- ICH GCP
- Amerikanska kliniska prövningsregistret
- Klinisk prövning NCT05776706
Klinisk prövning för validering av AR-baserat neuronavigationssystem
Exploratory Clinical Trial for the Evaluation of the Accuracy of Augmented Reality-Based Neuronavigation System
The goal of this clinical trial is to test augmented reality (AR) based neuronavigation system in surgeries for patients of brain neoplasm or cerebral vascular disease. The main questions it aims to answer are:
• AR based neuronavigation system can achieve accuracy that is not inferior to conventional intraoperative navigation system.
Participants will participate the study after informed consent. When participants undergo surgery for their brain tumor, we will set up 2 types of neuronavigation, conventional navigation system and developed AR based neuronavigation system. Surgeon will plan and conduct surgery based on only conventional navigation system, but 3D errors at several selected points between two types of navigation will be measured and analyzed.
Studieöversikt
Status
Betingelser
Intervention / Behandling
Detaljerad beskrivning
Being developed AR navigation were reported in our following papers:
- Dho YS, et al., Development of an inside-out augmented reality technique for neurosurgical navigation. Neurosurgical Focus. 2021 Aug 1;51(2):E21. (doi.org/10.3171/2021.5.FOCUS21184)
- Moon HC, et al., Navigation of frameless fixation for gamma knife radiosurgery using fixed augmented reality. Scientific Reports. 2022 Mar 16;12(1):1-0. (doi.org/10.3171/2021.5.FOCUS21184)
This study aims to evaluate the accuracy of anatomical localization of a newly developed augmented reality-based neurosurgery navigation. After 3-dimensional (3D) modeling of the brain of a patient with brain tumor or cerebral vascular disease through 3D image segmentation extraction and modeling, the Augmented Reality (AR)-based navigation developed by this research team can be used with a commercially available visualization device, such as iPad, iPhone, and Hololens2, to evaluate the accuracy.
The AR-based navigation obtained approval from the Ministry of Food and Drug Safety of Republic of Korea (class 2 medical device) in February 2022. Through this clinical tria that compare accuracy with existing conventional intraoperative neuronavigation system, we will evaluate whether it shows equivalent performance to replace the existing navigation.
Studietyp
Inskrivning (Beräknad)
Fas
- Inte tillämpbar
Kontakter och platser
Studiekontakt
- Namn: Chul-Kee Park, MD, PhD
- Telefonnummer: +82220720347
- E-post: nsckpark@snu.ac.kr
Studera Kontakt Backup
- Namn: Sun Mo Nam, MD
- Telefonnummer: +82020722357
- E-post: md.nsm.ns@gmail.com
Studieorter
-
-
-
Seoul, Sydkorea, 03080
- Rekrytering
- Seoul National University Hospital
-
Kontakt:
- Sun Mo Nam, MD
- Telefonnummer: 8220722357
- E-post: md.nsm.ns@gmail.com
-
Underutredare:
- Yun-Sik Dho, MD, PhD
-
-
Seoul
-
Seoul, Seoul, Sydkorea, 03080
- Har inte rekryterat ännu
- Seoul National University Hospital
-
Kontakt:
- Chul-Kee Park, MD, PhD
- Telefonnummer: +82220720347
- E-post: nsckpark@snu.ac.kr
-
Kontakt:
- Sun Mo Nam, MD
- Telefonnummer: +821096713692
- E-post: md.nsm.ns@gmail.com
-
-
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
Tar emot friska volontärer
Kön som är behöriga för studier
Beskrivning
Inclusion Criteria:
- Adults aged 18 years or older with a brain tumor or cerebrovascular disease who have undergone preoperative CT
- Scheduled for surgical treatment of the brain tumor or cerebrovascular disease
- Surgical position allowing exposure and measurement of bilateral Kocher's points and the glabella without significant head rotation
Exclusion Criteria:
- Patient or legal representative does not consent to participation, including consent to intraoperative video recording
- Anatomical deformity of the skull or facial region due to previous surgery or other causes that precludes identification of Kocher's points or the glabella
- Emergency surgery
- Surgical position (e.g., prone, lateral rotation of 90° or more) that does not allow simultaneous exposure of bilateral Kocher's points and the glabella
Studieplan
Hur är studien utformad?
Designdetaljer
- Primärt syfte: Övrig
- Tilldelning: N/A
- Interventionsmodell: Enskild gruppuppgift
- Maskning: Ingen (Open Label)
Vapen och interventioner
Deltagargrupp / Arm |
Intervention / Behandling |
|---|---|
|
Experimentell: Single arm
We will set up 2 navigation systems, newly developed AR-based navigation and conventional navigation system, to surgeries of all participants. Surgeons will perform all surgical planning and operations with reference to the conventional navigation system. In this process, the errors of the existing navigation and the newly developed navigation will be measured and compared in 3D at several points, and the points are as follows:
|
Patient-specific 3D models are generated from preoperative CT using MEDIP PRO and displayed on a tablet (iPad) via MEDIP PRO AR, superimposed on the patient using markerless surface registration. Registration and TRE measurement are video-recorded for retrospective analysis of registration time. The device is used for accuracy evaluation only and does not guide any surgical step. |
Vad mäter studien?
Primära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Target Registration Error (TRE) of AR-based navigation at anatomical landmarks
Tidsram: Intraoperative, immediately after AR registration before surgical draping on the day of surgery
|
Linear distance (mm) between the landmark position displayed by MEDIP PRO AR and the actual position on the patient's skin/skull surface, measured directly with a ruler at three fixed landmarks per participant (right Kocher's point, left Kocher's point, glabella). Measured immediately after registration (pre-drape). Reported as mean ± SD or median (IQR) overall and per landmark. |
Intraoperative, immediately after AR registration before surgical draping on the day of surgery
|
|
Registration time of AR-based navigation
Tidsram: Intraoperative, on the day of surgery
|
Time (minutes, seconds) from AR software launch to completion of registration, determined by retrospective analysis of intraoperative video recording. Reported separately for manual and automatic registration. |
Intraoperative, on the day of surgery
|
Sekundära resultatmått
Resultatmått |
Åtgärdsbeskrivning |
Tidsram |
|---|---|---|
|
Post-drape TRE after QR marker-based registration restoration
Tidsram: Intraoperative, immediately after draping, on the day of surgery
|
TRE (mm) at the same three landmarks measured after surgical draping and restoration of registration using QR markers attached to the skin. |
Intraoperative, immediately after draping, on the day of surgery
|
|
Incidence of registration drift after draping
Tidsram: Intraoperative, immediately after draping, on the day of surgery
|
Proportion of landmarks showing visible displacement of the AR overlay of ≥7.5 mm after draping, assessed on a second tablet held in a fixed mount. |
Intraoperative, immediately after draping, on the day of surgery
|
|
Adverse events and device malfunctions
Tidsram: Intraoperative through end of surgery, on the day of surgery
|
Number and description of adverse events and device errors occurring during AR navigation application. |
Intraoperative through end of surgery, on the day of surgery
|
|
One-step processing time from CT acquisition to 3D model delivery (SNUH only)
Tidsram: Preoperative, within 24 hours before surgery
|
Total elapsed time from CT image receipt on the server to delivery of the 3D mesh to the tablet, including automatic segmentation, calculated from system log timestamps. Success of automatic segmentation (no manual correction required) is also recorded. |
Preoperative, within 24 hours before surgery
|
Samarbetspartners och utredare
Studieavstämningsdatum
Studera stora datum
Studiestart (Faktisk)
Primärt slutförande (Beräknad)
Avslutad studie (Beräknad)
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Faktisk)
Uppdateringar av studier
Senaste uppdatering publicerad (Faktisk)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Nyckelord
Ytterligare relevanta MeSH-villkor
- Cerebrovaskulära störningar
- Hjärnsjukdomar
- Sjukdomar i centrala nervsystemet
- Sjukdomar i nervsystemet
- Kärlsjukdomar
- Hjärt-kärlsjukdomar
- Neoplasmer efter plats
- Neoplasmer
- Medfödda abnormiteter
- Kardiovaskulära avvikelser
- Neoplasmer i nervsystemet
- Neoplasmer i centrala nervsystemet
- Missbildningar i nervsystemet
- Intrakraniella arteriella sjukdomar
- Aneurysm
- Vaskulära missbildningar
- Arteriovenösa missbildningar
- Centrala nervsystemet Vaskulära missbildningar
- Medfödda, ärftliga och neonatala sjukdomar och abnormiteter
- Beteende
- Rumsligt beteende
- Neoplasmer i hjärnan
- Intrakraniell aneurysm
- Intrakraniella arteriovenösa missbildningar
- Spatial Navigation
Andra studie-ID-nummer
- 2204-094-1317
- 202012E08 (Annat bidrag/finansieringsnummer: Korea Medical Device)
Plan för individuella deltagardata (IPD)
Planerar du att dela individuella deltagardata (IPD)?
Läkemedels- och apparatinformation, studiedokument
Studerar en amerikansk FDA-reglerad läkemedelsprodukt
Studerar en amerikansk FDA-reglerad produktprodukt
produkt tillverkad i och exporterad från U.S.A.
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .