) at four data points; baseline (5 minutes post application of Temp-Traq) and every two hours (+/- 15 minutes) after baseline X 3. Previous testing of the TempTraq® on children and healthy adults has not resulted in any adverse skin reactions. However, as this is a new thermometer format and there is not a significant body of literature on potential skin reactions, the investigators will assess skin before and after application. Thus, at the end of the last reading the Temp-Traq thermometer will be removed and the skin will be assessed for erythema. If erythema is present the patient's nurse and physician will be notified by the researcher.

Protection of Human Subjects Since non-invasive temperature monitoring poses no risk to patients and is part of usual care and since the investigators will not collect or record any protected health information (PHI), written informed consent will not be sought. Patients will be identified by their study ID number only. The data will be collected by a registered nurse data collector familiar with the intensive care setting and data will be stored in a locked file cabinet in the researchers locked office. Temperature data for the Temp-Traq is collected through the Temp-Traq Application (AP) which will be placed on a Cleveland Clinic Approved and encrypted i-pad. Once the study is completed and all data has been transferred to the study database, the Temp-Traq application and all of its data will be deleted from the i-pad.

Analysis The investigators will use the TOST (Two One Sided t-tests) method for equivalence testing. This method requires a defined range of mean differences between two test methods, an estimate of the precision of the measurement of the two systems, and an estimate of the size of the possible difference in the means of the two methods under consideration. The TOST null hypothesis is a joint null hypothesis that the mean measurement differences between the two methods is greater than a critical lower bound and less than a critical upper bound. If the null is rejected then the investigators can conclude that the absolute difference of the means for the two groups falls within the specified range. It was determined that to be considered equivalent with respect to accuracy the mean measurements of the two methods should be within ± 0.2 degrees of each other. Precision tests for Temp-Traq thermometer using accepted ASTM test methods for various combinations of temperature and humidity provided a range of measurement variation between .0000435 and .019928 with a mean of 0.000254500. If the investigators assume both test methods exhibit the same levels of precision then the variance of the differences between their two means will reduce to two times the values listed previously. This will result in estimates of the standard deviations of the mean differences of .0093, .032, and .063 respectively.

The two methods for body temperature that will be examined are core temperature measurements with a PA catheter and temperature measurements using the Temp-Traq (measured in Celsius). Paired temperature measurements for both methods will be taken simultaneously every 2 hours for a 6 hour period. This will provide 4 repeated temperature measures per patient. The simultaneous measurements within a given patient should exhibit a high degree of correlation. If .7 is chosen as the lower bound for the correlation between the paired readings (higher correlations would result in an estimate of fewer patient samples thus the choice of .7 is conservative) and if the average standard deviation of the mean of the differences is used as an estimate of the expected variability, then it would be possible to declare measurement equivalence for a difference of ±.19 with a sample size of 40 patients. This assumes the ASTM measurements of instrument variation are representative of the within patient variance. Given the vagaries of patient-to-patient this may be too optimistic an estimate.

A sample size of 40 patients will provide 80% or greater power for testing for equivalence between methods where the mean difference is .1 or less for up to a ten-fold increase in the estimates of the variability of the differences based on the ASTM precision measures and for a mean difference of .15 for up to a five-fold increase in these same measures. If the sample size is 40 then the power for the test of equivalence, where the difference is .15 and the standard deviation is a ten-fold increase, is 35%. Agreement testing: The Bland-Altman test will be used to test the agreement of the two methods. This test will permit a check for significance of bias between the two measures (the average difference between the two measures) as well as a check for significant trending over the range of the measurements. If the bias is significant then the results of the two differ by an overall offset in their measurements. If the trending (slope of the regression line) is significant then the difference between the two measurements changes as their magnitude changes which means the two methods are not in agreement.

The only measurements of instrument precision are those provided by the ASTM assessment. It is reasonable to assume the within patient measurement variation will be greater than the measurements from the controlled testing. A sample size of 40 patients is, from a statistical standpoint, a reasonable number and this sample size provides acceptable power for a mean difference of .1 for a ten-fold increase in the estimate of minimum test variability. This sample size will also provide enough data for the Bland-Altman test for agreement. Because this is a repeated measures test and even one missing measurement will impact the results, the investigators plan to over-sample in order to assure the study has the power to answer the questions posed.

Studientyp

Beobachtungs

Einschreibung (Tatsächlich)

50

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

21 Jahre und älter (Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

N/A

Studienberechtigte Geschlechter

Alle

Probenahmeverfahren

Nicht-Wahrscheinlichkeitsprobe

Studienpopulation

To address the specific aims of this study the investigators will identify patients (N = 40) who have a pulmonary artery (PA) catheter in place (patients in intensive care units). Letters of support are attached from the unit managers where the study will take place.

Beschreibung

Inclusion Criteria:

Inclusion Criteria Adult patient (male or female) in intensive care unit who will have a PA catheter in place for at least the next 8 hours; and who have no visible skin condition to the axillary region upon inspection by the research nurse are eligible for this study

Exclusion Criteria:

  1. Patients younger than 21
  2. Patients who will not have a PA catheter in place for the next 8 hours
  3. Patients who have a visible skin condition to the axillary region upon inspection by the research nurse
  4. Patients who will be going for procedures or for some reason will not be available for the 6 hours of the study

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
Accuracy and precision of TempTraq thermometer- Baseline
Zeitfenster: Baseline
Skin surface temperature readings measured in Celsius using Temp Traq vs measured by Temp Traq and by Pulmonary artery catheter core thermister
Baseline
Accuracy and precision of TempTraq thermometer-Time 2
Zeitfenster: 2 hours post baseline
Skin surface temperature readings measured in Celsius using Temp Traq vs measured by Temp Traq and by Pulmonary artery catheter core thermister
2 hours post baseline
Accuracy and precision of TempTraq thermometer-Time 3
Zeitfenster: 4 hours post baseline
Skin surface temperature readings measured in Celsius using Temp Traq vs measured by Temp Traq and by Pulmonary artery catheter core thermister
4 hours post baseline
Accuracy and precision of TempTraq thermometer-Time 4
Zeitfenster: 6 hours post baseline
Skin surface temperature readings measured in Celsius using Temp Traq vs measured by Temp Traq and by Pulmonary artery catheter core thermister
6 hours post baseline

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Publikationen und hilfreiche Links

Die Bereitstellung dieser Publikationen erfolgt freiwillig durch die für die Eingabe von Informationen über die Studie verantwortliche Person. Diese können sich auf alles beziehen, was mit dem Studium zu tun hat.

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

15. Dezember 2016

Primärer Abschluss (Tatsächlich)

1. Februar 2017

Studienabschluss (Voraussichtlich)

1. September 2017

Studienanmeldedaten

Zuerst eingereicht

2. Februar 2017

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

25. August 2017

Zuerst gepostet (Tatsächlich)

29. August 2017

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

29. August 2017

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

25. August 2017

Zuletzt verifiziert

1. August 2017

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Andere Studien-ID-Nummern

  • 16-115

Plan für individuelle Teilnehmerdaten (IPD)

Planen Sie, individuelle Teilnehmerdaten (IPD) zu teilen?

NEIN

Beschreibung des IPD-Plans

IPD will not be shared

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .